Skip to content

StimRouter Implantable Kit ST2-1000Recalls & reported problems

Bioness Inc. · StimRouter Implantable Lead · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815962020016), received between Feb 17, 2022 and Jul 30, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Incomplete or Inadequate Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Incomplete or Inadequate Connection1
  • Insufficient Device Problem Information1
  • Material Separation1
  • Migration1

Patient problems

  • Erythema1
  • Pain1
  • Skin Infection1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 30, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    A patient implanted with a Talisman Implant Kit on (b)(6) 2025 reported during a follow up appointment ((b)(6) 2026) that the incision was not fully healed. The StimRouter Implantable Lead and Lead Introducer Kit were utilized during the Talisman Implantation. Decision by the patient's physician was to leave the dissolvable stitch for wound closure in place. Redness and swelling increased. While in (b)(6), patient visited emergency room for worsening local symptoms. The patient was prescribed Clindamycin and recommended to return to Duke Pain Clinic for follow up care as well as a potential explant. The StimRouter Lead was explanted without issue (b)(6) 2026, however the TalisMann (Implantable Pulse Generator Kit) IPG became detached during the explant process. The patient was referred to the implanting physician for surgical excision of the TalisMann IPG in the OR. On (b)(6) 2026 the device was successfully excised.The implant of the TalisMann model TM-1000 PNS device utilized the StimRouter model ST2-1000 StimRouter Implantable Lead and Lead Introducer Kit, Sterile. This report is for the StimRouter model ST2-1000 StimRouter Implantable Lead and Lead Introducer Kit, Sterile. A se

    Patient problems: Erythema; Skin Infection; Skin Inflammation/ Irritation

    MAUDE report 50091205 ↗
  • InjuryFeb 17, 2022Material Separation; Migration; Incomplete or Inadequate Connection; Material Separation; Migration; Incomplete or Inadequate Connection

    BIONESS RECEIVED A MEDWATCH REPORT FROM THE FDA ON 24 JAN 2022. AS REPORTED "WE IMPLANTED A BIONESS STIMROUTER DEVICE IN THE RIGHT SUPRASCAPULAR NERVE IN 2020, WHICH MIGRATED OUT OF THE SUPRASCAPULAR NOTCH. ON REMOVING THE ELECTRODE TODAY, WE CAREFULLY EXTRACTED THE PROXIMAL ASPECT, AND THE ELECTRODE BROKE IN HALF. SIGNIFICANTLY, MORE TIME AND DEEP TISSUE DISSECTION INTO THE SUPRASPINATUS REGION WAS REQUIRED TO EXTRACT THE FOREIGN BODY. PATIENT WAS HAVING SEVERE PAIN FOR THE LAST 9 MONTHS AND WILL HAVE A PROLONGED RECOVERY DUE TO THE LEAD EXTRACTION DIFFICULTY."

    Patient problems: Pain

    MAUDE report 13548820 ↗

Other STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF) devices

All STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page