Skip to content

ARROW IPN033729Recalls & reported problems

Teleflex Incorporated · EZ-IO(R) 15mm Needle Set · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816000011966), received between Nov 16, 2023 and Apr 14, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (3 reports).

3 other recalls cover devices cleared under the same FDA submission (K141117); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking3
  • Separation Failure2
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2024Mechanical Jam; Physical Resistance/Sticking; Mechanical Jam; Physical Resistance/Sticking

    IT WAS REPORTED "THEY REPORT INSERTING THE 15MM NEEDLE INTO THE PROXIMAL TIBIA AS DIRECTED AND BEING UNABLE TO UNSCREW THE STYLET FROM THE CANNULA. AFTER MULTIPLE PROVIDER ATTEMPTS TO UNSCREW THE STYLET, A SECOND 15MM WAS USED ON THE OPPOSITE LEG WITH SUCCESS. THE NEEDLE SET WAS RECOVERED FROM THE SHARPS SAFETY BIN AND IT WAS UNKNOWN HOW, BUT EVENTUALLY THE TWO PIECES WERE UNSCREWED FROM EACH OTHER. WHEN SPEAKING WITH THE REPORTING PARTY, I ASKED ABOUT THE CERTAINTY OF THE NEEDLE SET BEING CONNECTED TO THE CATHETER UPON INSERTION AND NOT JUST THE STYLET BEING PLACED INTO THE PATIENT'S BODY. THE REPORTING PARTY COULD NOT ANSWER IF THIS WAS THE POSSIBLE OCCURRENCE. THEY REPORTED HAVING BOTH PIECES OF THE NEEDLE SET. THE DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE DEATH. THE ON SCENE MEDIC SUPERVISOR. REPORTING PARTY STATED THIS WAS A SIDS CASE. RESUSCITATION WOULD NOT HAVE CHANGED THE OUTCOME, EVEN HAD THE INITIAL NEEDLE NOT FAILED." THE PATIENT'S CURRENT CONDITION IS REPORTED AS "DECEASED".

    Patient problems: Insufficient Information

    MAUDE report 19103115 ↗
  • MalfunctionNov 16, 2023Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking

    IT WAS REPORTED BY THE CUSTOMER THAT "[THE USER] HAD TWO 15MM EZ-IO NEEDLES FUNCTION INAPPROPRIATELY ON SCENE OF A CALL. THE PATIENT WAS A 2YO IN CARDIAC ARREST PRIOR TO THE ARRIVAL OF EMS, AND WAS NOT ABLE TO BE REVIVED AFTER RESUSCITATIVE EFFORTS. THE PARAMEDIC ON SCENE REPORTED TO HIM THAT THE INITIAL ATTEMPT FOR INTRAOSSEOUS ACCESS RESULTED IN THE INABILITY TO UNSCREW THE STYLET FROM THE HUB. IN A SECOND ATTEMPT, A DIFFERENT 15MM EZ-IO NEEDLE WAS REMOVED FROM PACKAGING WITH THE PARAMEDIC STATING THAT, PRIOR TO ATTEMPT, HE TESTED THE ABILITY OF THE STYLET TO SCREW AND UNSCREW FROM THE HUB. THE PARAMEDIC THEN STATED THAT THE ONCE THE NEEDLE WAS PLACED INTO THE PATIENT, THE STYLET ON THAT NEEDLE WOULD NOT UNSCREW FROM THE HUB, BY HAND OR WITH PLIERS. AN AMR UNIT THEN ARRIVED ON SCENE AND WAS ABLE TO SUCCESSFULLY PLACE AN EZ-IO NEEDLE FROM THEIR SUPPLY. HE ASSURED ME THAT THE PARAMEDIC ON SCENE WAS VERY EXPERIENCED IN EZ-IO, AND WOULD NOT HAVE BENT THE NEEDLE OR REMOVED THE STYLET PRIOR TO ATTEMPT". ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER STATES THAT THE PATIENT HAD BEEN SICK FOR 2 DAYS WITH FEVER AND A COUGH. UPON EMS'S ARRIVAL ON SCENE, THE PATIENT WAS IN ASYSTOLE WITH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18149358 ↗
  • MalfunctionNov 16, 2023Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking

    IT WAS REPORTED BY THE CUSTOMER THAT "[THE USER] HAD TWO 15MM EZ-IO NEEDLES FUNCTION INAPPROPRIATELY ON SCENE OF A CALL. THE PATIENT WAS A 2YO IN CARDIAC ARREST PRIOR TO THE ARRIVAL OF EMS, AND WAS NOT ABLE TO BE REVIVED AFTER RESUSCITATIVE EFFORTS. THE PARAMEDIC ON SCENE REPORTED TO HIM THAT THE INITIAL ATTEMPT FOR INTRAOSSEOUS ACCESS RESULTED IN THE INABILITY TO UNSCREW THE STYLET FROM THE HUB. IN A SECOND ATTEMPT, A DIFFERENT 15MM EZ-IO NEEDLE WAS REMOVED FROM PACKAGING WITH THE PARAMEDIC STATING THAT, PRIOR TO ATTEMPT, HE TESTED THE ABILITY OF THE STYLET TO SCREW AND UNSCREW FROM THE HUB. THE PARAMEDIC THEN STATED THAT THE ONCE THE NEEDLE WAS PLACED INTO THE PATIENT, THE STYLET ON THAT NEEDLE WOULD NOT UNSCREW FROM THE HUB, BY HAND OR WITH PLIERS. AN AMR UNIT THEN ARRIVED ON SCENE AND WAS ABLE TO SUCCESSFULLY PLACE AN EZ-IO NEEDLE FROM THEIR SUPPLY. HE ASSURED ME THAT THE PARAMEDIC ON SCENE WAS VERY EXPERIENCED IN EZ-IO, AND WOULD NOT HAVE BENT THE NEEDLE OR REMOVED THE STYLET PRIOR TO ATTEMPT". ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER STATES THAT THE PATIENT HAD BEEN SICK FOR 2 DAYS WITH FEVER AND A COUGH. UPON EMS'S ARRIVAL ON SCENE, THE PATIENT WAS IN ASYSTOLE WITH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18149334 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K141117. One submission often covers many models, so check the product and lot details in each record.

  • Z-0689-2020 · Oct 8, 2019 · Open, Classified
    Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides
  • Z-0690-2020 · Oct 8, 2019 · Open, Classified
    Arrow EZ-IO Intraosseous Vascular Access System 45mm Needle+Stabilizer Kit, Product Code 9079P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides
  • Z-0688-2020 · Oct 8, 2019 · Open, Classified
    Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.