TalisMann Implant Kit TM-1000Recalls & reported problems
Bioventus LLC · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816986020549), received between Jun 29, 2026 and Jul 30, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information2
- Break1
- Lack of Effect1
Patient problems
- Skin Infection2
- Erythema1
- Inadequate Pain Relief1
- Pain1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJul 30, 2026Break; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
A patient implanted with a Talisman Implant Kit on (b)(6) 2025 reported during a follow up appointment ((b)(6) 2026) that the incision was not fully healed. The StimRouter Implantable Lead and Lead Introducer Kit were utilized during the Talisman Implantation. Decision by the patient's physician was to leave the dissolvable stitch for wound closure in place. Redness and swelling increased. While in NY, patient visited emergency room for worsening local symptoms. The patient was prescribed Clindamycin and recommended to return to (b)(6) for follow up care as well as a potential explant. The StimRouter Lead was explanted without issue (b)(6) 2026, however the TalisMann (Implantable Pulse Generator Kit) IPG became detached during the explant process. The patient was referred to the implanting physician for surgical excision of the TalisMann IPG in the operating room. On (b)(6) 2026 the device was successfully excised.The implant of the TalisMann model TM-1000 PNS device utilized the StimRouter model ST2-1000 StimRouter Implantable Lead and Lead Introducer Kit, Sterile. This report is for the TalisMann model TM-1000 PNS device. A seperate report is submitted for the ST2-1000 StimRouter
Patient problems: Erythema; Skin Infection; Skin Inflammation/ Irritation
MAUDE report 50091206 ↗InjuryJun 29, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Lack of Effect
PATIENT EXPERIENCED LOCALIZED PAIN AND DISCHARGE APPROXIMATELY 10 DAYS POST IMPLANT OF THE STIMROUTER PNS DEVICE. THE PATIENT ALSO EXPERIENCED A LACK OF PAIN RELIEF FROM STIMULATION. THE PATIENT SENT PHOTOS TO THE SALES REP AND THE PHYSICIAN WHO PRESCRIBED KEFLEX AT A FREQUENCY OF FOUR TIMES PER DAY. AT THE PATIENTS IN PERSON FOLLOW-UP APPOINTMENT, THE PHYSICIAN RECOMMENDED AN EXPLANT DUE TO AN INFECTION AROUND THE LEAD, WHICH THE PATIENT DID NOT WANT AT THAT TIME. THE PHYSICIAN AND THE PATIENT DECIDE TO INCREASE THE INCISION SITE TO ALLOW FOR DRAINAGE AND PRESCRIBED A NEW ORAL ANTIBIOTIC TO CONTROL THE INFECTION AND SALVAGE THE IMPLANT. THE IMPLANT OF THE TALISMANN MODEL TM-1000 PNS DEVICE UTILIZED THE STIMROUTER MODEL ST2-1000 STIMROUTER IMPLANTABLE LEAD AND LEAD INTRODUCER KIT, STERILE. THIS REPORT IS FOR THE TALISMANN MODEL TM-1000 PNS DEVICE.
Patient problems: Inadequate Pain Relief; Pain; Skin Infection
MAUDE report 25654431 ↗
Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices
- Quattrode™ 3166St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)429 reports
- Quattrode™ 3169St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)401 reports
- PNS Freedom Permanent Tined Neurostimulator Receiver Kit Channel A STQ4-RCV-A0Curonix LLC · Stimulator, Peripheral Nerve, Implanted (Pain Relief)8 deaths364 reports
- Nalu Medical, Inc. 71019-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall225 reports
- StimQ Receiver Stimulator Kit Channel A US w/Receiver STQ4-RCV-A0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)3 deaths270 reports
- Nalu Medical, Inc. 71020-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)213 reports
- StimQ Spare Lead Kit Channel B US w/Receiver STQ4-SPR-B0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)126 reports
- Nalu Medical, Inc. 71005-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.