Stratum STRM-MDS-3550STRecalls & reported problems
Medartis Inc. · STRM MDS Screw 3.5x50mm ST · Class 2 · Implant
All models of Stratum (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Nov 11, 2019Class IITerminated
Components 2mm shorter than indicated on the package label
Z-0778-2020 · Root cause: Packaging process control · Recall details · FDA record ↗
FDA clearance and classification
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Stratum STRM-MDS-3550ST
Has the Stratum STRM-MDS-3550ST been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0778-2020 (Class II), began on Nov 11, 2019: Components 2mm shorter than indicated on the package label. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
Is the Stratum STRM-MDS-3550ST MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Stratum STRM-MDS-3550ST, and is it still sold?
The FDA's device database lists Medartis Inc. as the company responsible for it (brand Stratum, model STRM-MDS-3550ST). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page