StimQ Freedom-4A Trial Kit Channel A US w/ Receiver FR4A-TRL-A0Recalls & reported problems
Stimwave Technologies Incorporated · StimQ Peripheral Nerve Stimulator System StimQ Freedom-4A Trial Kit Channel A US w/ Receiver · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00818225020716), received between Apr 21, 2021 and Jun 29, 2023 — classified as 8 injuries.
6 other recalls cover devices cleared under the same FDA submission (K171366); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
Patient problems
- Skin Inflammation/ Irritation3
- Electric Shock2
- Pain2
- Blister1
- Discomfort1
- Erosion1
- Hemorrhage/Blood Loss/Bleeding1
- Itching Sensation1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED UNINTENDED STIMULATION WHEN MOVING THEIR ARM. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Electric Shock
MAUDE report 17233041 ↗InjuryJun 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED THE LEAD FELL OUT OF THEIR WRIST WHILE THEY WERE SLEEPING. INFORMATION REGARDING THE PATIENT OUTCOME IS NOT AVAILABLE.
Patient problems: Erosion
MAUDE report 17195016 ↗InjuryMay 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED THEY FELT A CURRENT/SHOCK SENSATION DURING AN MRI. THE MRI WAS STOPPED IMMEDIATELY. THE DEVICE IS WORKING FINE SINCE THE INCIDENT.
Patient problems: Electric Shock
MAUDE report 17021939 ↗InjuryApr 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT WENT TO THE ER DUE TO CONTINUOUS BLEEDING THE DAY AFTER THEIR IMPLANT PROCEDURE. ON (B)(6) 2023, WHEN THE ER STAFF REMOVED THE BANDAGE THE LEAD CAME OUT. THE PATIENT IS FINE AND NO FURTHER ISSUES HAVE BEEN REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16762825 ↗InjuryMar 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED AN ITCHING REACTION DURING THEIR TRIAL AND ASKED TO HAVE THE LEADS REMOVED EARLY. THEY WERE GIVEN CORTISONE CREAM TO TREAT THE AREA. THE PATIENT WAS ALSO COMPLAINING OF SEVERE RIGHT SIDED PAIN IN THEIR BACK AND LEG, WHICH THE DOCTOR SAID WAS RELATED TO THEIR L5-S1 HERNIATED DISC. AN EXPLANT PROCEDURE WAS PERFORMED ON (B)(6) 2023. THE PATIENT IS NO LONGER ITCHING AND THERE WAS NO INFECTION PRESENT.
Patient problems: Itching Sensation; Pain
MAUDE report 16501598 ↗InjuryFeb 20, 2022Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED A SKIN REACTION AND THE TRIAL LEAD WAS REMOVED ON (B)(6) 2022. THE PATIENT'S SKIN WAS BURNING AND DEVELOPED EXTENSIVE BLISTERS. THE PATIENT IS DOING MUCH BETTER AND ALMOST COMPLETELY HEALED.
Patient problems: Blister; Skin Inflammation/ Irritation
MAUDE report 13565460 ↗InjuryAug 2, 2021Adverse Event Without Identified Device or Use Problem
PATIENT IS HAVING DISCOMFORT RELATED TO THE LEAD EVEN WHEN STIMULATION IS TURNED OFF.
Patient problems: Discomfort; Skin Inflammation/ Irritation
MAUDE report 12264336 ↗InjuryApr 21, 2021Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2021, THE PATIENT ATTENDED A FOLLOW-UP APPOINTMENT WITH THE IMPLANTING CLINICIAN FOR A LEAD PULL FOR A SUSPECTED FOOT INFECTION.
Patient problems: Pain; Post Operative Wound Infection; Skin Inflammation/ Irritation; Swelling/ Edema
MAUDE report 11703701 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171366. One submission often covers many models, so check the product and lot details in each record.
- Z-2933-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models: 1. FR4A-RCV-AO US and 2. FR4A-RCV-A0 EU - Product Usage: is used for peripheral neural stimulation to provide therapeutic re
- Z-2932-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model STQ4-SPR-BO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractabl
- Z-2928-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model 1. STQ4-RCV-A0 US and 2.STQ4-RCV-A0 EU - Product Usage: is used for peripheral neural stimulation to provide therapeutic relie
- Z-2929-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Model FR4A-TRL-AO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractabl
- Z-2930-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models: 1. FR4A-TRL-BO US and 2. FR4A-TR-BO - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief
- Z-2931-2020 · Jul 15, 2020 · Open, ClassifiedStimQ Peripheral Nerve Stimulator System (PNS) Models FR4A-SPR-BO US - Product Usage: is used for peripheral neural stimulation to provide therapeutic relief for chronic, intractab
Other Stimulator, Peripheral Nerve, Implanted (Pain Relief) devices
- Quattrode™ 3166St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)429 reports
- Quattrode™ 3169St. Jude Medical, Inc. · STIMULATOR, PERIPHERAL NERVE, IMPLANTED (PAIN RELIEF)401 reports
- PNS Freedom Permanent Tined Neurostimulator Receiver Kit Channel A STQ4-RCV-A0Curonix LLC · Stimulator, Peripheral Nerve, Implanted (Pain Relief)8 deaths364 reports
- Nalu Medical, Inc. 71019-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall225 reports
- StimQ Receiver Stimulator Kit Channel A US w/Receiver STQ4-RCV-A0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)3 deaths270 reports
- Nalu Medical, Inc. 71020-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)213 reports
- StimQ Spare Lead Kit Channel B US w/Receiver STQ4-SPR-B0Stimwave Technologies Incorporated · Stimulator, Peripheral Nerve, Implanted (Pain Relief)126 reports
- Nalu Medical, Inc. 71005-USNalu Medical, Inc. · Stimulator, Peripheral Nerve, Implanted (Pain Relief)1 recall73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page