Skip to content

SOL-M™ Hypodermic Needle 112506Recalls & reported problems

Sol-Millennium Medical, Inc. · Sterile hypodermic needles intended to inject fluids into, or withdraw fluids from, parts of the body below the suface of the skin. · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00818392015140), received between May 16, 2025 and May 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2025Dull, Blunt; Dull, Blunt

    IT WAS REPORTED: ACCORDING TO BASTOS VIEGAS, DIFFICULTY WHEN PENETRATING THE SKIN. USED IN LOCAL ANESTHESIA AND SCLEROTHERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035526 ↗
  • MalfunctionMay 16, 2025Device-Device Incompatibility; Device-Device Incompatibility

    "INCIDENT DESCRIPTION: DURING DIRECT INTRAVENOUS INJECTION OF RADIOPHARMACEUTICAL MEDICINE, HEALTHCARE PROVIDER NOTES THAT NEEDLE IS BIASED IN ITS CAP. IT'S ALSO HARD TO SCREW THE NEEDLE HUB ON SYRINGES, ESPECIALLY THE LUER-LOCK 2.5 ML REF (B)(4) MANUFACTURED BY TERUMO, AND ALSO THE LUER-LOCK PLASTIPAK 5ML REF (B)(4) AND 10ML (B)(4) BECTON DICKINSON MANUFACTURER. IT HAS BEEN NOTED THAT NEEDLES WERE OFTEN SIDEWAYS IN THEIR PROTECTIVE CAPS, THAT THEY WERE VERY SOFT AND THERE IS A LOOSENESS IN THE SCREW THREAD AFTER THE SYRINGE HAS BEEN SCREWED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035503 ↗

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.