Skip to content

Cannulated Compression Headless Screw 04.334.440Recalls & reported problems

Tyber Medical, LLC · 4.0MM TI CANN COMPRESSION HEADLESS SCREW- LT- 40MM · Class 2 · Implant

All models of Cannulated Compression Headless Screw (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00819917024395), received between Apr 24, 2023 and Oct 21, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Foreign Body In Patient1
  • Implant Pain1
  • Nonunion/Delayed-union Bone Fracture1
  • Radiolucent Lines1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    THIS SUBMISSION IS FOR THE SECOND DEVICE ASSOCIATED WITH 3012966183-2024-00018. ADVERSE EVENT FROM THE CLINICAL STUDY: AT 9-MONTHS POST-OPERATIVE IMAGING DEMONSTRATED SCREW LOOSE, RADIOLUCENCY; THE SUBJECT WAS ORDERED A BONE GROWTH STIMULATOR AND TOE-OFF BRACE. AT 11-MONTHS POST-OPERATIVE SUBJECT CONTINUED TO HAVE PAIN WITH AMBULATION AND PLANNED TO CONTINUE TO USE BONE GROWTH STIMULATOR AND SIMULTANEOUSLY SCHEDULE SURGERY TO HAVE HARDWARE REMOVED IF NO IMPROVEMENT IS SEEN WITH BONE GROWTH STIMULATOR; IMAGING DEMONSTRATED THAT THE RADIOLUCENCY REMAINED VISIBLE HOWEVER NO NEW SIGNS OF LOOSENING. AT 12- MONTHS POST-OPERATIVE SUBJECT CONTINUED TO USE BONE GROWTH STIMULATOR WITH NO SIGNS OF INCREASED HEALING; IMAGING SHOWS RADIOLUCENCY REMAINED PRESENT. AT 13.5-MONTHS POST-OPERATIVE SUBJECT UNDERWENT HARDWARE REMOVAL SURGERY DUE TO THE NONUNION. ALL HARDWARE WAS REMOVED DURING THE PROCEDURE AND WAS REPLACED WITH A DIFFERENT BRAND'S SYSTEM. SUBJECT WAS TERMINATED FROM THE STUDY UPON REMOVAL OF HEADLESS SCREWS.

    Patient problems: Implant Pain; Nonunion/Delayed-union Bone Fracture; Radiolucent Lines

    MAUDE report 20498367 ↗
  • InjuryApr 24, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT ON (B)(6) 2023, DURING INTRA-OP AT THE TIME OF ORIF MEDIAL EPICONDYLE THE HEADLESS SCREW BROKE OFF IN PATIENT ON INSERTION. THERE WAS A SURGICAL DELAY OF TWENTY (20) MINUTES. THE ACTION WAS TAKEN AS USED BROKEN SCREW REMOVAL SET TO RETRIEVE BROKEN HALF OF THE SCREW. INSERTED ANOTHER SCREW. PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE FRAGMENTS GENERATED. THE FRAGMENT WAS DIFFICULT TO REMOVE, AND WE HAD TO USE BROKEN SCREW INSTRUMENTS TO RETRIEVE THE PIECE. THIS PART TOOK APPROXIMATELY 20 MIN. THERE WAS NO PATIENT CONSEQUENCES. THERE WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED WHICH IS SCREW REMOVAL PROCESS, LARGER INCISION, MORE BONE REMOVAL, INCREASED TIME IN SURGERY, INCREASED TOURNIQUET TIME. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: Foreign Body In Patient

    MAUDE report 16796439 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page