Skip to content

Cannulated Compression Headless Screw 04.333.642Recalls & reported problems

Tyber Medical, LLC · 5.5MM TI CANN COMPRESSION HEADLESS SCREW- ST- 42MM · Class 2 · Implant

All models of Cannulated Compression Headless Screw (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00819917025170), received between Oct 18, 2024 and Oct 18, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Implant Pain1
  • Nonunion/Delayed-union Bone Fracture1
  • Radiolucent Lines1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 18, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PER CLINICAL STUDY REPORT: AT 4-MONTHS POST-OPERATIVE THE PATIENT WAS NOT FUSED AND WAS DEEMED TO HAVE A PAINFUL NONUNION WITH THE SCREWS LOOSE AND RADIOLUCENCY AROUND THE SCREWS. AT 5-MONTHS POST-OPERATIVE THE SUBJECT HAD REVISION SURGERY TO REMOVE THE CCHS SCREWS: AN ANKLE ARTHRODESIS WITH AN ANTERIOR APPROACH WAS PERFORMED USING AN EFFUSION PLATE AND BONE GRAFT. TO REMOVE THE CCHS SCREWS: AN ANKLE ARTHRODESIS WITH AN ANTERIOR APPROACH WAS PERFORMED USING AN EFFUSION PLATE AND BONE GRAFT. SUBJECT WAS TERMINATED FROM THE STUDY UPON REMOVAL OF HEADLESS SCREWS.

    Patient problems: Implant Pain; Nonunion/Delayed-union Bone Fracture; Radiolucent Lines

    MAUDE report 20485618 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page