Cannulated Compression Headless Screw 04.333.765Recalls & reported problems
Tyber Medical, LLC · 6.5MM TI CANN COMPRESSION HEADLESS SCREW- ST- 65MM · Class 2 · Implant
All models of Cannulated Compression Headless Screw (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00819917025675), received between Oct 18, 2024 and Oct 21, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Nonunion/Delayed-union Bone Fracture2
- Implant Pain1
- Radiolucent Lines1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
ADVERSE EVENT FROM THE CLINICAL STUDY: AT 9-MONTHS POST-OPERATIVE IMAGING DEMONSTRATED SCREW LOOSE, RADIOLUCENCY; THE SUBJECT WAS ORDERED A BONE GROWTH STIMULATOR AND TOE-OFF BRACE. AT 11-MONTHS POST-OPERATIVE SUBJECT CONTINUED TO HAVE PAIN WITH AMBULATION AND PLANNED TO CONTINUE TO USE BONE GROWTH STIMULATOR AND SIMULTANEOUSLY SCHEDULE SURGERY TO HAVE HARDWARE REMOVED IF NO IMPROVEMENT IS SEEN WITH BONE GROWTH STIMULATOR; IMAGING DEMONSTRATED THAT THE RADIOLUCENCY REMAINED VISIBLE HOWEVER NO NEW SIGNS OF LOOSENING. AT 12-MONTHS POST-OPERATIVE SUBJECT CONTINUED TO USE BONE GROWTH STIMULATOR WITH NO SIGNS OF INCREASED HEALING; IMAGING SHOWS RADIOLUCENCY REMAINED PRESENT. AT 13.5-MONTHS POST-OPERATIVE SUBJECT UNDERWENT HARDWARE REMOVAL SURGERY DUE TO THE NONUNION. ALL HARDWARE WAS REMOVED DURING THE PROCEDURE AND WAS REPLACED WITH A DIFFERENT BRAND'S SYSTEM. SUBJECT WAS TERMINATED FROM THE STUDY UPON REMOVAL OF HEADLESS SCREWS.
Patient problems: Implant Pain; Nonunion/Delayed-union Bone Fracture; Radiolucent Lines
MAUDE report 20500212 ↗InjuryOct 18, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
FROM CLINICAL INVESTIGATION: AT 19-MONTHS POST-OPERATIVE THE SUBJECT SLIPPED IN THE SHOWER AND PRESENTED TO THE CLINIC WITH A PROBABLE ANTERIOR TALOFIBULAR LIGAMENT SPRAIN WITH QUESTIONABLE SYMPTOMATIC NONUNION OF THE SUBTALAR JOINT. THE SUBJECT WAS PLACED BACK IN A FRACTURE BOOT DURING THE VISIT. AT 23.5 MONTHS POST-OPERATIVE A CT SCAN WAS OBTAINED WHERE NONUNION WAS CONFIRMED. THE SUBJECT HAD THE DEVICES REMOVED 23.5 MONTHS POST-OPERATIVE AND REPLACED WITH OTHER HARDWARE. SUBJECT WAS TERMINATED FROM THE STUDY AFTER HARDWARE REMOVAL.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 20486422 ↗
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page