USA Elite EIS-HCF25Recalls & reported problems
Gyrus ACMI, LLC · USA Elite System Continuous Flow Inner Sheath 25.6 Fr, (8.5 mm) Color Coded Yellow · Class 2
Quick summary
This USA Elite hysteroscope sheath is the inner sheath for a continuous-flow hysteroscope system. The 25 reports mostly describe the tip breaking, cracking or going missing, often noticed during cleaning between uses, with no deaths and no injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00821925001893), received between Jul 10, 2020 and Mar 23, 2026 — classified as 25 malfunctions.
The most frequently reported specific device problem is “Break” (18 reports).
7 other recalls cover devices cleared under the same FDA submission (K890328); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break18
- Crack6
- Fracture2
- Mechanical Problem2
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2026Crack
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROSCOPE EXHIBITED THE CRACK BEAK. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24659931 ↗MalfunctionMar 26, 2025Break; Break
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD BROKEN TIP. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21701605 ↗MalfunctionJan 15, 2025Break; Break
IT WAS REPORTED THE SUBJECT DEVICE HAD A BROKEN TIP. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21162374 ↗MalfunctionNov 6, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE CF RESECTOSCOPE INNER SHEATH, 25FR HAS A BROKEN, MISSING TIP. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18081341 ↗MalfunctionOct 11, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CF RESECTOSCOPE INNER SHEATH, 25FR HAD A DAMAGED TIP. THE EVENT OCCURRED DURING REPROCESSING FOR A THERAPEUTIC PROCEDURE. THERE WAS NO DELAY, AND THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17911583 ↗MalfunctionSep 4, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE CF RESECTOSCOPE INNER SHEATH, HAD A BROKEN TIP. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17680574 ↗MalfunctionAug 22, 2023Break; Break
THE CUSTOMER REPORTED THAT WHILE REPROCESSING HIS OLYMPUS CF RESECTOSCOPE INNER SHEATH, THE TIP BROKE OFF. THIS EVENT HAD NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17596449 ↗MalfunctionJul 13, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS, THE TIP OF THE SHEATH WAS DAMAGED. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17322556 ↗MalfunctionMay 30, 2023Break; Mechanical Problem; Break; Mechanical Problem
DAMAGED TIP WAS REPORTED ON THE DEVICE . THE ISSUE FOUND DURING REPROCESSING. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED. DEVICE RETURN EVALUATION FOUND BROKEN TIP , SHARP. THIS REPORT IS BEING SUBMITTED DUE TO THE FINDING OF BROKEN TIP , SHARP IDENTIFIED DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17021076 ↗MalfunctionMay 24, 2023Break; Break
THE CUSTOMER RETURNED THE RESECTOSCOPE INNER SHEATH, TO OLYMPUS REPAIR CENTER FOR EVALUATION OF A BROKEN CERAMIC TIP THAT WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16995214 ↗MalfunctionMay 17, 2023Crack; Crack
IT WAS REPORTED THE DEVICE HAD A BROKEN TIP. THE ISSUE FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT ON THIS REPORTED ISSUE. NO HARM, NO USER INJURY REPORTED. DEVICE EVALUATION FOUND CRACKED BEAK (TIP). THIS REPORT IS BEING SUBMITTED DUE TO THE FINDING OF CRACKED BEAK (TIP) DURING DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16950877 ↗MalfunctionMar 31, 2023Break; Crack; Break; Crack
CUSTOMER REPRESENTATIVE REPORTED THE DEVICE CERAMIC TIP WAS BROKEN. THE ISSUE FOUND DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT ON THIS REPORTED EVENT. NO HARM WAS REPORTED. NO USER INJURY REPORTED .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16654143 ↗MalfunctionFeb 15, 2023Fracture; Fracture
THE CUSTOMER REPORTED THE CERAMIC TIP OF THE SUBJECT DEVICE BROKE DURING AN UNSPECIFIED PROCEDURE. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16380420 ↗MalfunctionFeb 15, 2023Fracture; Fracture
THE CUSTOMER REPORTED THE CERAMIC TIP OF THE SUBJECT DEVICE BROKE DURING AN UNSPECIFIED PROCEDURE. ADDITIONAL DETAILS HAVE BEEN REQUESTED REGARDING THE REPORTED EVENT. AT THIS TIME, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16380407 ↗MalfunctionFeb 15, 2023Crack; Crack
THE CUSTOMER REPORTED TO OLYMPUS THAT THERE WAS A DAMAGED TIP ON THE CF RESECTOSCOPE INNER SHEATH (25FR). THE ISSUE OCCURRED DURING REPROCESSING, AND THERE WAS NO PATIENT HARM OR PROCEDURAL IMPACT ASSOCIATED WITH THE EVENT. THE DEVICE WAS RETURNED AND EVALUATED, AND THE BEAK WAS BROKEN WITH MISSING SEGMENTS, AND THE METAL SHEATH TUBE WAS CRACKED AND DENTED AT THE PROXIMAL END CONNECTING WITH THE LOCKING MECHANISM. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16378031 ↗MalfunctionFeb 8, 2023Break; Break
CUSTOMER REPORTED WITH AN ISSUE OF THE "INNER SHEATH DAMAGED TIP". THE ISSUE OCCURRED DURING REPROCESSING. ACCORDING TO THE REPORT, THE DAMAGE WAS NOTICED WHILE REPROCESSING THE ITEM AND WAS RECEIVED IN THE DEPARTMENT WITHOUT ANY PATIENT INFORMATION OR ISSUES WITH THE SURGERY. NO FURTHER INFORMATION WAS PROVIDED. NO HARM WAS REPORTED. NO PATIENT HARM, NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16319736 ↗MalfunctionJan 20, 2023Break; Break
OLYMPUS REPORTED ON BEHALF OF THE CUSTOMER, THE RESECTOSCOPE INNER SHEATH TIP BROKE DURING REPROCESSING. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16219605 ↗MalfunctionJan 12, 2023Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THERE WAS A BROKEN TIP WHILE USING THE CF RESECTOSCOPE INNER SHEATH, 25FR. THE ISSUE OCCURRED DURING REPROCESSING, AND THERE WAS NO PROCEDURAL IMPACT OR PATIENT HARM ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16146325 ↗MalfunctionJan 12, 2023Break; Mechanical Problem; Break; Mechanical Problem
COMPANY REPRESENTATIVE REPORTED WITH AN ISSUE OF " BROKEN TIP". THE ISSUE OCCURRED DURING REPROCESSING. THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS EVENT. NO HARM REPORTED, NO USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16144130 ↗MalfunctionJan 12, 2023Crack; Crack
COMPANY REPRESENTATIVE SENT A REPAIR REQUEST REPORTED WITH AN ISSUE OF "INNER SHEATH DAMAGED, BROKEN TIP". THE ISSUE OCCURRED DURING REPROCESSING. THERE IS NO PATIENT INVOLVEMENT ASSOCIATED ON THIS REPORTED EVENT. NO HARM WAS REPORTED, NO USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16143933 ↗MalfunctionSep 12, 2022Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE TIP IS CHIPPING. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15406733 ↗MalfunctionAug 3, 2022Break; Break
THE CUSTOMER REPORTED TO OLYMPUS THAT THE CERAMIC TIP BROKE IN TO MULTIPLE PIECES. THE REPORTED PROBLEM WAS FOUND DURING AN UNSPECIFIED PROCEDURE. THERE WAS NO HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15160268 ↗MalfunctionMar 15, 2022Break; Break
THE CUSTOMER REPORTED THE INNER SHEATH TIP BROKE ON THEIR CF RESECTOSCOPE INNER SHEATH, 25FR. ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION FROM THE CUSTOMER ARE IN PROGRESS. THERE WERE NO REPORTS OF DEATH, HARM OR INJURIES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13773090 ↗MalfunctionDec 10, 2021Crack; Component Missing; Crack; Component Missing
CUSTOMER UPDATES ON THE EVENT REPORTED AS BELOW: THE MISSING CERAMIC WAS LOCATED IN THE CYSTO DRAPES AT THE END OF THE PROCEDURE. THE SCRUB TECHNICIAN NOTICED IT WAS MISSING AT THE END OF THE CASE. SCRUB TECH FOUND THE MISSING TIP IN THE CYSTO DRAPES IN TWO SEPARATE PIECES AFTER IT WAS NOTICED THE TIP WAS MISSING FROM RESECTOSCOPE. NO HARM , NO PATIENT IMPACT. BROKEN TIP WAS NOT INSIDE THE PATIENT. THE DOCTOR WAS NOTIFIED TIP FOUND IN DRAPES. THE PATIENT WAS CHECKED (RE-SCOPE) TO ENSURE NO PIECES WERE LEFT INSIDE THE BLADDER. NOTHING FOUND INSIDE THE BLADDER. THE INTENDED PROCEDURE WAS COMPLETED. PROCEDURE PERFORMED WAS TUR-BUTTON TURP (TRANSURETHRAL RESECTION OF THE PROSTATE).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12981234 ↗MalfunctionJul 10, 2020Break; Break
THE SERVICE CENTER WAS INFORMED THAT AT THE CONCLUSION OF A THERAPEUTIC PROCEDURE, THE BLACK DISTAL TIP OF THE INNER SHEATH CRACKED INTO TWO PIECES AND FELL INSIDE THE PATIENT. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT¿S BLADDER. THE INTENDED PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10257393 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K890328. One submission often covers many models, so check the product and lot details in each record.
- Z-3300-2026 · Aug 24, 2026 · Open, ClassifiedOlympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
- Z-3299-2026 · Aug 24, 2026 · Open, ClassifiedOlympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
- Z-2720-2026 · May 5, 2026 · Open, ClassifiedOlympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
- Z-2139-2025 · Jun 18, 2025 · Open, ClassifiedBlue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some
- Z-2140-2025 · Jun 18, 2025 · Open, ClassifiedGreen Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the work
- Z-0568-2024 · Oct 2, 2023 · Open, ClassifiedACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF M
- Z-1952-2018 · Apr 17, 2018 · TerminatedGyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs t
Other Hysteroscope (And Accessories) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.