Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00821925001992), received between May 4, 2022 and Mar 24, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

7 other recalls cover devices cleared under the same FDA submission (K890328); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Crack1
  • Dent in Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE HYSTEROSCOPE HAD A INSTRUMENT TIP IS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24681606 ↗
  • MalfunctionDec 26, 2025Break; Break

    IT WAS REPORTED THAT THE HYSTEROSCOPE HAD A BROKEN CERAMIC TIP. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23899000 ↗
  • MalfunctionMar 20, 2025Break; Break

    IT WAS REPORTED THAT THE PLASTIC TOP OF THE SUBJECT DEVICE WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21657343 ↗
  • MalfunctionNov 13, 2024Dent in Material; Dent in Material

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RESECTOSCOPE EXHIBITED DENTED PLASTIC DISTAL END COVER. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677561 ↗
  • MalfunctionJul 24, 2024Break; Break

    IT WAS REPORTED, THE SHEATH'S CERAMIC TIP WAS BROKE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822147 ↗
  • MalfunctionDec 15, 2023Crack; Crack

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE NON-CF RESECTOSCOPE METAL BROWN BEAK SHEATH HAD THE CERAMIC AT THE END CRACK. THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18334126 ↗
  • MalfunctionAug 3, 2022Break; Break

    CUSTOMER REPORTED WITH AN ISSUE OF " BROKEN WHITE END". THE ISSUE FOUND DURING REPROCESSING. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15161715 ↗
  • MalfunctionJul 27, 2022Break; Break

    A USER FACILITY REPORTED THAT THE WHITE CHIP AT THE END OF THE METAL SHEATH OF A SCOPE BROKE OFF. IT IS UNKNOWN WHEN THE EVENT OCCURRED. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT REPORTED TO OLYMPUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15112716 ↗
  • MalfunctionMay 4, 2022Break; Break

    THE DEVICE WAS RETURNED FOR EVALUATION AND REPAIR FOR THE BREAKING OF THE WHITE TIP THAT OCCURRED DURING INSTALLATION. THERE IS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14277656 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K890328. One submission often covers many models, so check the product and lot details in each record.

  • Z-3299-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
  • Z-3300-2026 · Aug 24, 2026 · Open, Classified
    Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.
  • Z-2720-2026 · May 5, 2026 · Open, Classified
    Olympus Ellik Evacuator Includes Glass Body, Male Connecting Tubing & Latex Bulb. Model Number: 194.
  • Z-2140-2025 · Jun 18, 2025 · Open, Classified
    Green Silicone Seal for Accessories, Reusable Silicone Seal 0 - 7 FR. Model Number: CS-G7. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the work
  • Z-2139-2025 · Jun 18, 2025 · Open, Classified
    Blue Silicone, Reusable Silicone Seal 6 - 12 FR. Model Number: CS-B612. The reusable silicone seals fit over the biopsy port of endoscopes to seal off the working channel. Some
  • Z-0568-2024 · Oct 2, 2023 · Open, Classified
    ACMI Roller Bar Electrodes bearing incorrect pouch label of ACMI USA Elite System and USA Series Right Angle Cutting Loop Electrode Stabilized, 0.012 Wire 24 Fr (8mm), Model: REF M
  • Z-1952-2018 · Apr 17, 2018 · Terminated
    Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs t

Other Hysteroscope (And Accessories) devices

All Hysteroscope (And Accessories) devices →