Skip to content

Olympus BD-400P-1080Recalls & reported problems

Gyrus ACMI, LLC · EZDilate - Fixed Wire Esophageal 3-Stage Balloon Dilatation Catheter 8.5-10.5 mm · Class 2

All models of Olympus (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00821925033177), received between Jul 1, 2021 and Jan 23, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 10, 2023Open, Classified

    Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.

    Z-0934-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Rupture1
  • Material Split, Cut or Torn1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2024

    DILATION BALLOON RUPTURED WHILE IN PATIENT. BALLOON WAS BEING INFLATED PAST 7 ATMOSPHERES FOR A 9.5MM DIAMETER WHEN IT HAPPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18562919 ↗
  • MalfunctionJul 25, 2023Mechanical Problem; Material Split, Cut or Torn; Mechanical Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT WHILE HOLDING PRESSURE AT 9.5 ATM (ATMOSPHERIC PRESSURE), THE DEVICE WAS SUDDENLY LOST PRESSURE . UPON INSPECTION, IT WAS FOUND THE BALLOON HAD SPLIT DOWN THE SIDE. THE DEVICE WAS REPLACED AND THE INTENDED DIAGNOSTIC AN UNSPECIFIED PROCEDURE WAS COMPLETED USING ANOTHER DILATOR BALLOON (SIMILAR DEVICE). THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17387122 ↗
  • MalfunctionJul 1, 2021Material Rupture; Material Rupture

    AS REPORTED FOR THIS EVENT BY THE CUSTOMER, DURING AN UNKNOWN THERAPEUTIC PROCEDURE, THE BALLOON WAS INFLATED AND DEFLATED PRIOR TO THE BALLOON RUPTURING AFTER INFLATION. ALL THE BALLOON PIECES WERE RECOVERED. THERE IS NO REPORTED HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12100430 ↗

Other Dilator, esophageal devices

All Dilator, esophageal devices →

Sources and updates