PeriView FLEX NA-403D-2021Recalls & reported problems

Gyrus ACMI, LLC · SINGLE USE ASPIRATION NEEDLE (4/PK) · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Gyrus ACMI single-use aspiration needle is used during bronchoscopy procedures to take samples from inside the lung. Most of the 54 reports describe mechanical trouble with the needle: bent or twisted parts, pieces breaking off, sticking or resistance during retraction, and missing components. A few mention patient effects such as a punctured plastic sheath, a foreign body left in the patient, and one each of pneumothorax, hypoxia, and pneumonia. Four reports involve injuries and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
51
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 51 adverse event reports that list this exact device (DI 00821925042513), received between Apr 27, 2021 and Jan 28, 2026 — classified as 3 injuries, 48 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 5 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent14
  • Physical Resistance/Sticking5
  • Retraction Problem5
  • Component Missing4
  • Mechanical Problem4
  • Break3
  • Deformation Due to Compressive Stress3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions49
  • Pneumothorax2
  • Hypoxia1
  • Pneumonia1

Recent adverse event reports

The 25 most recent of 51 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Mechanical Jam

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24203565 ↗
  • MalfunctionJan 5, 2026Material Twisted/Bent

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23965832 ↗
  • MalfunctionDec 22, 2025Material Twisted/Bent; Material Twisted/Bent

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23865931 ↗
  • MalfunctionDec 1, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE HAD A PLASTIC SHEATH THAT WAS PUNCTURED. THE ISSUE OCCURRED DURING A DIAGNOSTIC TRANSBRONCHIAL NEEDLE ASPIRATION PROCEDURE. THERE WAS LESS THAN A THIRTY MINUTE DELAY AND THE PROCEDURE WAS COMPLETED WITH A COMPETITORS' DEVICE. THERE WAS NO HARM TO THE PATIENT ASSOCIATED WITH THIS EVENT. THE PLASTIC SHEATH OF THE NEEDLE WAS PUNCTURED. THERE WAS NO HARM TO THE PATIENT. PROCEDURE COMPLETED WITH FB-433D BIOPSY FORCEPS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23688513 ↗
  • MalfunctionJul 15, 2025Device Damaged Prior to Receipt by User; Device Damaged Prior to Receipt by User

    IT WAS REPORTED THE SUBJECT DEVICE HAD A STYLET THAT WAS BENT DURING A DIAGNOSTIC PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22513704 ↗
  • MalfunctionJul 14, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THE SUBJECT DEVICE HAD AN INCOMPLETE RETRACTION ISSUE DURING A DIAGNOSTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22501033 ↗
  • MalfunctionJun 13, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THE SUBJECT DEVICE NEEDLE WAS BENT AT THE TIP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22203174 ↗
  • MalfunctionApr 25, 2025Material Twisted/Bent; Material Twisted/Bent

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21910477 ↗
  • InjurySep 4, 2024Firing Problem; Firing Problem

    IT WAS REPORTED THAT DURING TRANSBRONCHIAL LUNG BIOPSY PROCEDURE, NO SAMPLE COULD BE OBTAINED WITH A BIOPSY OR CYTOLOGY BRUSH. CONSEQUENTLY, THE CUSTOMER OPTED TO USE THIS ASPIRATION NEEDLE; HOWEVER, THE NEEDLE DID NOT COME OUT AND THE TEST ENDED. ADDITIONALLY, AS THE PATIENT'S CONDITION "NOT BEING GOOD AND IT WAS TAKING TIME," THE PROCEDURE WAS DISCONTINUED. AS REPORTED, NO ADDITIONAL BIOPSIES WILL BE PERFORMED, AND A SCHEDULED OPERATION ON AUGUST 30TH WAS NOT PERFORMED, AS A RESULT. THERE WERE NO REPORTS OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20140202 ↗
  • MalfunctionAug 12, 2024Device Stenosis; Device Stenosis

    IT WAS REPORTED, THE SUBJECT DEVICE WAS UNABLE TO ADVANCE THE NEEDLE IN THE GUIDE SHEATH, AS IF THE GUIDE SHEATH CHANNEL NARROWS. THE ISSUE WAS FOUND PRIOR TO THE DEVICE ENTERING THE SCOPE/PATIENT DURING A THERAPEUTIC RADIAL ENDOBRONCHIAL ULTRASOUND (EBUS) PROCEDURE AND WAS COMPLETED USING A SIMILAR DEVICE. NO SURGICAL DELAY AND NO PATIENT HARM WAS REPORTED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19959980 ↗
  • MalfunctionAug 9, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED, DURING A DIAGNOSTIC RADIAL PROBE PROCEDURE THE SUBJECT DEVICE A HAD PLASTIC SHEATH THAT WAS TIGHT ON THE NEEDLE AND WHEN PUSHING THE NEEDLE, THE PLASTIC SHEATH WAS EXTENDING. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19951429 ↗
  • MalfunctionAug 9, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    ADDITIONAL INFORMATION WAS PROVIDED BY THE CUSTOMER: THE ISSUE OCCURRED DURING A DIAGNOSTIC RADIAL PROBE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19951343 ↗
  • MalfunctionAug 9, 2024Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED, DURING A DIAGNOSTIC RADIAL PROBE PROCEDURE THE SUBJECT DEVICE A HAD PLASTIC SHEATH THAT WAS TIGHT ON THE NEEDLE AND WHEN PUSHING THE NEEDLE, THE PLASTIC SHEATH WAS EXTENDING. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19950816 ↗
  • MalfunctionAug 5, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING A THERAPEUTIC RADIAL EBUS PROCEDURE, THE ASPIRATION NEEDLE WAS UNABLE TO ADVANCE IN THE GUIDE SHEATH. IT SEEMS AS IF THE GUIDE SHEATH CHANNEL NARROWS, NOT ALLOWING THE DEVICE TO PASS. THE INTENDED PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19910506 ↗
  • MalfunctionJul 15, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED FOR SINGLE USE ASPIRATION NEEDLE, NEEDLE WAS HAD A BENT SHEATH AND WAS DAMAGED WHEN PACKAGE WAS FIRST OPENED DURING A NAVIGATION BRONCHOSCOPY PROCEDURE. NEEDLE WAS NOT USED ON PATIENT AND THROWN AWAY. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19746364 ↗
  • MalfunctionJun 17, 2024Break; Break

    IT WAS REPORTED THE NEEDLE WAS BROKEN WHEN THE PACKAGE WAS OPENED. THE EVENT WAS FOUND DURING PREPARATION FOR USE AND THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19548804 ↗
  • MalfunctionJun 14, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE SINGLE USE ASPIRATION NEEDLE DID NOT SLIDE WELL INSIDE THE GUIDE SHEATH. IT WAS VERY TIGHT, AND IT WAS HARD TO GET THE NEEDLE IN AND HARD TO GET IT OUT. WHEN THE RESPIRATION THERAPIST (RT) WANTED TO TAKE THE NEEDLE OUT IN ORDER TO COLLECT THE SAMPLE, IT DID NOT COME OUT BECAUSE THE GUIDE SHEATH ITSELF HAD STRETCHED DURING INTRODUCTION AND REMOVAL. THE ISSUE OCCURRED DURING A THERAPEUTIC RADIAL EBUS (ENDOBRONCHIAL ULTRASOUND) BRONCHOSCOPY PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19538746 ↗
  • MalfunctionMay 22, 2024Ejection Problem; Ejection Problem

    IT WAS REPORTED THAT DURING A RADIAL-EBUS THERAPEUTIC PROCEDURE, IT WAS DIFFICULT TO GET THE ASPIRATION NEEDLE IN AND STILL DIFFICULT TO GET IT OUT AND THE GUIDESHEATH OF THE NEEDLE HAD STRETCHED DURING INSERTION OR WITHDRAWAL. THE PROCEDURE WAS DELAYED AND COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19378770 ↗
  • MalfunctionApr 25, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE ASPIRATION NEEDLE COULD NOT BE INSERTED TO THE END. THE MALFUNCTION OCCURRED DURING A DIAGNOSTIC PROCEDURE FOR ASPIRATION BIOPSY WHEN USED IN COMBINATION WITH THE GUIDED SHEATH. THE PHYSICIAN USED A DIFFERENT ASPIRATION NEEDLE WHICH COULD BE INSERTED; HOWEVER, THE NEEDLE DID NOT COME OUT OF THE TIP. THE PROCEDURE WAS COMPLETED USING BIOPSY FORCEPS. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19187235 ↗
  • MalfunctionApr 9, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE SHEATH GUIDE OF THE SINGLE USE ASPIRATION NEEDLE WAS DRAWN WHEN RETRACTING THE NEEDLE FROM THE KIT, AND IT WAS IMPOSSIBLE TO REMOVE THE NEEDLE. THE ISSUE WAS FOUND DURING A DIAGNOSTIC RADIAL ENDOBRONCHIAL ULTRASOUND. THE PROCEDURE WAS PROLONGED FOR AN UNSPECIFIED AMOUNT OF TIME. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE AND THE CONDITION OF THE PATIENT WAS NOT IMPACTED BY THE FAILURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19071612 ↗
  • MalfunctionApr 9, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE SHEATH GUIDE OF THE SINGLE USE ASPIRATION NEEDLE WAS DRAWN WHEN RETRACTING THE NEEDLE FROM THE KIT, AND IT WAS IMPOSSIBLE TO REMOVE THE NEEDLE. THE ISSUE WAS FOUND DURING A RADIAL ENDOBRONCHIAL ULTRASOUND. THE PROCEDURE WAS PROLONGED FOR AN UNSPECIFIED AMOUNT OF TIME. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE AND THE CONDITION OF THE PATIENT WAS NOT IMPACTED BY THE FAILURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19071342 ↗
  • MalfunctionMar 25, 2024Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED THE SINGLE USE ASPIRATION NEEDLE PUNCTURED THE SCOPE SHEATH AND WAS DAMAGED DURING AN UNIDENTIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18973444 ↗
  • MalfunctionMar 7, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT WHEN RETRACTING THE SINGLE USE ASPIRATION NEEDLE FROM THE KIT, THE SHEATH GUIDE OF THE NEEDLE WAS DRAWN AND IT WAS IMPOSSIBLE TO REMOVE THE NEEDLE. THE ISSUE WAS FOUND DURING THE MIDDLE OF THE ENDOBRONCHIAL ULTRASOUND RADIAL. THE PROCEDURE WAS PROLONGED FOR AN UNSPECIFIED AMOUNT OF TIME BUT WAS COMPLETED WITH ANOTHER DEVICE AND THE CONDITION OF THE PATIENT WAS NOT IMPACTED BY THE FAILURE. NO MEDICAL INTERVENTION WAS REQUIRED AND THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18859600 ↗
  • InjuryDec 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED FROM THAT THERE WAS NO DEVICE MALFUNCTION. PER THE AUTHOR, IT IS POSSIBLE THAT AN OLYMPUS PRODUCT CONTRIBUTED; HOWEVER THE STUDY WAS NOT DESIGNED IN A WAY TO MORE DEFINITIVE.

    Patient problems: Pneumothorax

    MAUDE report 18361380 ↗
  • MalfunctionDec 5, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE DEVICE MALFUNCTION OCCURRED IN THE MIDDLE OF THE PROCEDURE AND THE PROCEDURE WAS EXTENDED FOR AN UNSPECIFIED AMOUNT OF TIME. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. ADDITIONALLY, PER THE CUSTOMER THERE WAS LESS LIKELY DIAGNOSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18265003 ↗

FDA clearance and classification

The FDA regulates product code KTI (bronchoscope accessory) under 21 CFR 874.4680 as a Class II device. 21 CFR 874.4680 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the PeriView FLEX NA-403D-2021

Has the PeriView FLEX NA-403D-2021 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00821925042513), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PeriView FLEX NA-403D-2021?

The FDA's MAUDE database lists 51 adverse event reports for this device, received between Apr 27, 2021 and Jan 28, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (14), “Physical Resistance/Sticking” (5) and “Retraction Problem” (5); the most common patient problems are “Pneumothorax” (2), “Hypoxia” (1) and “Pneumonia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PeriView FLEX NA-403D-2021?

None of the reports are classified as deaths, 3 as injuries and 48 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the PeriView FLEX NA-403D-2021, and is it still sold?

The FDA's device database lists Gyrus ACMI, LLC as the company responsible for it (brand PeriView FLEX, model NA-403D-2021). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.