CRE Pulmonary M00550350Recalls & reported problems
Boston Scientific Corporation · Pulmonary Balloon Dilatation Catheter · Class 2
All models of CRE Pulmonary (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 08714729456230), received between May 6, 2019 and May 27, 2026 — classified as 1 injury, 17 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel10
- Adverse Event Without Identified Device or Use Problem5
- Material Puncture/Hole5
- Leak/Splash3
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2026Material Puncture/Hole; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE PULMONARY ANATOMY (BRONCHI) DURING A BRONCHOSCOPY STENT PROCEDURE PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, A BALLOON PINHOLE WAS NOTICED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT MODEL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FULLY RECOVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25308398 ↗MalfunctionOct 15, 2025Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION (BSC) THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE TRACHEA DURING A BALLOON DILATATION PROCEDURE TO TREAT AIRWAY STENOSIS PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, A BALLOON DILATATION WAS PERFORMED TO A PATIENT BECAUSE OF AN AIRWAY STENOSIS. IT WAS REPORTED THAT THE BALLOON BURST IN THE TRACHEA WHEN IT WAS INFLATED AT 5 ATMOSPHERIC PRESSURE AND WITH THE HANDLE SPECIALLY USED BY BOSTON SCIENTIFIC (BSC). NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23280149 ↗MalfunctionAug 18, 2025Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE LEFT MAIN BRONCHUS DURING A BRONCHOSCOPIC BALLOON DILATION PROCEDURE TO TREAT BRONCHIAL STENOSIS PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, BRONCHOSCOPIC BALLOON DILATATION WAS PERFORMED 10 MINUTES AFTER ABLATION, AND THE BALLOON BURST WHEN IT WAS FILLED TO ITS MAXIMUM DIAMETER. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22835120 ↗MalfunctionAug 18, 2025Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE LEFT UPPER LOBE OF BRONCHUS DURING A BRONCHOSCOPIC BALLOON DILATATION PROCEDURE TO TREAT STENOSIS PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON SUDDENLY BURST DURING DILATION. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22834395 ↗MalfunctionDec 10, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE BRONCHUS DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE BALLOON LEAKED WATER AND COULD NOT HOLD PRESSURE. A VIDEO CLIP WAS PROVIDED SHOWING THE BALLOON LEAKED THROUGH A PINHOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20888803 ↗MalfunctionJun 7, 2024Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE PULMONARY TRACTS DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE BALLOON HAD A HOLE IN THE MIDDLE OF A SURGICAL PROCEDURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19484008 ↗MalfunctionNov 15, 2023Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE MAIN AIRWAY DURING AN AIRWAY STENOSIS AND RECANALIZATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE BALLOON BURST WHEN IT WAS INFLATED TO 13.5 MM. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18144275 ↗MalfunctionApr 25, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE TRACHEA DURING A TRACHEAL DILATION PROCEDURE PERFORMED ON MARCH 28, 2023. DURING THE PROCEDURE, THE BALLOON WAS NOTED TO BE LEAKING AND THE DILATION PRESSURE COULD NOT BE REACHED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16806848 ↗MalfunctionApr 17, 2023Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION (BSC) THAT A CRE PULMONARY BALLOON WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE SECOND USE OF THE BALLOON IN THE PROCEDURE, AS IT WAS BEING USED IN A NON-BSC TRACHEAL SILICON STENT TO RAISE IT BY ONE CENTIMETER, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY BALLOON. THERE WERE NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16751979 ↗MalfunctionMar 20, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED FOR A BRONCHIAL STENOSIS DURING A TRANSBRONCHOSCOPIC BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6), 2023. DURING THE PROCEDURE, THE DEVICE WAS PLACED AT THE STENOSIS AND WAS ABOUT TO BE INFLATED. HOWEVER, IT WAS FOUND THAT THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16574249 ↗MalfunctionOct 4, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE BRONCHI DURING A PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE BALLOON BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15540964 ↗MalfunctionOct 4, 2022Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE TRACHEA DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE WHEN THE BALLOON WAS PLACED IN THE STRICTURE, THE BALLOON BURST WHEN IT REACHED 13.5 MM. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15540895 ↗MalfunctionSep 15, 2022Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY BALLOON WAS USED IN THE MAIN AIRWAY DURING A BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTICED THAT THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15432348 ↗MalfunctionJan 26, 2022Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A PULMONARY CRE BALLOON WAS USED DURING AN UNKNOWN TYPE OF PROCEDURE PERFORMED ON AN UNKNOWN PROCEDURE DATE. DURING AN UNKNOWN TIME OF EVENT, THE DOCTOR REPORTED THAT THE BALLOON WAS PUNCTURED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13362102 ↗MalfunctionAug 12, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION (BSC) THAT A CRE PULMONARY DILATATION CATHETER WAS USED IN THE LEFT MAIN BRONCHUS DURING A TRANSTRACHEAL BALLOON DILATION PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON COULD NOT BE PRESSURIZED. REPORTEDLY, THE BALLOON WAS TAKEN OUT OF THE PATIENT AND TESTED AFTER RE-PRESSURIZATION, AND NOTICED THAT THE TIP OF THE BALLOON WAS LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12315818 ↗MalfunctionApr 22, 2020Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE MAIN TRACHEA DURING A DILATION OF MAIN AIRWAY PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WAS ATTEMPTED TO BE INFLATED; HOWEVER, IT WAS NOTICED THAT THE BALLOON HAD A HOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9986714 ↗MalfunctionMar 17, 2020Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED ABOVE THE CARINA AREA DURING A DILATION OF TRACHEA PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9841781 ↗InjuryMay 6, 2019Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE RIGHT MAIN STEM DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON (B)(6), 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, UPON INFLATION, THE BALLOON BURST AT THE RECOMMENDED PRESSURE FOR THE DILATION SIZE. REPORTEDLY, THE BALLOON WAS RETRIEVED WITH BIOPSY FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8583047 ↗
FDA clearance and classification
The FDA regulates product code KTI (bronchoscope accessory) under 21 CFR 874.4680 as a Class II device. 21 CFR 874.4680 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the CRE Pulmonary M00550350
Has the CRE Pulmonary M00550350 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729456230), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CRE Pulmonary M00550350?
The FDA's MAUDE database lists 18 adverse event reports for this device, received between May 6, 2019 and May 27, 2026. In those reports, the most common specific device problems are “Burst Container or Vessel” (10), “Material Puncture/Hole” (5) and “Leak/Splash” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CRE Pulmonary M00550350?
None of the reports are classified as deaths, 1 as an injury and 17 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the CRE Pulmonary M00550350, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand CRE Pulmonary, model M00550350). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BRONCHOSCOPE ACCESSORY devices
- ViziShot 2 FLEX NA-U403SX-4019Gyrus ACMI, LLC · Bronchoscope accessory1 recall348 reports
- PeriView FLEX NA-403D-2021Gyrus ACMI, LLC · Bronchoscope accessory51 reports
- INSPIRA AIR BC1040AZAcclarent, Inc. · BRONCHOSCOPE ACCESSORY1 recall0 reports
- CRE Pulmonary M00550340Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY26 reports
- CRE Pulmonary M00550330Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY19 reports
- CRE Pulmonary M00550310Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY19 reports
- CRE Pulmonary M00550320Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY13 reports
- CRE Pulmonary M00550300Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.