- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 08714729456209), received between Jul 7, 2017 and Mar 5, 2025 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel6
- Material Puncture/Hole4
- Leak/Splash2
- Material Integrity Problem2
- Component Missing1
- Deflation Problem1
- Human-Device Interface Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient3
- Insufficient Information1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS BEING PREPARED TO BE USED IN THE TRACHEA DURING A BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6) 2025. DURING PREPARATION, WHEN THE PACKAGE WAS OPENED, THE FRONT END OF THE CATHETER WAS FOUND TO BE DAMAGED. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AFTER THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21528670 ↗MalfunctionJan 10, 2025Material Puncture/Hole; Component Missing; Human-Device Interface Problem; Material Puncture/Hole; Component Missing; Human-Device Interface Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE TRACHEA DURING A BALLOON CATHETER DILATATION PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, WATER LEAK WAS NOTICED DURING BALLOON DILATION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NOTE: VIDEO OF THE COMPLAINT DEVICE OUTSIDE THE PATIENT WERE PROVIDED BY THE CUSTOMER AND SHOWED THE WATER LEAKED THRU A PINHOLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21131342 ↗MalfunctionJan 8, 2025Burst Container or Vessel; Material Integrity Problem; Burst Container or Vessel; Material Integrity Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE BRONCHIAL TREE DURING A DILATION PROCEDURE. THE EXACT PROCEDURE DATE IS UNKNOWN. DURING THE PROCEDURE, THE PHYSICIAN TRIED TO INFLATE THE CRE BALLOON BUT IT WAS BROKEN. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21102818 ↗MalfunctionFeb 21, 2024Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHILE THE CLINICAL STAFF WERE INFLATING THE BALLOON WITH WATER, THE TUBE CONNECTING TO THE TIP OF THE BALLOON BURST. THE SURGEON USED A LARYNGOSCOPE TO CHECK IF THERE WERE ANY RESIDUES LEFT FROM THE BALLOON IN THE PATIENT'S THROAT AND CHECKED FOR ANY DAMAGE TO THE THROAT OR LUNGS. NONE WERE NOTED. THE PROCEDURE WAS REPORTED TO BE RESCHEDULED DUE TO THIS EVENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE EVENT. NOTE: BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18754896 ↗MalfunctionDec 14, 2023Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED DURING A PROCEDURE PERFORMED ON NOVEMBER 22, 2023. DURING THE PROCEDURE, AFTER THE CATHETER WAS INSERTED AND THE BALLOON WAS INFLATED, THE BALLOON WAS LEAKING FLUID AND STARTED TO DEFLATE. UPON CLOSER INSPECTION, IT APPEARED TO HAVE SEVERAL HOLES IN IT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18332630 ↗MalfunctionOct 17, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2023. IT WAS REPORTED THAT WHEN THE CATHETER WAS INSERTED AND THE BALLOON INFLATED, THE BALLOON STARTED TO LEAK AND DEFLATE DURING THE FIRST MINUTE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17953046 ↗MalfunctionOct 16, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED DURING A PULMONARY DILATION PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE CATHETER WAS INSERTED AND THE BALLOON INFLATED, THE BALLOON STARTED TO LEAK AND DEFLATE DURING THE FIRST MINUTE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17946399 ↗MalfunctionJan 27, 2022Burst Container or Vessel; Use of Device Problem; Burst Container or Vessel; Use of Device Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED ON A BALLOON DILATION OF THE GASTROINTESTINAL TRACT PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, THE BALLOON STOPPED INFLATING AND CONTRAST LEAKED OUT. THE BALLOON WAS NOTED TO HAVE RUPTURED AT 4 ATM. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13366882 ↗MalfunctionJul 16, 2021Burst Container or Vessel; Material Puncture/Hole; Burst Container or Vessel; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE AIRWAY DURING A DIAGNOSTIC BRONCHOSCOPY PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON RUPTURED AROUND 2 ATM. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PULMONARY BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12181558 ↗MalfunctionMar 8, 2021Material Puncture/Hole; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY BALLOON WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. IT WAS REPORTED THAT THE BALLOON HAD A HOLE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.
Patient problems: Insufficient Information
MAUDE report 11436332 ↗MalfunctionOct 10, 2018Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE LUNGS DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WAS BROKEN. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7953027 ↗MalfunctionOct 10, 2018Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE LUNGS DURING A BRONCHOSCOPY PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON WAS BROKEN. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7953026 ↗InjuryJul 7, 2017Deflation Problem; Deflation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PULMONARY DILATATION BALLOON WAS USED IN THE TRACHEA DURING A BALLOON DILATION PROCEDURE ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON COULD NOT BE DEFLATED AND HAD TO BE CUT IN HALF TO BE PULLED OUT MANUALLY IN THE PATIENT'S AIRWAY USING THE FORCEPS. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6694624 ↗
FDA clearance and classification
The FDA regulates product code KTI (bronchoscope accessory) under 21 CFR 874.4680 as a Class II device. 21 CFR 874.4680 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the CRE Pulmonary M00550320
Has the CRE Pulmonary M00550320 been recalled?
No FDA recall record in our data names this exact device identifier (DI 08714729456209), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CRE Pulmonary M00550320?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Jul 7, 2017 and Mar 5, 2025. In those reports, the most common specific device problems are “Burst Container or Vessel” (6), “Material Puncture/Hole” (4) and “Leak/Splash” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CRE Pulmonary M00550320?
None of the reports are classified as deaths, 1 as an injury and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the CRE Pulmonary M00550320, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand CRE Pulmonary, model M00550320). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BRONCHOSCOPE ACCESSORY devices
- ViziShot 2 FLEX NA-U403SX-4019Gyrus ACMI, LLC · Bronchoscope accessory1 recall348 reports
- PeriView FLEX NA-403D-2021Gyrus ACMI, LLC · Bronchoscope accessory51 reports
- INSPIRA AIR BC1040AZAcclarent, Inc. · BRONCHOSCOPE ACCESSORY1 recall0 reports
- CRE Pulmonary M00550340Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY26 reports
- CRE Pulmonary M00550330Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY19 reports
- CRE Pulmonary M00550310Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY19 reports
- CRE Pulmonary M00550350Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY18 reports
- CRE Pulmonary M00550300Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page