Skip to content

DRIVEMEDICAL FLNP600Recalls & reported problems

Drive Devilbiss Healthcare · Patient RoomLifts Product Description: Gravis Floor lift,manual base,600 lbs · Class 1

All models of DRIVEMEDICAL (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00822383990873), received between Nov 19, 2020 and Nov 19, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loose or Intermittent Connection1

Patient problems

  • Abrasion1
  • Fall1
  • Vertebral Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 19, 2020Loose or Intermittent Connection; Loose or Intermittent Connection

    (B)(6) IS THE INITIAL IMPORTER OF THE DEVICE WHICH IS A PATIENT LIFT. WHILE BEING TRANSFERRED TO BED BY THE DEVICE THE NUT AND BOLTS OF THE SPREADER BAR HAD LOOSENED AND THE PATIENT WAS DROPPED. THE LIFT THEN FELL ONTO THE RESIDENT. SHE HIT HER HEAD. SHE WAS TAKEN TO THE ER AND DIAGNOSED WITH A FRACTURED C6 VERTEBRA AND OTHER MINOR ABRASIONS. AFTERWARDS SHE RETURNED TO HOSPICE CARE WHERE SHE PASSED AWAY 3 WEEKS LATER WITH CAUSE OF DEATH AS SENILE DEGENERATION. WHEN THE DRIVER ARRIVED TO PICK UP THE LIFT AFTER THE INCIDENT WAS REPORTED, HE DID FIND THAT THE CRADLE HAD COME OFF AND THE BOLT AND NUT WERE ABSENT FROM THE BOOM WHERE IT CONNECTS. HOWEVER, UPON RETURNING THE LIFT TO THE WAREHOUSE, THE SAME BOLT AND NUT WERE RETURNED TO THE LIFT, TIGHTENED, TESTED AND FOUND TO BE IN PROPER WORKING CONDITION AND THE LIFT WAS PUT BACK INTO SERVICE. THE DEVICE WAS THEN ISOLATED WHEN THE SERVICE PROVIDER BECAME AWARE OF THE INCIDENT.

    Patient problems: Abrasion; Fall; Vertebral Fracture

    MAUDE report 10871431 ↗

Other LIFT, PATIENT, NON-AC-POWERED devices

All LIFT, PATIENT, NON-AC-POWERED devices →

Sources and updates