Micropuncture G07410Recalls & reported problems
Cook Incorporated · Micropuncture, Introducer Set With Stainless Steel Wire Guide · Class 2
All models of Micropuncture (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002074101), received between Aug 3, 2016 and Jul 24, 2017 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Failure to Advance1
- Kinked1
- Material Separation1
- Physical Resistance1
- Stretched1
- Unraveled Material1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Code Available1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2017Kinked; Stretched; Difficult to Advance
IT WAS REPORTED THAT AFTER PUNCTURE, THE WIRE GUIDE WAS DIFFICULT TO ADVANCE INTO THE NEEDLE. THEREFORE, IT WAS REMOVED FROM THE PATIENT'S BODY. IT WAS CONFIRMED THAT THE WIRE GUIDE WAS STRETCHED AND KINKED. AS A RESULT, ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE ARE NO KNOWN ADVERSE EFFECTS TO THE PATIENT REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6737295 ↗MalfunctionJul 18, 2017Material Separation; Physical Resistance
THE PHYSICIAN FELT RESISTANCE WHEN ATTEMPTING TO ADVANCE THE DEVICE OVER THE WIRE GUIDE. A VISUAL EXAMINATION OF THE DEVICE CONFIRMED THAT THE SHEATH SEPARATED FROM THE HUB. THE DEVICE WAS REPLACED WITH ANOTHER ONE.
Patient problems: No Code Available
MAUDE report 6721658 ↗MalfunctionJul 12, 2017Break
IT WAS REPORTED THE USER FELT RESISTANCE DURING ADVANCEMENT OF THE DEVICE. SUBSEQUENTLY, THE DEVICE BROKE AT THE CONNECTOR. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6707637 ↗MalfunctionAug 3, 2016Unraveled Material; Failure to Advance; Unraveled Material; Failure to Advance
IT WAS REPORTED THAT AFTER PUNCTURE, THE MICRO-PUNCTURE ACCESS SET WIRE GUIDE WAS INSERTED FROM THE SUBCLAVIAN VEIN. AS IT WOULD NOT ADVANCE, THE DEVICE WAS REMOVED AND FOUND TO BE KINKED AND UNRAVELED. ANOTHER DEVICE WITH THE SAME SPECIFICATIONS WAS USED TO COMPLETE THE TREATMENT. THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5843005 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page