Performer G08956Recalls & reported problems
Cook Incorporated · Check-Flo Performer Introducer · Class 2
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00827002089563), received between Nov 25, 2014 and Sep 12, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Leak/Splash2
- Break1
- Kinked1
- Material Split, Cut or Torn1
- Physical Property Issue1
- Unsealed Device Packaging1
Patient problems
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- Blood Loss1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 12, 2024Unsealed Device Packaging; Unsealed Device Packaging
AS REPORTED, A PERFORMER INTRODUCER WAS MISSING FROM ITS PACKAGING. PER THE REPORTER, A SEAL WAS NOT EVIDENT ON THE DISTAL END OF THE PACKAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20207242 ↗InjuryMay 17, 2023Break; Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Foreign Body In Patient
MAUDE report 16949734 ↗MalfunctionAug 20, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ADDITIONAL INFORMATION WAS RECEIVED 29SEP2020. THE PHYSICIAN REPORTED THAT THE SKIN AT THE ACCESS SITE WAS TOUGH AND SCARRED, POSSIBLY RESULTING FROM PREVIOUS PROCEDURES. CALCIFICATION AND TORTUOSITY WERE NOT REPORTED. THE SHEATH WAS ADVANCED OVER AN UNKNOWN LUNDERQUIST WIRE GUIDE, AND ANOTHER MANUFACTURER'S CATHETER AND WIRE GUIDE WERE ADVANCED THROUGH THE SHEATH, AS WELL AS AN UNKNOWN FLUSH CATHETER.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10431674 ↗MalfunctionMar 8, 2019Fluid/Blood Leak; Fluid/Blood Leak
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8404638 ↗MalfunctionJan 8, 2019Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED, THE VALVE ON THE PERFORMER INTRODUCER WAS BROKEN. BLOOD WAS LEAKING. BLOOD LOSS WAS NOT QUANTIFIED; HOWEVER, NO MEDICAL INTERVENTION WAS REQUIRED. THE PHYSICIAN CHANGED TO ANOTHER MANUFACTURER'S SHEATH TO COMPLETE THE ABDOMINAL ANEURYSM REPAIR PROCEDURE. THE COMPLAINT DEVICE WAS INTRODUCED INTO THE PATIENT'S COMMON FEMORAL ARTERY IN AN IPSELATERAL APPROACH. A .035 GLIDE WIRE WAS IN PLACE THROUGH THE COMPLAINT DEVICE. IT IS NOT KNOWN IF THE PATIENT'S ANATOMY WAS TORTUOUS OR CALCIFIED. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. ADDITIONALLY, IT WAS REPORTED AN UNKNOWN NUMBER OF SIMILAR ISSUES OCCURRED WITH PERFORMER INTRODUCERS IN THE PAST. PLEASE REFERENCE 1820334-2019-00082 WHICH CAPTURES THESE EVENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8228929 ↗MalfunctionFeb 19, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) IN THE AORTA, AFTER INTRODUCING ALL THE EVAR COMPONENTS, THE INTRODUCER WAS SWITCHED FOR A 12 FR PERFORMER INTRODUCER. THE PERFORMER INTRODUCER WAS OBSERVED TO BE HEAVILY LEAKING BLOOD FROM THE VALVE AND HEMOSTASIS WAS UNABLE TO BE MAINTAINED, EVEN WHEN A 0.035 INCH DIAMETER WIRE GUIDE AND CATHETERS (REPORTER SPECIFIED THAT ANOTHER MANUFACTURER'S 5 FR CATHETER WAS USED) WERE PASSED THROUGH THE VALVE. DUE TO THE LEAKAGE, THE PROCEDURE WAS REQUIRED TO BE PERFORMED QUICKLY, BUT WAS ABLE TO BE SUCCESSFULLY COMPLETED. NO INTERVENTION WAS REQUIRED. EVERYTHING WAS PLACED IN THE CENTER OF THE INTRODUCER AND THE MEMBRANE WAS NOT PUNCTURED. ACCESS WAS GAINED THROUGH THE RIGHT LEG VIA A CONTRALATERAL APPROACH. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. IT WAS NOTED THAT SIMILAR EVENTS WITH LEAKAGE HAVE OCCURRED PREVIOUSLY WITH OTHER 12 FR PERFORMER INTRODUCERS (REFER TO MEDWATCH 1820334-2018-00540) AND A 14 FR PERFORMER INTRODUCER (REFER TO MEDWATCH 1820334-2018-00394). PER THE INITIAL REPORTER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7280708 ↗MalfunctionAug 8, 2017Kinked; Physical Property Issue
THE COMPLAINANT REPORTED THAT THE SHEATH WAS INSERTED IN THE LEFT CEPHALIC VEIN AND IT KINKED. IT WAS FURTHER NOTED THAT THERE WERE NICKS ON THE SHEATH THAT HAVE POTENTIAL TO LEAD TO AN ADVERSE CONSEQUENCE. HOWEVER, ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6773199 ↗MalfunctionNov 25, 2014Leak/Splash
ON (B)(6) 2014, DURING AN INFRA-RENAL GRAFT IMPLANT PROCEDURE ON A MALE PT, WHEN THE DEVICE WAS INTRODUCED IN THE FEMORAL ARTERY IT WAS STREAMING FROM HUB AND SHEATH CONNECTION. IT SEEMED AS THOUGH IT WAS NOT SEALED. THE PHYSICIAN DID AN ANGIOGRAM AND REPLACED WITH A MAIN BODY GRAFT AND COMPLETED THE PROCEDURE SUCCESSFULLY. THE SHEATH WAS IN ONLY TEMPORARILY. NO HARM TO THE PT, AND NO ADDITIONAL PROCEDURES OR INTERVENTION WAS REQUIRED DUE TO THIS OCCURRENCE. THE PT LOST APPROXIMATELY 10CC'S OF BLOOD, NO TRANSFUSION REQUIRED.
Patient problems: Blood Loss
MAUDE report 4321814 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.