Skip to content

Cook G09502Recalls & reported problems

Cook Incorporated · Multipurpose Drainage Catheter Ultrathane · Class U · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This is a Cook Ultrathane Mac-Loc multipurpose drainage catheter, placed in the body to drain fluid. Many of the 83 reports describe the hub separating from the catheter shaft, fluid leaking around the locking mechanism, and materials separating or cracking, and a few describe longer procedures. There are 40 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
79
Deaths reported
0
Injuries reported
38
Recalls naming it
0

Summary

The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 00827002095021), received between Apr 6, 2016 and Sep 3, 2026 — classified as 38 injuries, 41 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak22
  • Detachment of Device or Device Component14
  • Leak/Splash14
  • Material Separation13
  • Break5
  • Crack4
  • Component Falling3
  • Obstruction of Flow3

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • No Code Available19
  • No Known Impact Or Consequence To Patient16
  • Insufficient Information5
  • No Consequences Or Impact To Patient4
  • Foreign Body In Patient2
  • No Patient Involvement2
  • Unspecified Mental, Emotional or Behavioral Problem2

Recent adverse event reports

The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 3, 2026Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50598537 ↗
  • MalfunctionAug 26, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IN ADDITIONAL INFORMATION RECEIVED ON 27AUG2025, IT WAS REPORTED THAT THE COMPLAINT DEVICE WAS IN PLACE FOR 5 MINUTES BEFORE THE HUB SEPARATED. THE COMPLAINT DEVICE WAS THEN REMOVED FROM THE PATIENT, AND A NEW CATHETER WAS PLACED TO COMPLETE THE PROCEDURE. THIS LED TO AN INCREASED PROCEDURE TIME, RESULTING IN THE ADMINISTRATION OF MORE CONTRAST AND MORE RADIATION EXPOSURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22895721 ↗
  • MalfunctionMay 7, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER LEAKED FOLLOWING A NEPHROSTOMY TUBE EXCHANGE FOR A PATIENT OF UNKNOWN AGE AND GENDER. UPON SECURING THE MAC-LOC, THE CATHETER WAS FOUND TO BE LEAKING URINE AROUND THE LOCKING MECHANISM. THE CATHETER WAS SUBSEQUENTLY CUT AND REMOVED, AND A NEW CATHETER WAS INSERTED, WHICH HAD NO ISSUES. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21977234 ↗
  • MalfunctionMar 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IN ADDITIONAL INFORMATION RECEIVED ON 24APR2025, IT WAS CONFIRMED THAT THE DEVICE WAS INITIALLY PLACED IN TO LEFT PLEURAL EFFUSION BY THE RADIOLOGY STAFF ON (B)(6) 2025. THE DEVICE SEPARATED AT THE HUB, BUT THE TIP OF THE DEVICE REMAINED INTACT, NO FRAGMENTS OF THE COMPLAINT DEVICE WERE LEFT IN THE PATIENT. THE PATIENT HAD LIMITED MOBILITY AND WAS ACUTELY ADMITTED DUE TO COVID AND PLEURAL INFUSION. ADDITIONALLY, THE PATIENT WAS SUFFERING FROM MYOCARDITIS AND WAS UNDERGOING TREATMENT FOR LEUKEMIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21527523 ↗
  • InjurySep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER LEAKED AFTER IT WAS PLACED IN A PATIENT OF UNKNOWN AGE DURING AN UNKNOWN PROCEDURE. THE DRAIN WAS FOUND TO BE LEAKING FROM THE FLOOR AND THE HUB WAS COMPLETELY DISCONNECTED FROM THE CATHETER. THE DEVICE WAS USED AS AN EXTERNAL BILIARY DRAIN. IT WAS NOTED THAT THE DRAIN HAD BEEN EXCHANGED "IN (B)(6)." A NEW DEVICE WAS PLACED IN AN ADDITIONAL PROCEDURE. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20292991 ↗
  • MalfunctionAug 15, 2024Material Separation; Obstruction of Flow; Material Separation; Obstruction of Flow

    IT WAS REPORTED THAT DURING AN EXCHANGE AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS OCCLUDED. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19991143 ↗
  • InjuryMay 24, 2024Break; Break

    IT WAS REPORTED THE THE SUTURE STRING OF AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS RETAINED IN THE PATIENT THE CATHETER WAS PLACED FOR USE AS A CHEST DRAIN TO TREAT LEFT EMPYEMA. LATER, DURING REMOVAL OF THE CATHETER, THE MAC-LOC WAS ABLE TO BE UNLOCKED WITH NO ISSUE BY THE REGISTRAR. HOWEVER, THE SUTURE WAS RETAINED IN THE PLEURAL CAVITY. GUIDANCE FROM INTERVENTIONAL RADIOLOGY (IR) REGARDING POTENTIAL REMOVAL HAS BEEN REQUESTED. CURRENTLY, THE SUTURE REMAINS IN THE PATIENT. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Foreign Body In Patient; Unspecified Mental, Emotional or Behavioral Problem

    MAUDE report 19393690 ↗
  • InjuryMay 7, 2024Break; Break

    IT WAS REPORTED THAT THE SUTURE STRING OF AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS RETAINED IN THE PATIENT THE CATHETER WAS PLACED FOR USE AS A CHEST DRAIN TO TREAT LEFT EMPYEMA. LATER, DURING REMOVAL OF THE CATHETER, THE SUTURE WAS RETAINED IN THE PLEURAL CAVITY. GUIDANCE FROM INTERVENTIONAL RADIOLOGY (IR) REGARDING POTENTIAL REMOVAL HAS BEEN REQUESTED. CURRENTLY, THE SUTURE REMAINS IN THE PATIENT. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Foreign Body In Patient; Unspecified Mental, Emotional or Behavioral Problem

    MAUDE report 19260586 ↗
  • MalfunctionJan 19, 2024Fluid/Blood Leak; Fluid/Blood Leak

    ADDITIONAL INFORMATION WAS RECEIVED INDICATING ANOTHER CATHETER WAS USED TO COMPLETE THE PROCEDURE. IT WAS CONFIRMED THAT THE PATIENT DID NO EXPERIENCE ANY NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18545841 ↗
  • InjuryDec 27, 2023Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18402228 ↗
  • InjuryJul 25, 2023Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390833 ↗
  • MalfunctionJun 13, 2023Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING PATIENT AND/OR EVENT DETAILS HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Fluid Discharge

    MAUDE report 17118162 ↗
  • MalfunctionMay 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16872855 ↗
  • MalfunctionApr 4, 2023Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED A FOREIGN SUBSTANCE WAS FOUND INSIDE THE POUCH OF AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER PRIOR TO PATIENT CONTACT. NO ADVERSE EFFECTS HAVE BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16676238 ↗
  • InjuryMar 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16656703 ↗
  • InjuryMar 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16656698 ↗
  • InjuryMar 31, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16656519 ↗
  • InjuryOct 7, 2022Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT AN UNKNOWN PATIENT REQUIRED PLACEMENT OF THE ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER. THE PHYSICIAN OBSERVED AN AIR LEAK IN THE LOCKED CATHETER AFTER PLACEMENT. THE CATHETER WAS REPLACED WITH A NEW LIKE DEVICE IN AN ADDITIONAL PROCEDURE. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15566592 ↗
  • MalfunctionOct 4, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER LEAKED. DURING THE PROCEDURE, AFTER PLACING THE CATHETER FOR BILIARY DRAINAGE, LEAKAGE OF THE CATHETER WAS NOTED. A VIDEO PROVIDED BY THE CUSTOMER SHOWS FLUID LEAKAGE. A NEW DEVICE WAS THEN OBTAINED TO SUCCESSFULLY COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15539823 ↗
  • InjurySep 29, 2022Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT A PATIENT WAS SCHEDULED FOR A UNILATERAL NEPHROSTOMY TUBE EXCHANGE IN THE RIGHT KIDNEY ON (B)(6) 2022. THE NEPHROSTOMY TUBE WAS FUNCTIONING NORMALLY WITHOUT ANY LEAK. A COOK ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS PLACED IN THE PATIENT DURING THE EXCHANGE. THE PATIENT WAS SCHEDULED TO RETURN IN THREE MONTHS. APPROXIMATELY THREE HOURS AFTER THE EXCHANGE, THE PATIENT CALLED THE PHYSICIAN'S OFFICE AND INDICATED THAT THE NEPHROSTOMY TUBE WAS LEAKING, CAUSING HIM TO CHANGE HIS CLOTHES NUMEROUS TIMES. THE PATIENT RETURNED TO THE DEPARTMENT AND A LEAK WITHIN THE HUB OF THE ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS DISCOVERED. THE ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER WAS REMOVED AND REPLACED WITH A "FRESH TUBE."

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Kidney or Urinary Problem

    MAUDE report 15505298 ↗
  • InjurySep 20, 2022Leak/Splash; Migration; Leak/Splash; Migration

    IT WAS REPORTED THAT AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER LEAKED. AFTER THE DEVICE WAS PLACED FOR ABSCESS DRAINAGE, LEAKAGE FROM THE CATHETER WAS NOTED. THE LOCKING LOOP SUTURE LOOSENED, WHICH POSSIBLY CAUSED THE DRAIN TO MOVE FORWARD IN THE PATIENT. AS A RESULT, THE PATIENT RETURNED TO THE HOSPITAL ON (B)(6) 2022 FOR AN ADDITIONAL PROCEDURE TO REMOVE AND REPLACE THE DEVICE. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT. ADDITIONAL INFORMATION REGARDING EVENT DETAILS HAS BEEN REQUESTED, BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15455274 ↗
  • MalfunctionJul 8, 2022Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT A COOK ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER ACCIDENTALLY DISLODGED 6 DAYS AFTER PLACEMENT. THE DEVICE WAS REQUIRED FOR A STUDY PROCEDURE (MDR-1919) AND WAS SUCCESSFULLY PLACED ON (B)(6) 2021 IN THE RIGHT PELVIS FOR ABSCESS DRAINAGE. COMPUTED TOMOGRAPHY (CT) VISUAL GUIDANCE WAS USED DURING THE PLACEMENT OF THE CATHETER. PRIOR TO THE PROCEDURE, DILATION WAS PERFORMED. ON (B)(6) 2021 IT WAS DISCOVERED THAT THE CATHETER ACCIDENTALLY DISLODGED SIX DAYS AFTER PLACEMENT. THE ABSCESS WAS NOT COMPLETELY DRAINED WHEN IT WAS DISLODGED. A CT SHOWED SMALL RESIDUAL COLLECTION. THE DEVICE WAS REMOVED ON (B)(6) 2021 AND WAS NOT REPLACED. NO ADVERSE EVENTS WERE REPORTED PRIOR TO THE PATIENT EXITING THE STUDY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14959500 ↗
  • InjuryJul 6, 2022Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT A PATIENT'S ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER HAD INADEQUATE DRAINAGE. THE DEVICE WAS REQUIRED FOR A STUDY PROCEDURE (MDR-1919) FOR ABSCESS DRAINAGE AND WAS PLACED ON (B)(6) 2021. DILATION WAS PERFORMED PRIOR TO CATHETER PLACEMENT, THEN THE DEVICE WAS SUCCESSFULLY PLACED IN THE LEFT ANTEROLATERAL ABDOMINAL WALL VIA COMPUTED TOMOGRAPHY (CT) GUIDANCE AND DRAINAGE WAS ESTABLISHED. ON (B)(6) 2021, IT WAS DISCOVERED THAT THE DEVICE WAS NOT PROVIDING ADEQUATE DRAINAGE. THE DEVICE WAS THEN REMOVED AND REPLACED WITH A LARGER SIZE. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.

    Patient problems: Failure of Implant

    MAUDE report 14939764 ↗
  • InjuryJun 10, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A PATIENT'S ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER DISLODGED. THE DEVICE WAS REQUIRED FOR A STUDY PROCEDURE (MDR-1919) FOR SUPRAPUBIC URINARY DRAINAGE AND WAS PLACED ON (B)(6) 2021. DILATION WAS PERFORMED PRIOR TO CATHETER PLACEMENT, THEN THE DEVICE WAS SUCCESSFULLY PLACED VIA ULTRASOUND GUIDANCE. DRAINAGE WAS SUCCESSFULLY ESTABLISHED. ON (B)(6) 2021, THE DEVICE WAS ACCIDENTALLY DISLODGED. THE NEXT DAY, A NEW SUPRAPUBIC CATHETER WAS SUCCESSFULLY PLACED. NO OTHER ADVERSE EVENTS WERE REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Urinary Retention

    MAUDE report 14655131 ↗
  • InjuryJun 2, 2022Obstruction of Flow; Patient Device Interaction Problem; Obstruction of Flow; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT EXPERIENCED INSUFFICIENT DRAINAGE AND PAIN RELATED TO AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER. THE DEVICE WAS REQUIRED FOR A STUDY PROCEDURE ((B)(4)) TO TREAT PLEURAL EFFUSION. DILATION WAS PERFORMED PRIOR TO CATHETER PLACEMENT, THEN THE DEVICE WAS SUCCESSFULLY PLACED VIA ULTRASOUND GUIDANCE. DRAINAGE WAS ESTABLISHED; HOWEVER, COMPLETE FLUID EVACUATION WAS NOT ACHIEVED AND THE PLEURAL EFFUSION INCREASED IN SIZE. THE PATIENT ALSO EXPERIENCED CATHETER-RELATED PAIN. AS A RESULT, THE DEVICE WAS REMOVED TWO DAYS AFTER PLACEMENT RELATED TO ONGOING PAIN AT CATHETER SITE. NO OTHER ADVERSE EVENTS WERE REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: Pain

    MAUDE report 14578688 ↗

FDA clearance and classification

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Cook G09502

Has the Cook G09502 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002095021), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Cook G09502?

The FDA's MAUDE database lists 79 adverse event reports for this device, received between Apr 6, 2016 and Sep 3, 2026. In those reports, the most common specific device problems are “Fluid/Blood Leak” (22), “Detachment of Device or Device Component” (14) and “Leak/Splash” (14); the most common patient problems are “Foreign Body In Patient” (2) and “Unspecified Mental, Emotional or Behavioral Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Cook G09502?

None of the reports are classified as deaths, 38 as injuries and 41 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Cook G09502 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Cook G09502, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cook, model G09502). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, NEPHROSTOMY devices

All CATHETER, NEPHROSTOMY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page