Skip to content

Cook G09767Recalls & reported problems

Cook Incorporated · Multipurpose Drainage Catheter Ultrathane · Class U · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002097674), received between Feb 5, 2018 and Sep 13, 2024 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Difficult to Remove” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2024Open, Classified

    Product labels state the incorrect expiration dates that extends beyond their true expiration date

    Z-2223-2024 · Root cause: Labeling Change Control · FDA record ↗

Most reported problems

Device problems

  • Difficult to Remove8
  • Break2
  • Detachment of Device or Device Component1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient3

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2024Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20218110 ↗
  • MalfunctionAug 3, 2023Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    THE DEVICE WAS RETURNED TO COOK. UPON RECEIPT OF THE DEVICE, IT WAS NOTED THAT THE FLEXIBLE STIFFENER WAS LODGED INSIDE THE CATHETER, AND THE HUB WAS MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17454735 ↗
  • MalfunctionMar 27, 2023Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED IN ADDITIONAL INFORMATION THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE USING NEW INVENTORY/FRESH DRAIN. CLARIFICATION WAS RECEIVED THAT THE RIGID STIFFENER WAS UTILIZED DURING THE DRAIN PLACEMENT PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16620616 ↗
  • MalfunctionMar 20, 2023Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT A PATIENT REQUIRED PLACEMENT OF AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER. DURING PLACEMENT OF THE CATHETER, THE STIFFENER WAS DIFFICULT TO REMOVE FROM THE CATHETER. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE. TWO OTHER INCIDENTS OF DIFFICULT STIFFENER REMOVAL FOR THIS PATIENT HAVE BEEN CAPTURED IN REPORTS WITH PATIENT IDENTIFIER: (B)(6).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16573855 ↗
  • MalfunctionMar 20, 2023Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16573456 ↗
  • MalfunctionFeb 7, 2023Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16312379 ↗
  • MalfunctionNov 12, 2019Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9312422 ↗
  • MalfunctionNov 12, 2019Difficult to Remove; Difficult to Remove

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9312280 ↗
  • MalfunctionFeb 5, 2018Break; Break

    THE CUSTOMER REPORTED THAT THE PLASTIC STIFFENER BECAME STUCK INSIDE OF THE ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER DURING AN ILEAL CONDUIT DRAIN PLACEMENT. THE PHYSICIAN BEGAN TO PULL ON THE STIFFENER, AND IT REPORTEDLY BROKE OFF INSIDE THE DRAINAGE CATHETER. THE DRAINAGE CATHETER WAS SUBSEQUENTLY REMOVED AND A NEW DEVICE WAS PLACED TO COMPLETE THE PROCEDURE, WITH NO FURTHER ADVERSE EVENTS RESULTING AND NO ADDITIONAL PROCEDURES BEING NECESSITATED. THE DEVICE WAS ONLY IN PLACE IN THE PATIENT'S URETER FOR A FEW MINUTES. THE COMPLAINT DEVICE HAS BEEN RETURNED FOR EVALUATION; HOWEVER, AS OF THE DATE OF THIS REPORT, THE INVESTIGATION IS STILL PENDING.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7243012 ↗

Other CATHETER, NEPHROSTOMY devices

All CATHETER, NEPHROSTOMY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.