Cook G09767Recalls & reported problems
Cook Incorporated · Multipurpose Drainage Catheter Ultrathane · Class U · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002097674), received between Feb 5, 2018 and Sep 13, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Difficult to Remove” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 15, 2024Class IIOpen, Classified
Product labels state the incorrect expiration dates that extends beyond their true expiration date
Z-2223-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Remove8
- Break2
- Detachment of Device or Device Component1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 13, 2024Difficult to Remove; Difficult to Remove
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20218110 ↗MalfunctionAug 3, 2023Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
THE DEVICE WAS RETURNED TO COOK. UPON RECEIPT OF THE DEVICE, IT WAS NOTED THAT THE FLEXIBLE STIFFENER WAS LODGED INSIDE THE CATHETER, AND THE HUB WAS MISSING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17454735 ↗MalfunctionMar 27, 2023Break; Difficult to Remove; Break; Difficult to Remove
IT WAS REPORTED IN ADDITIONAL INFORMATION THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE PHYSICIAN WAS ABLE TO COMPLETE THE PROCEDURE USING NEW INVENTORY/FRESH DRAIN. CLARIFICATION WAS RECEIVED THAT THE RIGID STIFFENER WAS UTILIZED DURING THE DRAIN PLACEMENT PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16620616 ↗MalfunctionMar 20, 2023Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT A PATIENT REQUIRED PLACEMENT OF AN ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER. DURING PLACEMENT OF THE CATHETER, THE STIFFENER WAS DIFFICULT TO REMOVE FROM THE CATHETER. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE. TWO OTHER INCIDENTS OF DIFFICULT STIFFENER REMOVAL FOR THIS PATIENT HAVE BEEN CAPTURED IN REPORTS WITH PATIENT IDENTIFIER: (B)(6).
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16573855 ↗MalfunctionMar 20, 2023Difficult to Remove; Difficult to Remove
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 16573456 ↗MalfunctionFeb 7, 2023Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16312379 ↗MalfunctionNov 12, 2019Difficult to Remove; Difficult to Remove
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9312422 ↗MalfunctionNov 12, 2019Difficult to Remove; Difficult to Remove
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9312280 ↗MalfunctionFeb 5, 2018Break; Break
THE CUSTOMER REPORTED THAT THE PLASTIC STIFFENER BECAME STUCK INSIDE OF THE ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER DURING AN ILEAL CONDUIT DRAIN PLACEMENT. THE PHYSICIAN BEGAN TO PULL ON THE STIFFENER, AND IT REPORTEDLY BROKE OFF INSIDE THE DRAINAGE CATHETER. THE DRAINAGE CATHETER WAS SUBSEQUENTLY REMOVED AND A NEW DEVICE WAS PLACED TO COMPLETE THE PROCEDURE, WITH NO FURTHER ADVERSE EVENTS RESULTING AND NO ADDITIONAL PROCEDURES BEING NECESSITATED. THE DEVICE WAS ONLY IN PLACE IN THE PATIENT'S URETER FOR A FEW MINUTES. THE COMPLAINT DEVICE HAS BEEN RETURNED FOR EVALUATION; HOWEVER, AS OF THE DATE OF THIS REPORT, THE INVESTIGATION IS STILL PENDING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7243012 ↗
FDA clearance and classification
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Cook G09767
Has the Cook G09767 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2223-2024 (Class II), began on May 15, 2024: Product labels state the incorrect expiration dates that extends beyond their true expiration date. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Cook G09767?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Feb 5, 2018 and Sep 13, 2024. In those reports, the most common specific device problems are “Difficult to Remove” (8), “Break” (2) and “Detachment of Device or Device Component” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Cook G09767?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Cook G09767 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the Cook G09767, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Cook, model G09767). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, NEPHROSTOMY devices
- Cook G09501Cook Incorporated · CATHETER, NEPHROSTOMY277 reports
- Dawson Mueller G09705Cook Incorporated · CATHETER, NEPHROSTOMY2 deaths214 reports
- Cook G09497Cook Incorporated · CATHETER, NEPHROSTOMY127 reports
- Dawson Mueller G11020Cook Incorporated · CATHETER, NEPHROSTOMY99 reports
- Cook G09502Cook Incorporated · CATHETER, NEPHROSTOMY79 reports
- NephroMax Kit M0062101180Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall25 reports
- NephroMax Kit M0062101600Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall11 reports
- Dawson Mueller G13768Cook Incorporated · CATHETER, NEPHROSTOMY55 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page