Skip to content

NephroMax Kit M0062101600Recalls & reported problems

Boston Scientific Corporation · High Pressure Nephrostomy Balloon Catheter Kit · Class U

All models of NephroMax Kit (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729834540), received between Jun 8, 2017 and Apr 2, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Open, Classified

    The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

    Z-2535-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel6
  • Material Rupture2
  • Material Split, Cut or Torn2
  • Break1
  • Fluid/Blood Leak1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient4
  • Discomfort1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026Burst Container or Vessel

    IT WAS REPORTED THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. DURING THE PROCEDURE, THE PHYSICIAN WAS TRYING TO ACCESS TO THE TRACK; HOWEVER, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24769102 ↗
  • MalfunctionOct 16, 2025Break; Fluid/Blood Leak; Material Split, Cut or Torn; Break; Fluid/Blood Leak; Material Split, Cut or Torn

    IT WAS REPORTED THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER DEVICE WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. DURING THE PROCEDURE, AFTER THE PATIENT UNDERWENT RIGHT PERCUTANEOUS NEPHROLITHOTOMY (PCNL) THE NEPHROSTOMY TUBE FELL APART WHERE THE TUBING MEETS THE LUER LOCK HUB. REPORTEDLY, IT WAS FOUND THAT THE TUBE NEAR THE HUB WAS BROKEN AND IT WAS NOTICING A LEAKAGE FROM THE TUBE. THIS REQUIRES REPLACING THE TUBE AND THE PATIENT FELT DISCOMFORT AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort

    MAUDE report 23318485 ↗
  • MalfunctionJul 7, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER DEVICE WAS USED DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. DURING THE PROCEDURE, THE BALLOON BURST WHILE INFLATING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22422691 ↗
  • MalfunctionSep 18, 2023Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER WAS USED INT HE RIGHT KIDNEY DURING A PERCUTANEOUS NEPHROLITHONOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, THE BALLOON BURST AS IT WAS BEING INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX NEPHROSTOMY BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17765745 ↗
  • MalfunctionSep 18, 2023Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER WAS USED INT HE RIGHT KIDNEY DURING A PERCUTANEOUS NEPHROLITHONOMY (PCNL) PROCEDURE PERFORMED ON (B)(6), 2023. DURING THE PROCEDURE, THE BALLOON BURST BEFORE IT WAS FULLY INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX NEPHROSTOMY BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17765547 ↗
  • MalfunctionMar 2, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX NEPHROSTOMY BALLOON CATHETER WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROLITHONOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, IT WAS NOTICED THAT THERE WAS A NARROW SPOT IN THE BALLOON AND THE RENAL SHEATH DID NOT CROSS WELL. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX NEPHROSTOMY BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF BALLOON TORN LONGITUDINALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13652909 ↗
  • MalfunctionOct 14, 2021Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX BALLOON CATHETER KIT WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROLITHONOMY (PCNL) PROCEDURE PERFORMED ON (B)(6) 2021. DURING THE PROCEDURE, WHEN THE BALLOON WAS INFLATED TO 17 ATM, THE PHYSICIAN PROCEEDED TO INSERT THE SHEATH INTO THE INCISION; HOWEVER, THE BALLOON SUDDENLY BURST EVEN BEFORE THE SHEATH COULD BE FULLY INSERTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX BALLOON CATHETER KIT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12631329 ↗
  • MalfunctionNov 9, 2020Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX BALLOON DILATATION CATHETER DEVICE WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROLITHOTOMY (PNL) PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, A PINHOLE WAS NOTED. REPORTEDLY, THE BALLOON WAS CHECKED OUTSIDE THE PATIENT AND CONFIRMED THAT THE PINHOLE WAS IN THE MIDDLE PART OF THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX BALLOON DILATATION CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10808911 ↗
  • MalfunctionJul 25, 2018Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX BALLOON KIT WAS USED DURING A PERCUTANEOUS URINARY TRACT STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON RUPTURED AT 18 ATMOSPHERIC PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX BALLOON KIT . THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE NO PROBLEM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7717735 ↗
  • MalfunctionNov 30, 2017Material Rupture; Material Rupture

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX BALLOON KIT WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROSTOMY PROCEDURE PERFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE DEFLATION, THE BALLOON RUPTURED. THERE WERE NO PIECES OF THE BALLOON THAT DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. REGARDING THE PATIENT¿S CONDITION AT THE CONCLUSION OF THE PROCEDURE, IT WAS NOTED THAT THERE WERE NO ADVERSE EVENTS OR PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7070411 ↗
  • MalfunctionJun 8, 2017Material Rupture; Material Rupture

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NEPHROMAX BALLOON CATHETER WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE PREFORMED ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE BALLOON BURST WHEN IT WAS INFLATED AT 14 OR 15 ATM. NO PIECES WERE DETACHED FROM THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER NEPHROMAX BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED AS "NO PATIENT CONSEQUENCES".

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6623290 ↗

Other CATHETER, NEPHROSTOMY devices

All CATHETER, NEPHROSTOMY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page