Flexor G09601Recalls & reported problems
Cook Incorporated · Flexor, Check-Flo, Introducer Set · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002096011), received between Feb 26, 2015 and Sep 4, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Material Separation2
- Break1
- Detachment Of Device Component1
- Difficult to Advance1
- Kinked1
- Material Split, Cut or Torn1
- Split1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient2
- Blood Loss1
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING AN UNKNOWN PROCEDURE, THE HUB OF A FLEXOR INTRODUCER SHEATH SEPARATED UPON REMOVAL OF THE DEVICE. THERE WERE NO SERIOUS CONSEQUENCES FOR THE PATIENT OR USER. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 22966152 ↗MalfunctionJul 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19654620 ↗MalfunctionOct 25, 2021Material Separation; Material Split, Cut or Torn; Material Separation; Material Split, Cut or Torn
NO ADDITIONAL INFORMATION HAS BEEN RECEIVED SINCE THE LAST REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12692580 ↗MalfunctionJul 26, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE CHECK-FLO VALVE OF A FLEXOR INTRODUCER SHEATH SEPARATED FROM THE DEVICE DURING A RIGHT EXTERNAL ILIAC RECANALIZATION. ACCESS WAS GAINED IN THE RIGHT COMMON FEMORAL ARTERY TARGETING THE RIGHT EXTERNAL ILIAC ARTERY. THE PATIENT DID NOT HAVE ANY TORTUOUS, CALCIFIED, OR SCARRED ANATOMY. THE DEVICE WAS USED OVER A 0.035 INCH AMPLATZ WIRE GUIDE. DURING RETRIEVAL OF THE DEVICE, THE CHECK-FLO VALVE DETACHED FROM THE DEVICE. RESISTANCE WAS NOT FELT DURING REMOVAL. THE DEVICE WAS PULLED BY THE HUB DURING REMOVAL, AND THE DILATOR WAS NOT IN PLACE. ANOTHER 9FR INTRODUCER WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12225618 ↗InjurySep 27, 2017Detachment Of Device Component; Kinked; Split; Detachment Of Device Component; Kinked; Split
A FLEXOR INTRODUCER SHEATH WAS USED IN AN UNSPECIFIED PROCEDURE. IT WAS REPORTED THAT, DURING THE PROCEDURE, THE TIP OF THE INTRODUCER DETACHED. AS A RESULT, THE ENTRY POINT WAS ENLARGED AND THE PATIENT HAD EXCESSIVE BLOOD LOSS AND HAD TO BE RESUSCITATED. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED.
Patient problems: Blood Loss
MAUDE report 6897354 ↗MalfunctionOct 3, 2016Break; Break
IT WAS REPORTED THAT DURING A FEMORAL ACCESS PROCEDURE ON A (B)(6) MALE LED PATIENT, THE FLEXOR INTRODUCER SHEATH DEVICE BROKE BETWEEN THE DESILET AND THE INTRODUCER. THE PATIENT REPORTEDLY HAD A TORTUOUS/CALCIFIED ANATOMY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5993773 ↗MalfunctionFeb 26, 2015Material Separation; Difficult to Advance
WHILE ATTEMPTING TO PERFORM A CROSSOVER PROCEDURE OVER THE SUPERFICIAL FEMORAL ARTERY, THE ROTAREX DID NOT PUSH OVER THE BIFURCATION THROUGH THE 8 F INTRODUCER. WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE DEVICES, THE CHECK FLO VALVES OF BOTH SHEATHS RIPPED WHILE PULLING OUT THE INTRODUCERS. THE PROCEDURE WAS SUCCESSFULLY PERFORMED WITH A TERUMO INTRODUCER. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DEU TO THIS OCCURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 4577509 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page