Performer G12664Recalls & reported problems
Cook Incorporated · Micropuncture Check-Flo Performer Introducer Set · Class 2
All models of Performer (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002126640), received between Jul 13, 2020 and Apr 23, 2025 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Puncture/Hole1
- Material Separation1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING A LOWER EXTREMITY ARTERIOGRAM WITH BALLOON DILATION VIA THE FEMORAL ARTERY, THE HUB OF A PERFORMER INTRODUCER SEPARATED FROM THE DEVICE. THE PATIENT HAD A HISTORY OF ARTERIOSCLEROSIS AND OCCLUSIVE DISEASE OF THE LOWER EXTREMITIES (FONTAINE STAGE IV). AFTER THE HUB SEPARATED, THE SHAFT OF THE DEVICE ¿FELL¿ INTO THE PATIENT¿S FEMORAL ARTERY AND WAS UNABLE TO BE REMOVED. THE PATIENT WAS TRANSFERRED TO ANOTHER HOSPITAL FOR SURGICAL REMOVAL OF THE SHAFT. THE SHAFT WAS SUCCESSFULLY REMOVED DURING THE ¿SECOND¿ SURGICAL PROCEDURE. A PHOTO PROVIDED BY THE CUSTOMER SHOWS HUB SEPARATION. THE SHAFT MATERIAL ALSO APPEARS TO HAVE BEEN CUT OR PARTIALLY TORN. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 21889758 ↗MalfunctionDec 11, 2024Unraveled Material; Unraveled Material
IT WAS ORIGINALLY REPORTED THAT PRIOR TO USE FOR AN UNKNOWN PROCEDURE THE USER FOUND THE WIRE GUIDE OF A PERFORMER INTRODUCER WAS BENT. THE PROCEDURE WAS COMPLETED BY USING ANOTHER UNSPECIFIED NEW DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. UPON PRELIMINARY EVALUATION OF THE RETURNED DEVICE ON 02DEC2024, IT WAS NOTED THAT WIRE GUIDE HAD UNRAVELED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20906470 ↗MalfunctionDec 9, 2020Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, PRIOR TO PATIENT CONTACT DURING A "TASE" PROCEDURE, A HOLE OR SEPARATION WAS NOTED BETWEEN THE PERFORMER INTRODUCER SHEATH AND HUB. AS THE USER ATTEMPTED TO INSERT THE DILATOR, IT DID NOT ENTER THE LUMEN OF THE SHEATH, BUT RATHER CAME OUT BETWEEN THE HUB AND SHEATH MATERIAL. THE DEVICE DID NOT MAKE CONTACT WITH THE PATIENT, AND ANOTHER DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10975463 ↗InjuryJul 13, 2020Material Separation; Material Separation
AS REPORTED, DURING AN ATTEMPTED PLACEMENT OF ANOTHER MANUFACTURER'S PERMCATH IN A PATIENT WITH AN AV FISTULA, A PERFORMER INTRODUCER SEPARATED IN THE PATIENT. THE INTRODUCER WAS HANDLED USING LATEX GLOVES AND FLUSHED WITH SALINE SOLUTION PRIOR TO USE. THE INTRODUCER WAS INSERTED INTO THE LEFT NECK OVER A GUIDE WIRE INCLUDED WITH THE PERMCATH. DURING REMOVAL OF THE INTRODUCER, THE DEVICE SEPARATED. A CUT-DOWN WAS PERFORMED TO EXPOSE THE VESSEL, AND A SNARE WAS USED TO SUCCESSFULLY RETRIEVE THE SEPARATED DEVICE PIECE FROM THE PATIENT. THE PATIENT'S ANATOMY WAS NOT CALCIFIED OR TORTUOUS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Code Available
MAUDE report 10263974 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page