Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002138162), received between Jul 19, 2019 and Jul 19, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Positioning Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 19, 2019Positioning Problem; Positioning Problem
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK STANDARD WIRE GUIDE (THSF-35-480). THE ERCP WAS PERFORMED TO TREAT A BILE DUCT STENOSIS. THE WIRE GUIDE WAS ADVANCED THROUGH ANOTHER MANUFACTURER'S ERCP CATHETER (FOR 0.035"), BUT THE THSF BECAME UNMOVABLE DURING THE ERCP. THEREFORE, THE THSF AND THE CATHETER WERE REMOVED TOGETHER FROM THE ENDOSCOPE [LOSS OF WIRE GUIDE ACCESS]. THEN, THE USER ATTEMPTED TO PULL OUT THE THSF FROM THE CATHETER BUT THE THSF GOT UNRAVELED DURING THE ATTEMPT. A COMPETITOR'S WIRE GUIDE WAS USED INSTEAD AND THE PROCEDURE WAS COMPLETED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8809157 ↗
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.