NCircle G18778Recalls & reported problems
Cook Incorporated · NCircle, Nitinol Tipless Stone Extractor · Class 2
All models of NCircle (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Cook NCircle ureteral stone extractor has 231 adverse event reports, including 24 injuries and no deaths. The reports mostly describe the wire or basket breaking and baskets that are difficult to open or close. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 222
- Deaths reported
- 0
- Injuries reported
- 24
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 222 adverse event reports that list this exact device (DI 00827002187788), received between Mar 4, 2015 and Aug 11, 2026 — classified as 24 injuries, 198 malfunctions.
The most frequently reported specific device problem is “Break” (103 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break103
- Difficult to Open or Close56
- Material Separation22
- Device Operates Differently Than Expected14
- Detachment of Device or Device Component9
- Loss of or Failure to Bond9
- Material Deformation9
- Device Damaged Prior to Use6
Patient problems
- No Consequences Or Impact To Patient101
- No Clinical Signs, Symptoms or Conditions56
- No Known Impact Or Consequence To Patient35
- No Patient Involvement13
- Device Embedded In Tissue or Plaque8
- No Information8
- Foreign Body In Patient4
- Complaint, Ill-Defined1
Recent adverse event reports
The 25 most recent of 222 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50210247 ↗MalfunctionJun 29, 2026Break
IT WAS REPORTED THE DISTAL END OF NCIRCLE TIPLESS STONE EXTRACTOR BROKE OFF INSIDE THE PATIENT DURING URETEROSCOPY PROCEDURE FOR A STONE EXTRACTION. THE BROKEN WIRE WAS RETRIEVED WITH ANOTHER EXTRACTOR. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THE DEVICE WAS TESTED PRIOR TO USE AND LASER WAS USED WHILE THE BASKET WAS USED. IT IS UNKNOWN IF THE PATIENT HAS TORTUOUS, CALCIFIED OR SCARRED ANATOMY. NO OTHER ADVERSE EFFECTS WERE REPORTED FOR THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25651991 ↗MalfunctionJun 23, 2026Difficult to Open or Close
IT WAS REPORTED THAT DURING A TRANSURETHRAL URETEROSCOPIC LITHOTRIPSY WITH STONE FRAGMENTATION AND REMOVAL AN NCIRCLE TIPLESS STONE EXTRACTOR BASKET COULD NOT BE OPENED AFTER FIVE STONE REMOVALS. THE BASKET WAS TESTED PRIOR TO USE AND WAS NORMAL. A NEW BASKET WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25597378 ↗MalfunctionJun 19, 2026Break
IT HAS BEEN REPORTED THAT DURING AN URETEROSCOPY PROCEDURE WITH LASER LITHOTRIPSY, A NCIRCLE TIPLESS STONE EXTRACTOR "BROKE" WHILE BEING USED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25562510 ↗MalfunctionApr 21, 2026Loss of or Failure to Bond
IT WAS REPORTED THAT ONE BASKET WIRE SEPARATED FROM THE DISTAL NOTCHED CANNULA OF AN NCIRCLE TIPLESS STONE EXTRACTOR DURING A KIDNEY PROCEDURE FOR A FEMALE PATIENT. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED BY USING A NEW DEVICE. A PHOTO PROVIDED BY THE CUSTOMER CONFIRMS WIRE SEPARATION. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24942405 ↗MalfunctionApr 15, 2026Loss of or Failure to Bond; Break
IT WAS REPORTED, DURING A FLEXIBLE URETEROSCOPY + PYELOSCOPY/LASERTRIPSY (KIDNEY STONE REMOVAL) THE BASKET OF AN NCIRCLE TIPLESS STONE EXTRACTOR HAD RUPTURED. THE PATIENT WAS X-RAYED AND THE DRAPE CLOSELY INSPECTED BUT THERE WAS NO SIGN OF THE BASKET. THE EXTRACTOR WAS USED TO REMOVE SEVERAL KIDNEY STONES. WHEN REMOVING THE LAST STONE, THE BASKET WAS OPENED TO GRAB THE STONE AND THE STONE WAS PULLED TO THE BLADDER. THE STONE WAS ABLE TO BE RETRIEVED. WHEN THE BASKET WAS PULLED OUT IT WAS FOUND THAT THE END OF THE BASKET WAS "FLIMSY." AS REPORTED, NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY AND THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24886087 ↗MalfunctionApr 6, 2026Break
IT HAS BEEN REPORTED THAT THE BASKET OF A NCIRCLE TIPLESS STONE EXTRACTOR BROKE DURING THE REMOVAL OF THE STONE FRAGMENT DURING A RETROGRADE INTRARENAL SURGERY (RIRS) PROCEDURE OF THE LEFT KIDNEY. ANOTHER SAME TYPE OF DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790840 ↗MalfunctionMar 31, 2026Material Deformation
IT WAS REPORTED THE BASKET WIRES OF AN NCIRCLE TIPLESS STONE EXTRACTOR LOST SHAPE DURING AN UNSPECIFIED PROCEDURE. THE CASE WAS BILATERAL, AND THE DEVICE FUNCTIONED AS EXPECTED ON THE FIRST SIDE BY SUCCESSFULLY CAPTURING AND EXTRACTING ALL STONE FRAGMENTS IN A SINGLE ATTEMPT. HOWEVER, ON THE SECOND SIDE, THE PRONGS OF THE BASKET FAILED TO MAINTAIN THE INTENDED 90-DEGREE SEPARATION UPON CLOSURE; INSTEAD, THEY SLID DOWNWARD TO APPROXIMATELY 10 DEGREES OF SEPARATION. THE PROCEDURE WAS COMPLETED WITH THE COMPLAINT DEVICE; HOWEVER, IT REQUIRED APPROXIMATELY A DOZEN ATTEMPTS TO CAPTURE THE FRAGMENT. IT WAS REPORTED THE LOSS OF SHAPE INCREASED THE PROCEDURE TIME BY A FEW MINUTES. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24742361 ↗MalfunctionMar 20, 2026Loss of or Failure to Bond
IT WAS REPORTED THAT THE PROXIMAL PORTION OF AN NCIRCLE TIPLESS STONE EXTRACTOR DETACHED DURING A TRANSURETHRAL LITHOTRIPSY (TUL). AFTER LASER LITHOTRIPSY OF THE URETERAL STONE, THE BASKET WAS USED FOR STONE EXTRACTION. AFTER ONE TO TWO RETRIEVAL ATTEMPTS, THE BASKET COULD NO LONGER BE OPENED OR CLOSED. UPON INSPECTION, IT WAS FOUND THAT THE 'PROXIMAL PORTION' OF THE BASKET HAD DETACHED. THE DEVICE WAS REPLACED WITH ANOTHER SAME-LIKE DEVICE TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654756 ↗MalfunctionMar 9, 2026Break
IT WAS REPORTED THAT THE BASKET TIP OF AN NCIRCLE TIPLESS STONE EXTRACTOR SEPARATED DURING A RETROGRADE INTRARENAL SURGERY (RIRS) PROCEDURE IN THE RIGHT KIDNEY. WHEN CHECKING THE OPEN AND CLOSE FUNCTION OF THE DEVICE PRIOR TO USE IN THE PATIENT, THE BASKET TIP SEPARATED. A NEW SIMILAR-LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24544966 ↗MalfunctionMar 2, 2026Difficult to Open or Close
IT WAS REPORTED, DURING A FLEXIBLE URETEROSCOPY, THE BASKET OF AN NCIRCLE TIPLESS STONE EXTRACTOR DID NOT OPEN CORRECTLY. NO DAMAGE TO THE DEVICE WAS REPORTED. THE ISSUE WAS DISCOVERED WHILE TESTING THE DEVICE, PRIOR TO PATIENT CONTACT AND THE DEVICE WAS NOT USED IN THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24483091 ↗MalfunctionFeb 24, 2026Break
IT WAS REPORTED ONE OF THE BASKET WIRES AT THE LOOP OF AN NCIRCLE TIPLESS STONE EXTRACTOR BROKE WHEN FRAGMENTING A CALCULUS DURING A FLEXIBLE URETERUS PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24428901 ↗MalfunctionFeb 23, 2026Loss of or Failure to Bond
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR¿S BASKET WIRE "BROKE" AND BASKET CANNOT BE PUSHED OUT PRIOR TO USE AN UNSPECIFIED KIDNEY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. A CUSTOMER ATTACHED PHOTO APPEARS TO SHOW PART OF ONE OF THE BASKET WIRES. THIS IS SEPARATED FROM THE DISTAL NOTCHED CANNULA. THE SMALL LOOP THAT JOINS THE TWO BASKET WIRES IS VISIBLE AT THE DISTAL TIP OF THE BASKET SHEATH INDICATING ONE OF THE BASKET LOOPS MAY STILL BE IN THE SHEATH. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE DEVICE DID NOT MAKE PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24414413 ↗MalfunctionFeb 17, 2026Break
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR¿S BASKET "MESH" BROKE DURING THE URETERORENOLITHOTRIPSY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24367687 ↗MalfunctionFeb 13, 2026Break
IT WAS REPORTED THE BASKET OF AN NCIRCLE TIPLESS STONE EXTRACTOR BROKE DURING A RIGHT URETEROSCOPY PROCEDURE. THE DEVICE BROKE INSIDE OF THE PATIENT UPON THE FIRST ATTEMPT TO REMOVE A SMALL URETERAL STONE. ALL FRAGMENTS WERE RETRIEVED USING A URETERAL GRASPER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. THE BASKET WAS USED ONCE. THE SIZE OF THE STONE IS UNKNOWN. THE LASER WAS NOT USED AT THAT TIME. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24333798 ↗MalfunctionFeb 9, 2026Loss of or Failure to Bond
IT WAS REPORTED, DURING TRANSURETHRAL FLEXIBLE URETEROSCOPIC LITHOTRIPSY, THE BASKET TIP OF AN NCIRCLE TIPLESS STONE EXTRACTOR BROKE. THE EXTRACTOR WAS TESTED PRIOR TO USE AND FUNCTIONED NORMALLY. AFTER APPROXIMATELY 5 OR 6 STONES WERE REMOVED, THE BASKET TIP BROKE DURING A STONE RETRIEVAL, RENDERING THE DEVICE UNUSABLE. A PHOTO WAS PROVIDED THAT APPEARS TO SHOW A BASKET WIRE PULLED FREE FROM BASKET CANNULA. A NEW BASKET WAS USED TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24297580 ↗MalfunctionJan 29, 2026Difficult to Open or Close
IT WAS REPORTED THAT THE BASKET DIAMETER OF AN NCIRCLE TIPLESS STONE EXTRACTOR BECAME SMALLER IN THE OPEN POSITION AFTER REMOVING STONES ABOUT 8 TO 10 TIMES MAKING IT UNABLE TO MEET THE REQUIREMENTS FOR STONE REMOVAL DURING A TRANSURETHRAL URETEROSCOPY LITHOTRIPSY PROCEDURE. THE DEVICE WAS REPLACED WITH A NEW BASKET TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24215819 ↗MalfunctionJan 15, 2026Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24084873 ↗MalfunctionJan 15, 2026Difficult to Open or Close
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24083014 ↗MalfunctionDec 17, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED, DURING AN UNSPECIFIED PROCEDURE, TWO NCIRCLE TIPLESS STONE EXTRACTORS (REFERENCE THIS REPORT) FAILED TO FULLY CLOSE WHILE THE SURGEON WAS EXTRACTING STONES FROM A PATIENT´S KIDNEY. THE SURGEON THEN CHANGED TO A THIRD SAME DEVICE, OF A DIFFERENT LOT, (REFERENCE PATIENT IDENTIFIER (B)(6) AND THE THIRD ONE ALSO FAILED. THE SURGEON THEN OPENED AN EQUIVALENT DEVICE FROM A DIFFERENT MANUFACTURER TO COMPLETE THE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, HOWEVER, AT THE TIME OF THIS REPORT, NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23832135 ↗MalfunctionDec 9, 2025Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23764896 ↗MalfunctionDec 1, 2025Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR¿S WIRE WAS BROKEN DURING A PROCEDURE IN THE BILE DUCT. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. A CUSTOMER PROVIDED PHOTO APPEARS TO SHOW ONE BASKET WIRE HAS PULLED FREE FROM BASKET CANNULA ON THE END. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23688161 ↗MalfunctionNov 20, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED, THERE WAS A FOREIGN MATERIAL FOUND INSIDE THE STERILE PACKAGING OF AN NCIRCLE TIPLESS STONE EXTRACTOR. THE ISSUE WAS DISCOVERED BY A COOK DISTRIBUTOR AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23606681 ↗MalfunctionNov 12, 2025Break; Difficult to Open or Close; Break; Difficult to Open or Close
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR WAS UNABLE TO OPEN COMPLETELY DURING THE FLEXIBLE URETEROLITHOSTRIPSY PROCEDURE. THIS ISSUE OCCURRED TWICE WITH TWO SEPARATE DEVICES. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23537877 ↗MalfunctionNov 12, 2025Break; Difficult to Open or Close; Break; Difficult to Open or Close
IT WAS REPORTED THAT THE NCIRCLE TIPLESS STONE EXTRACTOR WAS UNABLE TO OPEN COMPLETELY DURING THE FLEXIBLE URETEROLITHOSTRIPSY PROCEDURE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23537687 ↗
FDA clearance and classification
The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NCircle G18778
Has the NCircle G18778 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002187788), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NCircle G18778?
The FDA's MAUDE database lists 222 adverse event reports for this device, received between Mar 4, 2015 and Aug 11, 2026. In those reports, the most common specific device problems are “Break” (103), “Difficult to Open or Close” (56) and “Material Separation” (22); the most common patient problems are “Device Embedded In Tissue or Plaque” (8), “Foreign Body In Patient” (4) and “Complaint, Ill-Defined” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NCircle G18778?
None of the reports are classified as deaths, 24 as injuries and 198 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NCircle G18778 FDA approved?
Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.
Who makes the NCircle G18778, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand NCircle, model G18778). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL286 reports
- NCircle G46206Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL195 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL127 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL87 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
- NGage G48295Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL74 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.