NCircle G46206Recalls & reported problems
Cook Incorporated · NCircle, Nitinol Tipless Stone Extractor · Class 2
All models of NCircle (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Cook NCircle Tipless Stone Extractor has 204 adverse event reports, with 6 injuries and no deaths. The reports mostly describe the wire or tip breaking, the basket being hard to open or close, and a grip switch that could not be operated during kidney and ureter stone procedures. Many reports say the patient had no adverse effects. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 195
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 195 adverse event reports that list this exact device (DI 00827002462069), received between Mar 19, 2015 and Aug 11, 2026 — classified as 6 injuries, 189 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (77 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close77
- Break69
- Material Separation14
- Device Operates Differently Than Expected13
- Material Deformation11
- Device Damaged Prior to Use7
- Loss of or Failure to Bond7
- Detachment of Device or Device Component5
Patient problems
- No Consequences Or Impact To Patient93
- No Clinical Signs, Symptoms or Conditions53
- No Known Impact Or Consequence To Patient26
- No Patient Involvement17
- Insufficient Information3
- No Information3
- Device Embedded In Tissue or Plaque2
- Foreign Body In Patient1
Recent adverse event reports
The 25 most recent of 195 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50209379 ↗MalfunctionAug 7, 2026Loss of or Failure to Bond
It was reported, during a Ureteroscopy, the tip of an NCircle Tipless Stone Extractor was found broken.The issue was discovered prior to patient contact, when the device was removed from its product packaging. Preliminary evaluation of the returned device appears to show a basket wire has pulled out of the distal cannula. As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50181244 ↗MalfunctionMay 28, 2026Loss of or Failure to Bond
IT WAS REPORTED THAT UPON OPENING THE PACKAGING OF AN NCIRCLE TIPLESS STONE EXTRACTOR, THE WIRE OF THE EXTRACTOR WAS OBSERVED TO BE BROKEN PRIOR TO USE. THE USER REPLACED THE DEVICE WITH A NEW SIMILAR-LIKE DEVICE TO COMPLETE THE URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE WITHOUT DIFFICULTIES. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25321960 ↗MalfunctionMay 7, 2026Difficult to Open or Close
IT WAS REPORTED \"THE WIRE IN THE HANDLE PORTION PROTRUDED FROM THE GROOVE\" OF THE NCIRCLE TIPLESS STONE EXTRACTOR DURING A TRANSURETHRAL URETEROSCOPIC LITHOTRIPSY PROCEDURE WHILE USED WITH A URETEROSCOPE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25117061 ↗MalfunctionMay 5, 2026Break
AS REPORTED, UPON OPENING THE PACKAGE OF AN NCIRCLE TIPLESS STONE EXTRACTOR, THE USER FOUND ONE LINE OF THE BASKET WIRE TO BE BROKEN. THE URETEROSCOPIC LITHOTRIPSY (URSL) PROCEDURE WAS COMPLETED WITHOUT DIFFICULTIES BY USING A NEW NCIRCLE TIPLESS STONE EXTRACTOR. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25078785 ↗MalfunctionMar 24, 2026Break; Difficult to Open or Close
AS REPORTED, THE TIP OF A NCIRCLE TIPLESS STONE EXTRACTOR BROKE, RENDERING IT IMPOSSIBLE TO DEPLOY THE EXTRACTOR TO CAPTURE KIDNEY STONES DURING AN UNKNOWN PROCEDURE. A SECOND EXTRACTOR WAS OPENED TO CONTINUE AND COMPLETE THE PROCEDURE. THERE HAVE BEEN NO REPORT OF A PATIENT EXPERIENCING ADVERSE EFFECTS OR REQUIRING ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Insufficient Information
MAUDE report 24675477 ↗MalfunctionMar 24, 2026Difficult to Open or Close
No additional information regarding the patient and/or event has been received since the previous MedWatch report was sent.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24675162 ↗MalfunctionMar 17, 2026Difficult to Open or Close
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR WAS UNABLE TO OPEN COMPLETELY PRIOR TO AN UNSPECIFIED PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE DEVICE DID NOT MAKE PATIENT CONTACT. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615912 ↗MalfunctionMar 11, 2026Material Deformation
IT WAS REPORTED, DURING TRANSURETHRAL URETEROLITHOTRIPSY, THE NCIRCLE TIPLESS STONE EXTRACTOR WAS UNABLE TO GRASP THE STONE AFTER ONE OR TWO SUCCESSFUL RETRIEVALS OF FRAGMENTED CALCULI. THE PROCEDURE WAS PERFORMED UNDER FLEXIBLE URETEROSCOPY FOR RENAL AND URETERAL STONES. WHEN THE DEVICE COULD NO LONGER GRASP THE STONE, THE REMAINING FRAGMENTS WERE SUCCESSFULLY REMOVED USING ANOTHER DEVICE OF THE SAME TYPE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24567236 ↗MalfunctionMar 4, 2026Break
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR WAS DAMAGE-BROKEN PRIOR TO AN UNSPECIFIED PROCEDURE. THE ISSUE WAS FOUND WHEN THE PACKAGE WAS OPENED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE DEVICE DID NOT MAKE PATIENT CONTACT. THE COMPLAINT WAS ORIGINALLY NOT REPORTABLE DUE TO LACK OF INFORMATION. HOWEVER, COMPLAINT BECOME REPORTABLE FOLLOWING DEVICE EVALUATION AND INVESTIGATION COMPLETION.
Patient problems: Insufficient Information
MAUDE report 24511593 ↗MalfunctionMar 4, 2026Break; Difficult to Open or Close
IT WAS REPORTED THAT THE BASKET OF AN NCIRCLE TIPLESS STONE EXTRACTOR WOULD NOT DEPLOY NOR OPEN NORMALLY PRIOR TO USE DURING AN URETEROSCOPY WITH LASER LITHOTRIPSY (URSL) PROCEDURE. THE COMPLAINT DEVICE WAS REPLACED WITH A NEW SIMILAR-LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER DIFFICULTIES. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24509497 ↗MalfunctionFeb 25, 2026Loss of or Failure to Bond
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR¿S BASKET WIRE BROKE DURING A TRANSURETHRAL URETEROLITHOTRIPSY PROCEDURE. THE DEVICE WAS USED TO REMOVE A 3MM STONE. A SECOND STONE WAS ABLE TO BE REMOVED AND THE ISSUE WAS DISCOVERED. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. THE COMPLAINT DEVICE WAS INSPECTED PRIOR TO USE TO ENSURE NO DAMAGE HAD OCCURRED. THE BASKET WAS TESTED IN THE UNCOILED POSITION PRIOR TO USE AND FUNCTIONED PROPERLY. A LASER WAS NOT USED WHILE THE BASKET WAS USED. NO PART OF THE DEVICE SEPARATED. THERE WAS NOTHING WITH THE PATIENT¿S ANATOMY KEEPING THE BASKET FROM OPENING OR CLOSING. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24442900 ↗MalfunctionFeb 17, 2026Break; Difficult to Open or Close
IT WAS REPORTED THAT THE NCIRCLE TIPLESS STONE EXTRACTOR¿S HANDLE WAS STIFF AND WAS UNABLE TO FUNCTION DURING A TRANSURETHRAL LITHOTOMY (TUL) PROCEDURE. THE DEVICE WAS INSERTED INTO THE SCOPE CHANNEL. WHEN ATTEMPTING TO OPEN AND CLOSE THE BASKET, THE HANDLE WAS STIFF AND DIFFICULT TO OPERATE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. THE STONE WAS LOCATED IN THE URINARY TRACT. IT IS UNKNOWN IF THE BASKET WAS TESTED PRIOR TO USE. NO PART OF THE DEVICE SEPARATED IN THE PATIENT. THE PATIENT¿S ANATOMY WAS NOT KEEPING THE BASKET FROM OPENING OR CLOSING. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24365165 ↗MalfunctionFeb 9, 2026Break; Material Deformation
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR¿S BASKET WAS DEFORMED DURING A TRANSURETHRAL URINARY STONE REMOVAL PROCEDURE, MAKING IT DIFFICULT TO EXTRACT THE CALCULUS. THE CALCULUS WAS PREVIOUSLY FRAGMENTED USING A LASER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-TYPE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24298926 ↗MalfunctionFeb 9, 2026Break
IT WAS REPORTED, THE WIRE OF AN NCIRLCE TIPLESS STONE EXTRACTOR BROKE DURING A TRANS URETERAL LITHOTRIPSY (TUL) ENDOSCOPIC PROCEDURE FOR URETERAL STONES FOLLOWED BY STONE RETRIEVAL. DURING USE, THE WIRE OF THE BASKET BOKE AFTER GRASPING A URETERAL STONE AND CLOSING THE BASKET. THE DEVICE WAS THEREFORE NOT USED FURTHER. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. THE DEVICE WAS TESTED PRIOR TO USE AND NO ISSUES WERE OBSERVED. DURING THE INITIAL ATTEMPT, NO EXCESSIVE MANIPULATION, ANGULATION, OR FORCE WAS USED. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24298255 ↗MalfunctionFeb 9, 2026Difficult to Open or Close
IT WAS REPORTED THAT THE HANDLE OF AN NCIRCLE TIPLESS STONE EXTRACTOR WAS DAMAGED. AFTER OPENING THE PACKAGE AND PRIOR TO USE, A FUNCTIONAL TEST WAS CONDUCTED AND THE HANDLE WAS FOUND TO BE DAMAGED. PER PROVIDED CUSTOMER PHOTOS, THE METAL CANNULATED HANDLE COMPONENT WAS BENT. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24296158 ↗MalfunctionJan 9, 2026Break
IT WAS REPORTED AFTER ADVANCING THE NCIRCLE TIPLESS STONE EXTRACTOR INTO THE URETER AND ATTEMPTING TO PLACE THE STONE IN THE BASKET, THE PHYSICIAN FOUND THAT THE BASKET WIRE HAD SEVERED. THE COMPLAINT DEVICE WAS REPLACED WITH A NEW SAME-LIKE DEVICE (LOT UNKNOWN) TO COMPLETE THE PROCEDURE. THE DEVICE WAS INSPECTED PRIOR TO USE AND WAS UNDAMAGED. THE DEVICE WAS TESTED IN ITS UNCOILED POSITION AND THE BASKET OPERATED CORRECTLY. NO LASER WAS USED DURING BASKET OPERATION, AND THERE WERE NO ANATOMICAL ISSUES PREVENTING THE BASKET FROM OPENING OR CLOSING. THE HANDLE WAS NOT POSITIONED STRAIGHT TOWARD THE PATIENT OR POINTED AT THE CEILING. NO PARTS SEPARATED. THE PATIENT WAS UNDER ANESTHESIA. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24027624 ↗MalfunctionDec 5, 2025Break; Break
AS REPORTED, DURING A TRANSURETHRAL UROLITHOTRIPSY PROCEDURE, AN NCIRCLE TIPLESS STONE EXTRACTOR WAS USED TO RETRIEVE A STONE. AFTER OPENING THE PRODUCT, A TEST OPERATION WAS PERFORMED, AND NO PROBLEMS WERE OBSERVED. WHEN ATTEMPTING TO GRASP THE STONE, THE DEVICE WAS UNABLE TO DO SO ON THE FIRST ATTEMPT. UPON REMOVAL AND INSPECTION, IT WAS FOUND THAT THE BASE OF THE BASKET HAD SEVERED. THE WIRE HAD BROKEN OFF, DESPITE NO EXCESSIVE FORCE BEING APPLIED DURING USE. ANOTHER NCIRCLE TIPLESS STONE EXTRACTOR FROM A DIFFERENT LOT WAS UTILIZED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT FURTHER ISSUES. THERE WERE NO CONCERNS REGARDING DEVICE HANDLING AT THE FACILITY OR BY THE PHYSICIANS, WHO ARE EXPERIENCED WITH THIS PRODUCT. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THERE WAS NO UNINTENDED SECTION OF THE DEVICE THAT REMAINED INSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23729669 ↗MalfunctionDec 1, 2025Break; Break
IT WAS REPORTED THE NCIRCLE TIPLESS STONE EXTRACTOR BROKEN DURING AN UNSPECIFIED PROCEDURE. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23687841 ↗MalfunctionNov 24, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED DURING A TRANSURETHAL LITHOTRIPSY (TUL) PROCEDURE, THE STONE WAS FRAGMENTED USING A LASER AND ONE FRAGMENT WAS RETRIEVED USING A NCIRCLE TIPLESS STONE EXTRACTOR, BUT THE BASKET THEN BECAME UNABLE TO BE OPENED OR CLOSED. ANOTHER SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE COMPLAINANT HAS NOT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630155 ↗MalfunctionNov 19, 2025Break; Break
AS REPORTED, PRIOR TO AN URETEROSCOPY LITHOTRIPSY (URSL) PROCEDURE, THE USER OPENED THE PACKAGE OF AN 'NCIRCLE TIPLESS STONE EXTRACTOR' AND FOUND THAT ONE END OF THE BASKET WIRE WAS PULLED FREE FROM THE BASKET CANNULA. THE PROCEDURE WAS COMPLETED WITHOUT DIFFICULTIES BY USING ANOTHER 'NCIRCLE TIPLESS STONE EXTRACTOR'. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23597529 ↗MalfunctionNov 12, 2025Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress
IT WAS REPORTED, AN NCIRCLE TIPLESS STONE EXTRACTOR¿S OUTER SUPPORT SHEATH KINKED NEAR THE HANDLE ON THE PROTECTION SHEATH. THE DEVICE WAS UNABLE TO OPEN AND CLOSE DURING A TRANSURETHRAL LITHOTRIPSY (TUL) PROCEDURE WHILE ATTEMPTING TO REMOVE A RENAL STONE. THE PROCEDURE WAS COMPLETED BY USING ANOTHER UNSPECIFIED DEVICE. IT IS UNKNOWN WHETHER THE DEVICE WAS INSPECTED BEFORE USE, TESTED IN DIFFERENT POSITIONS, FUNCTIONED PROPERLY, USED WITH A SCOPE, THE SCOPE¿S CHANNEL SIZE, STONE SIZE, LASER USAGE, PATIENT ANATOMY IMPACT, USE OF OTHER DEVICES, DEVICE ORIENTATION, NUMBER OF STONES CAPTURED BEFORE FAILURE, ANY PART SEPARATION, OR IF THE PATIENT WAS UNDER ANESTHESIA. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23533574 ↗MalfunctionNov 5, 2025Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress
ADDITIONAL INFORMATION WAS RECEIVED 29OCT2025. INITIALLY, THE BASKET WAS STIFF AND DID NOT OPEN OR CLOSE SMOOTHLY. AFTER ATTEMPTING TO OPEN AND CLOSE THE DEVICE ABOUT FIVE TIMES, THE BASKET SHEATH SEVERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23479630 ↗MalfunctionNov 5, 2025Break; Break
AS REPORTED, DURING A TRANSURETHRAL URINARY STONE REMOVAL PROCEDURE, THE BASKET SHEATH OF 'NCIRCLE TIPLESS STONE EXTRACTOR' CAME OFF. A TRANSURETHRAL APPROACH WAS PERFORMED USING A URETEROSCOPE TO REMOVE THE STONE FROM THE URETER. THE DEVICE WAS INSPECTED FOR DAMAGE AND TESTED PRIOR TO USE IN AN UNCOILED POSITION. REPORTEDLY, THE BASKET FUNCTIONED PROPERLY. ANOTHER MANUFACTURER'S URETEROSCOPE WAS USED. A LASER WAS NOT USED. THERE WAS NOTHING WITH THE PATIENT'S ANATOMY KEEPING THE BASKET FROM OPENING OR CLOSING. THE HANDLE WAS NOT IN THE STRAIGHT POSITION. THE BASKET CAPTURED STONES MORE THAN FIVE TIMES. THEN, THE SHEATH CAME OFF FROM THE HANDLE WHILE REMOVING STONE APPROXIMATELY THREE MILLIMETER IN SIZE. REPORTEDLY, THE PHYSICIAN TOOK THE SHEATH "BY HAND" AND PULLED THE ENTIRE BASKET OUT OF THE ENDOSCOPIC CHANNEL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER SIMILAR DEVICE FROM THE IN-HOUSE INVENTORY. AS PER PHYSICIAN'S COMMENT, THE PHYSICIAN WAS FAMILIAR WITH THE DEVICE AND USED AS USUAL, BUT THE SHEATH BECAME DETACHED AFTER SEVERAL USES. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23475896 ↗MalfunctionOct 29, 2025Break; Break
IT WAS REPORTED THAT THE BASKET WIRE OF AN NCIRCLE TIPLESS STONE EXTRACTOR BROKE DURING A FLEXIBLE TRANSURETHRAL LITHOTRIPSY PROCEDURE. AFTER LASER CRUSHING, THE DEVICE WAS USED TO GRASP A 3MM STONE IN THE URINARY DUCT. ONE OF THE FOUR BASKET WIRES BROKE WHEN PULLING IT INTO THE ACCESS SHEATH. THE PROCEDURE WAS COMPLETED BY USING A SIMILAR-LIKE DEVICE. ADDITIONALLY, NO DAMAGE TO THE DEVICE WAS OBSERVED PRIOR TO USE. THE DEVICE WAS TESTED IN AN UNCOILED POSITION AND FUNCTIONED PROPERLY PRIOR TO USE. THE LASER WAS NOT USED WHILE THE BASKET WAS USED. PATIENT'S URETERAL STENOSIS KEPT THE BASKET FROM OPENING OR CLOSING. THE HANDLE WAS NOT IN THE STRAIGHT POSITION TOWARDS THE PATIENT¿S BODY NOR POINTED TOWARDS THE CEILING. THE BASKET WAS NOT ABLE TO CAPTURE ANY STONES. NO PART OF THE BASKET SEPARATED. THE PATIENT DID NOT HAVE TO RETURN FOR A SEPARATE PROCEDURE. AS REPORTED, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23417377 ↗
FDA clearance and classification
The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NCircle G46206
Has the NCircle G46206 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002462069), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NCircle G46206?
The FDA's MAUDE database lists 195 adverse event reports for this device, received between Mar 19, 2015 and Aug 11, 2026. In those reports, the most common specific device problems are “Difficult to Open or Close” (77), “Break” (69) and “Material Separation” (14); the most common patient problems are “Device Embedded In Tissue or Plaque” (2) and “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NCircle G46206?
None of the reports are classified as deaths, 6 as injuries and 189 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NCircle G46206 FDA approved?
Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.
Who makes the NCircle G46206, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand NCircle, model G46206). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DISLODGER, STONE, BASKET, URETERAL, METAL devices
- Zero Tip M0063901050Boston Scientific Corporation · DISLODGER, STONE, BASKET, URETERAL, METAL286 reports
- NCircle G18778Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL222 reports
- NCircle G17520Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL127 reports
- NGage G48297Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL96 reports
- NGage G48294Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL92 reports
- NCircle G19110Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL87 reports
- Bard® SkyLite™ Tipless Nitinol Stone Basket 041900C. R. Bard, Inc. · DISLODGER, STONE, BASKET, URETERAL, METAL78 reports
- NGage G48295Cook Incorporated · DISLODGER, STONE, BASKET, URETERAL, METAL74 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.