Zero Tip M0063901050Recalls & reported problems

Boston Scientific Corporation · Nitinol Stone Retrieval Basket · Class 2

All models of Zero Tip (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Zero Tip is a stone retrieval basket used to remove stones from the urinary tract. Many of its 288 reports describe the basket breaking or its wire detaching during procedures, and some describe broken pieces left in the patient. The reports include 111 injuries and no deaths. No recalls name this device. Reports fell from 47 in the previous 12 months to 27 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
286
Deaths reported
0
Injuries reported
110
Recalls naming it
0

Summary

The FDA's MAUDE database contains 286 adverse event reports that list this exact device (DI 08714729469643), received between Dec 20, 2016 and Sep 1, 2026 — classified as 110 injuries, 176 malfunctions.

The most frequently reported specific device problem is “Break” (129 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 4 reportsApr 2024: 2 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 5 reportsAug 2024: 11 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 4 reportsJan 2025: 2 reportsFeb 2025: 4 reportsMar 2025: 1 reportsApr 2025: 8 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 14 reportsAug 2025: 1 reportsSep 2025: 4 reportsOct 2025: 2 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 5 reportsApr 2026: 4 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break129
  • Detachment of Device or Device Component82
  • Adverse Event Without Identified Device or Use Problem62
  • Difficult to Open or Close27
  • Material Split, Cut or Torn17
  • Nonstandard Device9
  • Tear, Rip or Hole in Device Packaging9
  • Unsealed Device Packaging9

Patient problems

  • No Clinical Signs, Symptoms or Conditions153
  • No Consequences Or Impact To Patient67
  • No Known Impact Or Consequence To Patient19
  • Urinary Tract Infection18
  • Foreign Body In Patient12
  • Sepsis7
  • No Patient Involvement5
  • Device Embedded In Tissue or Plaque3

Recent adverse event reports

The 25 most recent of 286 reports, in the reporter's own words as published by the FDA.

  • InjurySep 1, 2026Device Contamination with Chemical or Other Material; Adverse Event Without Identified Device or Use Problem

    It was reported that a patient with this Inflatable Penile Prosthesis (IPP) experienced device erosion. A surgical procedure was performed during which the device was explanted and replaced with a malleable prosthesis to allow the patient to heal prior to replacement with a new inflatable implant There were no further patient complications reported.

    Patient problems: Erosion; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50553492 ↗
  • InjuryJul 8, 2026Break

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A PROCEDURE, REPORTEDLY, WHEN THE BASKET WAS USED TO RETRIEVE THE STONE FRAGMENT, THE TIP BROKE INSIDE THE PATIENT. THE PHYSICIAN WAS USING THE BASKET WITH A LASER AND THERE WERE NO PIECES OF FRAGMENT WAS LEFT INSIDE THE PATIENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25755185 ↗
  • MalfunctionJul 8, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING URETEROLITHOTOMY PROCEDURE, A ZERO TIP BASKET WAS USED, AND IT WAS FOUND THAT THE WHOLE BASKET FELL OFF OUTSIDE OF THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25751066 ↗
  • MalfunctionMay 15, 2026Break; Nonstandard Device; Material Split, Cut or Torn

    IT WAS REPORTED THAT ZERO TIP BASKET DEVICE WAS USED DURING A TRANSURETHRAL RENOLITHOTRIPSY PROCEDURE. DURING PROCEDURE, THE WIRE GOT DAMAGE IN THE BASKET SECTION. THE PROCEDURE WAS COMPLETED WITH THIS DEVICE AND THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ANALYSIS OF THE RETURNED DEVICE REVEALED THE SHEATH WAS FOUND TO BE MELTED AT THE DISTAL END.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25191484 ↗
  • MalfunctionApr 14, 2026Nonstandard Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PROCEDURE A ZERO TIP WAS USED. IT WAS FOUND THAT THE WIRE WAS DAMAGED AND WAS SEPARATED AND DETACHED FROM THE BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24872666 ↗
  • MalfunctionApr 13, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING PROCEDURE A ZERO TIP NITINOL STONE RETRIEVAL WAS USED. THE BASKET BROKE OFF INSIDE THE URETER DURING ATTEMPT TO REMOVE THE BROKEN NEPHROSTOMY TUBE. THE PROCEDURE WAS COMPLETED WITHOUT FURTHER INCIDENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24866763 ↗
  • MalfunctionApr 2, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT TWO ZERO TIP BASKET DEVICES WERE USED DURING A URETEROSCOPY (URS) PROCEDURE. DURING THE PROCEDURE, THE BASKET PORTION BROKE OFF. THE SAME ISSUE OCCURRED WITH THE SECOND ZERO TIP BASKET DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. THIS REPORT PERTAINS TO THE SECOND OF TWO ZERO TIP BASKET DEVICES USED IN THE SAME PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24767737 ↗
  • MalfunctionApr 1, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT TWO ZERO TIP BASKET DEVICES WERE USED DURING A URETEROSCOPY (URS) PROCEDURE. DURING THE PROCEDURE, THE BASKET PORTION BROKE OFF. THE SAME ISSUE OCCURRED WITH THE SECOND ZERO TIP BASKET DEVICE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. THIS REPORT PERTAINS TO THE FIRST OF TWO ZERO TIP BASKET DEVICES USED IN THE SAME PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24753973 ↗
  • MalfunctionMar 11, 2026Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT A ZERO TIP WAS USED DURING A URETEROSCOPY LASER LITHOTRIPSY PROCEDURE. PRIOR TO PROCEDURE, A BLACK SPECS WERE FOUND FLOATING IN THE PACKAGE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568785 ↗
  • InjuryMar 10, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP WAS USED DURING PROCEDURE. IT WAS FOUND THAT THE BASKET BROKE INSIDE THE PATIENT RESULTING IN ADDITIONAL SURGERY. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24562941 ↗
  • InjuryMar 10, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A URETEROSCOPY (URS) PROCEDURE. DURING THE PROCEDURE, TONE RETRIEVAL BASKET WAS ADVANCED THROUGH A SEMIRIGID URETEROSCOPE AND DEPLOYED TO CAPTURE THE STONE; HOWEVER, A RESISTANCE WAS ENCOUNTERED DURING ATTEMPTED WITHDRAWAL OF THE BASKET WITH THE CAPTURED STONE, AND THE BASKET BROKE IN THIS ATTEMPT. REPORTEDLY, THE URETEROSCOPE WAS WITHDRAWN AND A 16 FR FOLEY CATHETER WAS PLACED FOR URINARY DRAINAGE, AND A GUIDEWIRE WAS LEFT IN PLACE VIA THE CATHETER. A CLEAR URINE DRAINAGE WAS OBSERVED AT THE CONCLUSION OF THE CASE WITH NO PATIENT COMPLICATIONS.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24562929 ↗
  • MalfunctionMar 10, 2026Nonstandard Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A URETEROLITHOTOMY (TUL) PROCEDURE. POST PROCEDURE, THE FRONT FOUR CLAWS AND THE ENTIRE BASKET DETACHED. REPORTEDLY, THE DETACHED BASKET DID FALL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24562905 ↗
  • MalfunctionMar 3, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET WAS USED DURING A URETEROLITHOTOMY PROCEDURE. DURING THE PROCEDURE CLOSURE, IT WAS FOUND THAT THE WHOLE BASKET FELL OFF BUT IT DID NOT FALL INTO THE PATIENT'S BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24501036 ↗
  • MalfunctionFeb 27, 2026Nonstandard Device; Detachment of Device or Device Component; Material Split, Cut or Torn

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A URETEROLITHOTOMY PROCEDURE. REPORTEDLY, POST PROCEDURE IT WAS FOUND THAT THE FRONT FOUR CLAWS AND THE COATING OF THE BASKET WAS DETACHED OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24468870 ↗
  • MalfunctionFeb 18, 2026Nonstandard Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A URETEROLITHOTOMY PROCEDURE A ZERO TIP BASKET WAS INTENDED TO BE USED, IT WAS FOUND THAT FRONT FOUR CLAWS WERE DETACHED, HOWEVER, IT DID NOT FALL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379308 ↗
  • MalfunctionFeb 6, 2026Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THE PLASTIC AROUND THE ZERO TIP BASKET DEVICE HAD A HOLE IN IT, DOWN TOWARDS THE BOTTOM, MAKING IT NON-STERILE. THERE WERE NO PATIENT AND PROCEDURE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283313 ↗
  • MalfunctionFeb 4, 2026Break

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A PROCEDURE. REPORTEDLY, THE WIRE WAS DAMAGED AND SEPARATED FROM THE BASKET SECTION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24264158 ↗
  • InjuryJan 26, 2026Nonstandard Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE DURING A RETROGRADE INTRARENAL SURGERY (RIRS) PROCEDURE. DURING THE PROCEDURE, THE ENTIRE BASKET BREAK AND DETACHED. REPORTEDLY, IT WAS RETRIEVED USING ANOTHER BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24184914 ↗
  • MalfunctionJan 15, 2026Break

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS SET TO BE USED DURING A KIDNEY STONE REMOVAL PROCEDURE. REPORTEDLY, AFTER UNPACKING IT WAS FOUND THAT ONE OF THE BASKET WIRES COMPLETELY FELL OFF FROM THE BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24088620 ↗
  • MalfunctionDec 31, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT, AT THE END OF THE PROCEDURE, IT WAS OBSERVED THAT THE "FRONT FOUR [BASKET WIRES]" WERE DETACHED BUT HAD NOT FALLEN INTO THE PATIENT. IT WAS REPORTED THAT THE PATIENT'S CONDITION FOLLOWING THE PROCEDURE WAS "STABLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23935941 ↗
  • MalfunctionDec 26, 2025Break; Nonstandard Device; Break; Nonstandard Device

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USE DURING A PROCEDURE. REPORTEDLY, ONE OF THE BASKET WIRES BROKE; HOWEVER, ANALYSIS OF THE RETURNED DEVICE REVEALED THAT ONE OF THE BASKET WIRES BREAK AND FULLY DETACHED FROM THE BASKET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23897640 ↗
  • InjuryDec 4, 2025Break; Break

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A PROCEDURE FOR STONE REMOVAL. DURING THE PROCEDURE, A PIECE OF BASKET BROKE OFF WHILE IN PATIENT. REPORTEDLY, THEY WERE ABLE TO RETRIEVE THE BROKEN PIECE OUT OF THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716777 ↗
  • MalfunctionNov 28, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS SET TO BE USE DURING A PROCEDURE. DURING UNPACKING, A HOLE WAS FOUND IN THE STERILE PACKAGING. THERE WERE NO PROCEDURE AND PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23671867 ↗
  • MalfunctionNov 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS SET TO BE USED DURING A URETERAL CALCULI PROCEDURE. DURING PREPARATION, IT WAS FOUND THAT ONE END OF THE BASKET WAS FRACTURED. IT WAS FURTHER CLARIFIED THAT THE ENTIRE BASKET WAS DETACHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23617023 ↗
  • InjuryNov 3, 2025Nonstandard Device; Detachment of Device or Device Component; Nonstandard Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT A ZERO TIP BASKET DEVICE WAS USED DURING A URETERIC STONE REMOVAL. DURING THE PROCEDURE, THE BASKET WAS INSERTED INSIDE THE PATIENT AND OPENED TO GRAB THE STONE. HOWEVER, WHEN IT WAS BEING REMOVED THE BASKET SNAPPED PART WAY ALONG. REPORTEDLY, ALL PARTS WERE RETRIEVED USING ANOTHER BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23455818 ↗

FDA clearance and classification

The FDA regulates product code FFL (dislodger, stone, basket, ureteral, metal) under 21 CFR 876.4680 as a Class II device. 21 CFR 876.4680 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Zero Tip M0063901050

Has the Zero Tip M0063901050 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08714729469643), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Zero Tip M0063901050?

The FDA's MAUDE database lists 286 adverse event reports for this device, received between Dec 20, 2016 and Sep 1, 2026. In those reports, the most common specific device problems are “Break” (129), “Detachment of Device or Device Component” (82) and “Difficult to Open or Close” (27); the most common patient problems are “Urinary Tract Infection” (18), “Foreign Body In Patient” (12) and “Sepsis” (7). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Zero Tip M0063901050?

None of the reports are classified as deaths, 110 as injuries and 176 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Zero Tip M0063901050 FDA approved?

Devices of this type (product code FFL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Boston Scientific Corporation.

Who makes the Zero Tip M0063901050, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand Zero Tip, model M0063901050). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other DISLODGER, STONE, BASKET, URETERAL, METAL devices

All DISLODGER, STONE, BASKET, URETERAL, METAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page