Flexor G19174Recalls & reported problems
Cook Incorporated · Flexor, Ureteral Access Sheath with AQ, Hydrophilic Coating · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002191747), received between Apr 3, 2025 and Jun 10, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2025Peeled/Delaminated; Peeled/Delaminated
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22179537 ↗MalfunctionJun 4, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THE INNER WALL OF THE DISTAL END OF THE OUTER SHEATH OF A FLEXOR URETERAL ACCESS SHEATH AND DILATOR DETACHED DURING A RETROGRADE INTRARENAL SURGERY (RIRS). THE SHEATH WAS SUCCESSFULLY PLACED. DURING ADVANCE OF THE SOFT ENDOSCOPE, RESISTANCE WAS ENCOUNTERED AT THE DISTAL END OF THE DEVICE WHICH PREVENTED THE ENDOSCOPE FROM ADVANCING. THE PROCEDURE WAS COMPLETED BY USING ANOTHER SAME-LIKE DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION DUE TO THIS OCCURRENCE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22140920 ↗MalfunctionApr 3, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED DURING A RETROGRADE INTRARENAL SURGERY (RIRS) TO REMOVE A STONE IN THE RIGHT KIDNEY, A PIECE OF PLASTIC WAS FOUND INSIDE THE DILATOR OF THE FLEXOR URETERAL ACCESS SHEATH AND DILATORS. THE PHYSICIAN DISCONTINUED ITS USE AND COMPLETED THE SURGERY USING THE ACCESS SHEATH ONLY. NO SECTION OF THE DEVICE REMAINED INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. THE COMPLAINANT HAS NOT REPORTED ANY ADVERSE EFFECTS ON THE PATIENT DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21762499 ↗
Other endoscopic access overtube, gastroenterology-urology devices
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · endoscopic access overtube, gastroenterology-urology1 recall42 reports
- Trokamed WA2PS20STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS20LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS18LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Trokamed WA2PS18STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Navigator HD M0062502220Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology1 death31 reports
- Navigator HD M0062502230Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology29 reports
- Proxis™ Ureteral Access Sheath 231045C. R. Bard, Inc. · endoscopic access overtube, gastroenterology-urology22 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.