Roadrunner G22419Recalls & reported problems
Cook Incorporated · Roadrunner Wire Guide · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002224193), received between Sep 9, 2016 and Sep 9, 2016 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture1
- Improper or Incorrect Procedure or Method1
- Off-Label Use1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2016Fracture; Off-Label Use; Improper or Incorrect Procedure or Method; Fracture; Off-Label Use; Improper or Incorrect Procedure or Method
DURING A RENDEZVOUS PROCEDURE IN THE STOMACH, THE PHYSICIAN USED A COOK ROADRUNNER WIRE GUIDE. THE WIRE GUIDE SHREDDED AFTER BEING PULLED THROUGH A COVIDIEN 22G SHARK CORE FNA/FNB NEEDLE, WHICH IS NOT INDICATED TO BE USED WITH A WIRE GUIDE IN THIS MANNER [OFF LABEL USE]. THE TIP WAS MISSING. THERE WAS NO HARM TO THE PATIENT, HOWEVER THE TIP OF THE WIRE GUIDE REMAINED IN THE PATIENT'S STOMACH. THE DEVICE WAS RECEIVED FOR EVALUATION ON 08/25/2016. TWO (2) DEVICES WERE RECEIVED WITH THE SAME MALFUNCTION (BOTH TIPS DETACHED). THE FOLLOWING CLARIFICATION ON THE SECOND DEVICE WAS RECEIVED FROM THE SALES REPRESENTATIVE: "THIS WIRE WAS USED IN THE EXACT SAME CASE WITH THE EXACT SAME RESULTS." THE FOLLOWING INFORMATION ON THE RETRIEVAL OF THE DETACHED WIRE GUIDE TIPS WAS RECEIVED 08/29/2016 VIA THE SALES REPRESENTATIVE: "THE TIPS WERE ALLOWED TO PASS NATURALLY THROUGH THE PATIENT". ON 10/12/2016, WE RECEIVED A MEDWATCH MW5065042 VIA THE FDA THAT STATED THE FOLLOWING: "PATIENT WAS UNDERGOING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PHYSICIAN ATTEMPTED MAIN PANCREATIC CANNULATION BUT WAS UNSUCCESSFUL. HE THEN ATTEMPTED AN ENDOSCOPIC ULTRASONOGRAPHY (EUS) GUIDED ERCP, BUT THE
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 5939691 ↗
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.