Saeed G22692Recalls & reported problems
Cook Incorporated · Six Shooter Saeed Multi-Band Ligator · Class 2
All models of Saeed (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Cook Six Shooter multi-band ligator is used to band varicose veins in the esophagus. The 285 reports mostly describe bands that fail to fire or fire early, bands that come loose, and parts of the device left inside the patient. The reports include 8 injuries and no deaths. No recalls name this device, and reports fell to 27 in the last year from 47 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 283
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 283 adverse event reports that list this exact device (DI 00827002226920), received between Feb 26, 2016 and Sep 14, 2026 — classified as 8 injuries, 275 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (135 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire135
- Premature Activation88
- Device Dislodged or Dislocated19
- Unintended Movement13
- Device-Device Incompatibility10
- Positioning Failure10
- Break9
- Material Too Soft/Flexible8
Patient problems
- No Clinical Signs, Symptoms or Conditions199
- No Consequences Or Impact To Patient63
- Foreign Body In Patient20
- No Patient Involvement16
- Device Embedded In Tissue or Plaque2
- Hemorrhage/Blood Loss/Bleeding2
- Gastrointestinal Hemorrhage1
- Hemostasis1
Recent adverse event reports
The 25 most recent of 283 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026Premature Activation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50763004 ↗MalfunctionJun 26, 2026Premature Activation
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT WHEN THE THIRD BAND WAS RELEASED AND THE FOURTH BAND WAS READY TO BE RELEASED, IT COULD NOT BE ROTATED FOR RELEASE. EVEN AFTER ADJUSTING THE POSITION, IT COULD NOT BE RELEASED. USER HAD TO RETRACT THE ENDOSCOPE, REPLACE THE LIGATION DEVICE WITH A NEW ONE, AND CONTINUE THE TREATMENT. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25641615 ↗MalfunctionJun 17, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED TWO (2) COOK 6 SHOOTER SAEED MULTI-BAND LIGATORS. IT WAS REPORTED THAT THE BAND ENTANGLED WITH TRIGGER CORD, BANDS CANNOT BE DEPLOYED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25536489 ↗MalfunctionJun 10, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PHYSICIAN INSTALLED THE DEVICE NORMALLY, AFTER ASPIRATION THE PHYSICIAN RELEASED THE DEVICE AND FOUND THAT THE RELEASED WHITE BAND STUCK UNDER BAND AND THE ENDOSCOPE CANNOT BE RETRACTED. AFTER RELEASING ALL THE RUBBER RINGS [BANDS] AT ONCE, THE ENDOSCOPE CAN BE RETRACTED. OTHER THAN THE DEPLOYED BANDS WHICH WILL PASS NATURALLY, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25459625 ↗MalfunctionJun 5, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT AFTER TWO BANDS OF LIGATION, THE THIRD BAND COULD NOT BE RELEASED. AFTER REMOVING THE ENDOSCOPE AND REINSTALLING THE DEVICE, IT STILL COULD NOT BE RELEASED. THE ENDOSCOPE BODY HAD BEEN SEVERELY TWISTED BY THE TRIGGER CORD OF THE DEVICE. AFTER REPLACING WITH ANOTHER BOX OF MBL-6-F, THE REMAINING VARICOSE VEIN LIGATION WAS COMPLETED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25406609 ↗MalfunctionJun 4, 2026Premature Activation
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT DURING INSTALLATION THE BANDS DETACHED. USER CHANGED TO A NEW DEVICE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25397388 ↗MalfunctionJun 4, 2026Premature Activation
DURING AN ENDOSCOPIC VARICEAL LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT DURING INSTALLATION THE BANDS DETACHED. USER CHANGED TO A NEW DEVICE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25394821 ↗MalfunctionMay 12, 2026Premature Activation
DURING AN ESOPHOGEAL LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PATIENT HAS A HISTORY OF HEPATITIS B-RELATED CIRRHOSIS ACCOMPANIED BY ESOPHAGOGASTRIC VARICES. HE PRESENTED WITH "HEMATEMESIS" AND WAS ADMITTED TO THE OPERATING ROOM FOR "ESOPHAGEAL VARICEAL LIGATION + GASTRIC VARICEAL TISSUE GLUE CLOSURE" UNDER GENERAL ANESTHESIA WITH TRACHEAL INTUBATION FROM 23:20 ON (B)(6) 2026, TO 23:39 ON THE SAME DAY. DURING THE ESOPHAGEAL VARICEAL LIGATION PROCEDURE, AN ADVERSE EVENT OCCURRED WHEN SIX BANDS WERE RELEASED SIMULTANEOUSLY [BANDS PREMATURELY DEPLOY - SUBJECT OF REPORT] DURING THE LIGATION OF THE BLOOD VESSELS AT THE CARDIA. THEREFORE, THE SECOND LIGATION DEVICE WAS PROMPTLY REPLACED, AND THE BLOOD VESSELS WERE LIGATED IN A CIRCULAR MANNER FROM THE BOTTOM UP. AFTER THE PROCEDURE, THE VARICES IN THE UPPER ESOPHAGUS WERE SIGNIFICANTLY REDUCED. THE PATIENT'S VITAL SIGNS WERE STABLE, AND HE WAS ESCORTED BACK TO THE WARD AFTER RESUSCITATION. THIS EVENT OCCURRED DURING THE PROCEDURE, WITH SIX BANDS RELEASED SIMULTANEOUSLY AROUND THE BLOOD VESSELS, RESULTING IN SIGNIFICANT EDEMA AROUND THE VESSELS BUT NO VESSEL RUPTURE OR MAJOR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25149106 ↗MalfunctionApr 27, 2026Premature Activation
DURING PREPARATION FOR AN ESOPHOGEAL LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE USER OPENED THE PACKAGE AND DISCOVERED THAT THE BANDS DETACHED [PREMATURE DEPLOYMENT-SUBJECT OF REPORT]. A NEW LIGATION DEVICE WAS SUBSTITUTED TO COMPLETE THE PROCEDURE THE OBSERVATION WAS MADE PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. IN ADDITION, THE PATIENT SUSTAINED NO CLINICAL CONSEQUENCE AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25000041 ↗MalfunctionApr 21, 2026Premature Activation
DURING AN ENDOSCOPIC VARICEAL LIGATION FOR CIRRHOTIC ESOPHAGOGASTRIC VARICES, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT FOUR BANDS WERE RELEASED AT ONE TIME. THE DEVICE FUNCTIONED NORMALLY AFTER REPLACEMENT WITH A NEW LIGATOR. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24939105 ↗MalfunctionMar 4, 2026Off-Label Use
DURING AN ENDOSCOPIC PROCEDURE IN THE STOMACH, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT UPON RELEASING THE FIRST RUBBER BAND FROM THE LIGATURE, HE LOOSENED THE CORD AND RELEASED ALL THE RUBBER BANDS [PREMATURE DEPLOYMENT] OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24505526 ↗MalfunctionMar 4, 2026Failure to Fire
DURING AN ENDOSCOPIC LIGATION PROCEDURE FOR INTERNAL HEMORRHOIDS IN THE ANORECTAL REGION, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PATIENT WAS ALREADY IN POSITION, AND THE NURSE INSTALLED DEVICE NORMALLY. HOWEVER, THE FIRST ATTEMPT TO RELEASE [BANDS] WAS UNSUCCESSFUL AS THE PATIENT'S POSITION WAS ADJUSTED THE ANGLE. WHEN ATTEMPTING TO RELEASE IT AGAIN, THE DEVICE GOT STUCK. THE NURSE QUICKLY REPLACED IT WITH A NEW SET OF LIGATING DEVICES AND COMPLETED THE PROCEDURE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24505273 ↗MalfunctionMar 2, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION (EVL) PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE USER ATTACHED THE COOK BAND LIGATION DEVICE MBL-6-F TO THE ENDOSCOPE AND PROCEEDED WITH THE BAND LIGATION TREATMENT THROUGH THE PATIENT'S MOUTH INTO THE ESOPHAGUS. WHEN STARTING WITH THE FIRST BAND, USER FELT THAT THE TENSION OF THE TRIGGER CORD WAS STRONG MAKING IT DIFFICULT TO TRIGGER THE BAND TO RELEASE. WHEN RELEASING THE SECOND BAND, USER DESCRIBED THAT EVEN THOUGH SHE PULLED HARD ON THE TRIGGER CORD, SHE COULDN'T TRIGGER THE BAND TO RELEASE. IT MIGHT BE THAT THE THIRD BAND RELEASED UNEXPECTEDLY UNDER FORCE, FAILING TO BAND THE TARGET AREA, AND DUE TO THE TIGHTNESS OF THE TRIGGER CORD, IT WAS DIFFICULT TO TRIGGER. THEREFORE, THE DEVICE HAD TO BE REMOVED AND REPLACED WITH A NEW ONE TO COMPLETE THE PROCEDURE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24485121 ↗MalfunctionFeb 9, 2026Failure to Fire
DURING AN ENDOSCOPIC LIGATION FOR INTERNAL HEMORRHOIDS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE USER ADVANCED THE DEVICE TO THE DESIRED POSITION AND READY TO DEPLOY WHEN IT WAS OBSERVED THAT THE BANDS COULD NOT BE DEPLOYED AND THE TRIGGER CORD WAS TANGLED. [UNABLE TO DEPLOY BANDS - SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292621 ↗MalfunctionFeb 6, 2026Failure to Fire
DURING AN ENDOSCOPIC LIGATION PROCEDURE FOR MIXED HEMORRHOIDS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT DURING THE PROCEDURE, THE FIRST BAND DEPLOYMENT WAS DIFFICULT, AFTER THE FIRST BAND RELEASED THE SECOND BAND IMMEDIATELY DETACHED, AND THIRD BAND CANNOT BE DEPLOYED, THE ENDOSCOPE IS BENT. USER RETRACTED THE DEVICE FROM PATIENT AND TESTED WHICH CONFIRMED THE BAND CANNOT BE RELEASED. [UNABLE TO EFFECTIVELY DEPLOY BANDS - SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24280818 ↗MalfunctionJan 30, 2026Failure to Fire; Unintended Movement
DURING AN ENDOSCOPIC VARICEAL LIGATION (EVL) IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PRODUCT WAS CONNECTED NORMALLY AND FOLLOWING THE MANUFACTURER'S INSTRUCTIONS. WHEN ATTEMPTING TO PULL THE FIRST LIGATURE [BAND] FROM THE LIGATURE KIT, THE CORD BROKE, PREVENTING THE NECESSARY PULL TO RELEASE THE LIGATURES." A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24225234 ↗MalfunctionJan 21, 2026Break; Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION (EVL) IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THROUGH GASTROSCOPY, THE LOCATION OF THE VARICOSE VEIN WAS DISCOVERED. AFTER RETRACTING THE ENDOSCOPE, THE DEVICE WAS INSTALLED OUTSIDE THE BODY ACCORDING TO THE PROCEDURE, AND SUCTION WAS ATTEMPTED. IT ENTERED THE ESOPHAGUS AND REACHED THE VARICOSE VEIN LOCATION. THE FIRST TO THIRD BANDS WERE RELEASED NORMALLY, BUT THE FOURTH BAND WAS FOUND TO BE UNABLE TO BE RELEASED NORMALLY. PRESSURE WAS APPLIED TO THE OPERATING HANDLE, ROTATED, AND THE TRIGGER CORD WAS BROKEN. AFTER RETRACTING, SCISSORS WERE USED TO CUT THE TRIGGER CORD TIE, AND THE PRODUCT WAS REPLACED TO COMPLETE THE PROCEDURE. OTHER THEN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24131230 ↗MalfunctionJan 15, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION (EVL) IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE HANDLE COULD NOT BE LOCKED [DIFFICULT OR UNABLE TO DEPLOY BANDS - SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24079794 ↗MalfunctionJan 9, 2026Premature Activation
IN PREPARATION FOR AN ENDOSCOPIC HEMORRHOIDAL LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE USER OPENED THE PACKAGE AND FOUND OUT THE BAND IS PREMATURE DETACHED [PREMATURE DEPLOYMENT - SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. IN ADDITION, THE PATIENT SUSTAINED NO CLINICAL CONSEQUENCE AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24020483 ↗MalfunctionJan 8, 2026Failure to Fire; Unintended Movement
DURING AND ENDOSCOPIC VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE DOCTOR USED THE FUJI GASTROSCOPE TO INSTALL THE LIGATOR AND ENTERED THE ESOPHAGUS, AFTER 3 BANDS OF LIGATION, THE FOURTH BAND COULD NOT BE RELEASED, AND THE LENS BODY HAD BEEN PULLED AND TWISTED BY THE THREAD OF THE LIGATOR EVEN MORE. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24002983 ↗MalfunctionJan 5, 2026Failure to Fire
DURING AN ENDOSCOPIC VARICEAL LIGATION (EVL), THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE USER INSTALLED THE DEVICE ONTO A FUJI ENDOSCOPE AND ADVANCED INTO THE ESOPHAGUS AND RELEASED 3 BANDS BUT THE FOURTH BAND CANNOT BE DEPLOYED. THE USER RETRACTED THE DEVICE FROM THE ENDOSCOPE AND RE-INSTALLED THE DEVICE, BUT THE BAND STILL CANNOT BE DEPLOYED, AND THE ENDOSCOPE BODY IS TWISTED AS THE TRIGGER CORD PULLING [UNABLE TO DEPLOY BANDS - SUBJECT OF REPORT]. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23964475 ↗MalfunctionDec 22, 2025Failure to Fire; Failure to Fire
DURING AND ENDOSCOPIC VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT BEFORE USE, INSTALL THE DEVICE ACCORDING TO THE IFU. WHEN RELEASING IT [DEPLOY BAND], IT WAS FOUND THAT THE BAND CANNOT BE DEPLOYED. AFTER ROTATING SEVERAL TIMES, IT COULD NOT BE DEPLOYED. REMOVE THE TRANSPARENT CAP RUBBER RING FROM THE FRONT END, CUT THE TRIGGER CORD, AND RETRACTED THE DEVICE. USER CHANGED TO ANOTHER DEVICE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865279 ↗MalfunctionDec 4, 2025Failure to Fire; Failure to Fire
DURING AND ENDOSCOPIC VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT AT 3:00 PM ON (B)(6) 2025 DURING AN ENDOSCOPIC HEMOSTASIS PROCEDURE IN THE OPERATING ROOM, THE PHYSICIAN PREPARED FOR TREATMENT AFTER INSTALLING THE LIGATOR. UPON RELEASING THE RUBBER BAND, IT WAS FOUND THAT THE PULL CORD HAD BECOME ENTANGLED WITH THE BAND AND COULD NOT BE SEPARATED. ATTEMPTS TO PULL IT FREE PROVED UNSUCCESSFUL, PREVENTING THE RELEASE OF THE NEXT BAND [UNABLE TO DEPLOY BANDS - SUBJECT OF REPORT]. THE PHYSICIAN IMMEDIATELY REPLACED THE LIGATOR WITH ANOTHER FROM THE SAME BATCH NUMBER AND SUCCESSFULLY COMPLETED THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23720156 ↗MalfunctionNov 17, 2025Failure to Fire; Failure to Fire
DURING AND ENDOSCOPIC VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PATIENT UNDERWENT A GASTROINTESTINAL BLEEDING LIGATION PROCEDURE. THE LIGATOR WAS INSTALLED WITHOUT INCIDENT. HOWEVER, DURING THE ESOPHAGEAL VARICEAL LIGATION, THE LIGATOR FAILED TO DEPLOY PROPERLY. A NEW LIGATOR HAD TO BE REPLACED TO COMPLETE THE LIGATION PROCEDURE. THIS PROLONGED THE PROCEDURE DURATION, INCREASED THE PATIENT'S BLOOD LOSS, EXTENDED THE PATIENT'S EXPOSURE TIME, AND HEIGHTENED THE RISK OF INFECTION, POTENTIALLY CAUSING SERIOUS CONSEQUENCES. USER CHANGED TO A NEW DEVICE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23569682 ↗MalfunctionNov 5, 2025Premature Activation; Premature Activation
DURING AND ENDOSCOPIC VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PATIENT WAS PREPARED FOR THE PROCEDURE. THE USER ATTACHED THE COOK MBL-6-F LIGATION DEVICE TO THE ENDOSCOPE AND ADVANCE IT THROUGH THE PATIENT'S MOUTH INTO THE ESOPHAGUS FOR LIGATION. THE FIRST THREE BANDS WERE RELEASED NORMALLY. WHILE APPLYING THE FOURTH BAND, THE OPERATOR REPORTED TWISTING IT ONLY ONCE, YET BOTH THE FOURTH AND FIFTH BAND RELEASED SIMULTANEOUSLY AND BECAME LODGED IN THE TISSUE. SUBSEQUENTLY, IT WAS DISCOVERED THAT NO FURTHER BANDS REMAINED. USER NOTED THAT THE DEVICE WAS MISSING ONE BAND. SINCE THE PATIENT REQUIRED MORE THAN FOUR LIGATION BANDS, THE DEVICE HAD TO BE REMOVED AND REPLACED WITH A NEW LIGATOR TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23477336 ↗
FDA clearance and classification
The FDA regulates product code MND (ligator, esophageal) under 21 CFR 876.4400 as a Class II device. 21 CFR 876.4400 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Common questions about the Saeed G22692
Has the Saeed G22692 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002226920), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Saeed G22692?
The FDA's MAUDE database lists 283 adverse event reports for this device, received between Feb 26, 2016 and Sep 14, 2026. In those reports, the most common specific device problems are “Failure to Fire” (135), “Premature Activation” (88) and “Device Dislodged or Dislocated” (19); the most common patient problems are “Foreign Body In Patient” (20), “Device Embedded In Tissue or Plaque” (2) and “Hemorrhage/Blood Loss/Bleeding” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Saeed G22692?
None of the reports are classified as deaths, 8 as injuries and 275 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Saeed G22692 FDA approved?
Devices of this type (product code MND) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.
Who makes the Saeed G22692, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Saeed, model G22692). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Ligator, esophageal devices
- Speedband Superview Super 7 M00542250Boston Scientific Corporation · Ligator, esophageal6 deaths4,632 reports
- Saeed G22553Cook Incorporated · Ligator, esophageal286 reports
- SINGLE USE LIGATING DEVICE HX-400U-30Olympus Medical Systems Corp. · Ligator, Esophageal1 recall234 reports
- Saeed G31917Cook Incorporated · Ligator, esophageal118 reports
- SINGLE USE LIGATING DEVICE HX-400U-30Olympus Medical Systems Corp. · Ligator, Esophageal95 reports
- Single Use Rotatable Clip Fixing Device HX-201UR-135Olympus Medical Systems Corp. · Ligator, esophageal1 recall1 death10 reports
- Single Use Rotatable Clip Fixing Device HX-201UR-135LOlympus Medical Systems Corp. · Ligator, esophageal1 recall8 reports
- Single Use Rotatable Clip Fixing Device HX-201UR-135Olympus Medical Systems Corp. · Ligator, esophageal1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page