Quick summary

This Cook Six Shooter multi-band ligator has 122 adverse event reports, with no deaths and 8 injuries. Reports most often describe bands that fail to fire or fire early, the device coming loose or moving out of position, and parts breaking off, and some mention pieces left in the patient. Reports fell from 15 in the previous year to 8 in the latest year. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
118
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 118 adverse event reports that list this exact device (DI 00827002319172), received between Feb 19, 2016 and Jul 1, 2026 — classified as 8 injuries, 110 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 4 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire29
  • Device Dislodged or Dislocated23
  • Material Too Soft/Flexible19
  • Premature Activation18
  • Material Separation13
  • Positioning Failure11
  • Detachment of Device or Device Component10
  • Break7

Patient problems

  • No Clinical Signs, Symptoms or Conditions58
  • No Consequences Or Impact To Patient41
  • No Patient Involvement12
  • Foreign Body In Patient11
  • Hemorrhage/Blood Loss/Bleeding3
  • Gastrointestinal Hemorrhage2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 118 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 1, 2026Device Dislodged or Dislocated

    DURING A LIGATION PROCEDURE FOR GASTRIC / ESOPHAGEAL BLEEDING, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE PATIENT REQUIRED A GASTROSCOPY TO RULE OUT AN UPPER GI BLEED DUE TO MELENA STOOLS. PATIENT WAS HEMODYNAMICALLY STABLE. CONSCIOUS SEDATION ADMINISTERED. LARGE ESOPHAGEAL VARICES WITH HIGH-RISK SPOTS VISUALIZED. ONE OF WHICH APPEARED THAT IT HAD BEEN BLEEDING (RED WHEAL SPOT). THERE WAS AN ATTEMPT TO DEPLOY A BAND TO THE VARIX, BUT IT FELL OFF. CONSEQUENTLY, VARIX RUPTURED AND THERE WAS PROFUSE BLEEDING. SCOPE WAS REMOVED, BANDING DEVICE REMOVED AND SCOPE INSERTED AGAIN TO ATTEMPT THROMBOJET INJECTION TO BLEEDING SITE, THIS WAS UNSUCCESSFUL AS IT WAS IMPOSSIBLE TO LOCATE THE BLEEDING SITE DUE TO SEVERE BLEEDING. 300ML OF FRESH BLOOD SUCTIONED FROM PATIENT'S MOUTH/NOSE. AIRWAYS WERE MAINTAINED. PATIENT BECAME HYPOXEMIC DESPITE OXYGEN 3L/MIN VIA NASAL CANNULA, AND CODE 66 WAS CALLED. WHILE WAITING FOR THE CODE TEAM, OXYGENATION AND AIRWAYS WERE MAINTAINED.

    Patient problems: Gastrointestinal Hemorrhage

    MAUDE report 25680534 ↗
  • MalfunctionMar 30, 2026Device Handling Problem

    IN PREPARATION FOR AN ESOPHAGOGASTRODUODENOSCOPY WITH BANDING, THE PHYSICIAN SELECTED FOUR COOK 6 SHOOTER SAEED MULTI-BAND LIGATORS. IT WAS REPORTED THAT PRIOR TO USE THE PHYSICIAN ALWAYS FIRES OFF ONE BAND, THE BANDS ARE NOT THE SIZE THEY SHOULD BE WHEN COMING OFF THE BARREL. THEY APPEAR TO BE BIGGER THAN NORMAL. IN THE PICTURE [PROVIDED FOR EVALUATION], THE SMALL BAND CAME FROM [THE DEVICE] THAT WAS USED TO COMPLETE THE PROCEDURE. ALL FOUR DEVICES WERE USED IN THE SAME PROCEDURE AND THE DEVICES WERE DISCARDED AFTER THE INITIAL BAND WAS FIRED OFF BEFORE MAKING PATIENT CONTACT. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. IN ADDITION, THE USER SUSTAINED NO CLINICAL CONSEQUENCE AND THERE WERE NO ADVERSE EFFECTS TO THE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24717882 ↗
  • MalfunctionJan 26, 2026Failure to Fire

    DURING AN ESOPHAGODUODENOSCOPY WITH VARICEAL BANDING IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT ONE OF THE LIGATORS MALFUNCTIONED AND WOULD NOT FIRE THE BAND. A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24179305 ↗
  • MalfunctionNov 18, 2025Detachment of Device or Device Component

    FAULTY THREADER BLUE HOOKS CAME OFF BOTH ENDS TRYING TO THREAD THE STRING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583996 ↗
  • MalfunctionOct 16, 2025Failure to Fire; Failure to Fire

    DURING AN UNKNOWN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE "BANDER STRING KEPT SLIPPING OFF." [INTERPRETED AS UNABLE TO EFFECTIVELY DEPLOY BANDS - SUBJECT OF REPORT]. THEY HAD TO ATTEMPT 4 TIMES TO GET IT TO STAY IN PLACE AND REALLY HAD TO CRANK ON THE DIAL TO GET THE BAND TO RELEASE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23313131 ↗
  • MalfunctionSep 17, 2025Premature Activation; Premature Activation

    DURING PREPARATION FOR AN ENDOSCOPIC VARICEAL LIGATION (EVL) PROCEDURE, THE PHYSICIAN SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE "RUBBER BAND WASN'T UNDER CONTROL." TWO BANDS WAS [WERE] DROPPED OFF BY ITSELF [PREMATURE DEPLOYMENT] AFTER FIXED THE BARREL AND BEFORE INTRODUCED INTO PATIENT. AFTER INSERTED TO PATIENT, THE DEVICE WAS UNABLE TO FIRE THE BANDS. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. IN ADDITION, THE PATIENT SUSTAINED NO CLINICAL CONSEQUENCE AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23072760 ↗
  • MalfunctionAug 14, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING A VARICEAL LIGATION (EVL) PROCEDURE FOR ESOPHAGEAL VARICES, THE PHYSICIAN CHOSE A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE BANDS WERE PLACE ON THE VARIX, THEN RELEASED AND WENT INSIDE THE BARREL. IT SEEMS LIKE THE BANDS SLIPPED OFF TISSUE RIGHT AWAY. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN IN THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22803279 ↗
  • MalfunctionJun 11, 2025Device Handling Problem; Device Handling Problem

    IN PREPARATION FOR A GASTROSCOPY PROCEDURE FOR LIGATION OF ESOPHAGEAL VARICES, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE CLINICIAN DEPLOYED A BAND OUTSIDE OF THE PATIENT TO TEST THE BAND. THIS IS NORMAL PROCEDURE FOR THIS CLINICIAN. THE CLINICIAN OBSERVED THE BAND DID NOT RESTRICT IN SIZE AS IT USUALLY WOULD [SUBJECT OF REPORT]. THE CLINICIAN THEN USED THIS DEVICE ON A PATIENT'S VARICES, DEPLOYING 2 X BANDS. ANOTHER OF THE SAME DEVICE WAS SET UP ON ANOTHER GASTROSCOPE AND ALL 6 X BANDS WERE DEPLOYED TO TREAT ESOPHAGEAL VARICES. HEMOSTASIS WAS ACHIEVED AND THE PATIENT REQUIRED NO FURTHER TREATMENT. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22183153 ↗
  • MalfunctionMay 23, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING AN ESOPHAGEAL VARICES' LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE BANDS DIDN'T FUNCTION AND DID NOT HOLD ON TO THE VARIX. 4 BANDS DEPLOYED BUT NONE WERE ABLE TO LIGATE ANY VARIX. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22080431 ↗
  • MalfunctionApr 18, 2025Premature Activation; Premature Activation

    DURING AN UNSPECIFIED LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE FIRST TWO BANDS DEPLOYED CORRECTLY BUT THE LAST 4 MISFIRED. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21867252 ↗
  • MalfunctionApr 4, 2025Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IN PREPARATION FOR AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE USER SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT WHEN PASSING THE COOK 6 SHOOTER BANDING KIT THROUGH THE SCOPE, THE THREADING DEVICE BROKE (BLUE THREADER BROKE OFF WHEN PULLING IT THROUGH DEVICE IN SCOPE). THEY ATTEMPTED TO USE THE OTHER END OF THE BLUE THREADER, AND IT BROKE OFF WHEN PULLING THROUGH SCOPE ATTACHMENT WHEEL AS WELL. USED A SECOND KIT AND HAD NO FURTHER PROBLEMS WITH IT. THIS OCCURRED PRIOR TO PATIENT CONTACT; THERE WAS NO IMPACT TO THE PATIENT. IN ADDITION, THE PATIENT SUSTAINED NO CLINICAL CONSEQUENCE AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21774612 ↗
  • MalfunctionMar 21, 2025Break; Failure to Fire; Break; Failure to Fire

    DURING AN UNSPECIFIED LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT WHEN THE ENDOSCOPE WAS REMOVED, WE [USERS] NOTICED THAT THE "WIRES HOLDING THE CAP (ON WHICH THE LIGATURES ARE PLACED) WERE SEVERED" [TRIGGER CORD BROKEN]. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21668000 ↗
  • MalfunctionMar 19, 2025Premature Activation; Premature Activation

    DURING AN UNSPECIFIED LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE LAST LIGATION WAS NEVER FOUND, NOR INITIATED, NOR SEEN [INTERPRETED AS BAND PREMATURELY DEPLOYS]. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21646594 ↗
  • MalfunctionMar 19, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING AN UNSPECIFIED LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT ONE LIGATURE WAS TRIGGERED BUT DID NOT STAY IN PLACE [BANDS FALLS OFF VARIX]. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21646592 ↗
  • MalfunctionMar 5, 2025Failure to Fire; Failure to Fire

    DURING A LIGATION PROCEDURE FOR ESOPHAGEAL VARICES, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE BANDS WOULD NOT DEPLOY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21526983 ↗
  • MalfunctionJan 7, 2025Failure to Fire; Failure to Fire

    DURING AN UNSPECIFIED LIGATION PROCEDURE THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS INITIALLY REPORTED THAT THE DEVICE WAS DEFECTIVE AFTER ASSEMBLY ON THE INSTRUMENT. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. ADDITIONAL INFORMATION ABOUT THE EVENT WAS RECEIVED ON 11 DEC 2024 STATING THAT THE RELEASE SYSTEM LOCKED AFTER MOUNTING ON THE INSTRUMENT [INTERPRETED AS UNABLE TO ROTATE THE HANDLE LEADING TO UNABLE TO DEPLOY BANDS]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21093040 ↗
  • MalfunctionNov 21, 2024Device Dislodged or Dislocated; Material Too Soft/Flexible; Device Dislodged or Dislocated; Material Too Soft/Flexible

    DURING A VARICEAL LIGATION PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED 2 COOK 6 SHOOTER SAEED MULTI-BAND LIGATORS. IT WAS REPORTED THAT THERE WAS A COMPLAINT BUT NO FURTHER DETAILS WERE PROVIDED. NO DETAILS WERE PROVIDED THAT HAS HISTORICALLY CAUSED OR CONTRIBUTED TO A SERIOUS INJURY TO THE PATIENT, THEREFORE THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. ADDITIONAL INFORMATION WAS RECEIVED ON 28 OCT 2024 STATING THAT THE FIRST RINGS WERE EASY TO SET AND THE OTHER RINGS COULD BE REMOVED, BUT DID NOT FIT OVER THE SUCKED-IN TISSUE. THE TISSUE WAS OPTIMALLY SUCKED IN AS THE PHYSICIAN HAS A LOT OF EXPERIENCE WITH OUR 6-SHOOTER. THE RINGS THEN LAY LOOSE IN THE PATIENT. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20745915 ↗
  • MalfunctionNov 21, 2024Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IN PREPARATION FOR A LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT DURING THE EXTRACTION THE BLUE HOOK, WHERE THE TRIGGER CHORDS IS ATTACHED TO, BROKE [INTERPRETED AS BLUE HOOK DETACHING FROM THE LOADING CATHETER]. THIS OCCURRED PRIOR TO PATIENT CONTACT. OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION REGARDING PATIENT OUTCOME WERE UNSUCCESSFUL. WHILE THE COMPLAINANT DID NOT SPECIFY IF THE PATIENT EXPERIENCED ANY ADVERSE EFFECTS OR REQUIRED ADDITIONAL MEDICAL PROCEDURES DUE TO THIS EVENT, THE INFORMATION ABLE TO BE COLLECTED DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20745671 ↗
  • MalfunctionNov 1, 2024Material Separation; Detachment of Device or Device Component; Misassembly by Users; Material Separation; Detachment of Device or Device Component; Misassembly by Users

    DURING A LIGATION PROCEDURE, THE PHYSICIAN SELECTED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT BOTH ENDS OF THE LOADING CATHETER WERE TORN OFF THE THREAD RETRACTION. WHEN THE CORD WAS PULLED THROUGH THE WORKING CHANNEL, BOTH ENDS [BLUE HOOKS] OF THE [LOADING] CATHETER CAME LOOSE IN THE WORKING CHANNEL. THIS HAPPENED IN THE MIDDLE OF THE WORKING CHANNEL AND NOT AT THE EXIT FROM THE VALVE. ANOTHER 6-SHOOTER HAD TO BE OPENED TO COMPLETE THE INVESTIGATION. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20588582 ↗
  • MalfunctionOct 25, 2024Failure to Fire; Failure to Fire

    DURING A LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT DURING THE PROCEDURE, "DISADAPTATION OF THE DISPOSITIVE BEFORE JETTISONING THE 1ST ELEMENT (THE PRACTITIONER BEGAN TO TURN THE KNOB)" [UNABLE TO EFFECTIVELY DEPLOY BANDS]. THIS LED TO THE USE OF A NEW DEVICE AND PROLONGED THE INTERVENTION [PROCEDURE] BY ABOUT 15 MINUTES. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20535902 ↗
  • MalfunctionSep 12, 2024Device Dislodged or Dislocated; Material Too Soft/Flexible; Device Dislodged or Dislocated; Material Too Soft/Flexible

    DURING VARICEAL ESOPHAGEAL BANDING, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT AS HE ATTEMPTED TO CAPTURE THE VARIX IN THE CAP WHILE SUCTIONING AND TO RELEASE THE BAND OVER THE TISSUE. THE BAND RELEASED BUT DID NOT HAVE THE USUAL TIGHT DIAMETER WHEN DEPLOYED, UPON SEVERAL ATTEMPTS THE BANDS APPEARED TO BE STRETCHED, I.E., THE RESTING INNER DIAMETER WAS MUCH LARGER AS IT APPEARED TO BE EXPANDED, UNABLE TO GRAB AND SECURE THE TISSUE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20205601 ↗
  • MalfunctionAug 13, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING A VARICEAL VEIN LIGATION IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT THE BANDS DID NOT HAVE ENOUGH ELASTICITY, SO THEY WOULDN'T STAY ON. THE PROCEDURE WAS COMPLETED BY USING ANOTHER OF THE SAME DEVICE. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19972346 ↗
  • MalfunctionJun 25, 2024Premature Activation; Premature Activation

    DURING A LIGATION PROCEDURE, THE PHYSICIAN USED A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT WHEN THEY RELEASED THE BANDS, THE BANDS CAME OUT TWO BY TWO (IN EACH SHOT TWO BANDS LIFT) [PREMATURE DPLOYMENT]. THERE ARE 6 BANDS IN THE DEVICE. SO, AFTER 3 RELEASES (ALL 3 SUCCESSFUL) THEY HAD TO OPEN ANOTHER DEVICE TO FINISH THE PROCEDURE. OTHER THAN THE DEPLOYED BANDS, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19604507 ↗
  • MalfunctionApr 11, 2024Failure to Fire; Failure to Fire

    DURING VARICEAL BANDING, THE PHYSICIAN USED TWO 6 SHOOTER SAEED MULTI-BAND LIGATORS. IT WAS REPORTED THAT BOTH BANDERS MISFIRED AND LOST TENSION ON THE STRING [UNABLE TO EFFECTIVELY DEPLOY BANDS]. OUR ATTEMPTS TO COLLECT ADDITIONAL INFORMATION REGARDING PATIENT OUTCOME WERE UNSUCCESSFUL. WHILE THE COMPLAINANT DID NOT SPECIFY IF THE PATIENT EXPERIENCED ANY ADVERSE EFFECTS OR REQUIRED ADDITIONAL MEDICAL PROCEDURES DUE TO THIS EVENT, THE INFORMATION ABLE TO BE COLLECTED DOES NOT REASONABLY SUGGEST THE PATIENT WAS ADVERSELY IMPACTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19086338 ↗
  • MalfunctionFeb 2, 2024Product Quality Problem; Material Twisted/Bent; Product Quality Problem; Material Twisted/Bent

    DURING A LIGATION PROCEDURE, THE PHYSICIAN PREPARED TO USE A COOK 6 SHOOTER SAEED MULTI-BAND LIGATOR. IT WAS REPORTED THAT WHEN THEY OPENED THE PACKAGE, THE TRIGGER CORD WAS KNOTTED. THE LOCATION OF THE KNOT WOULD PREVENT DEPLOYMENT OF THE BANDS. ON (B)(6) 2024 THE FACILITY COMMUNICATED THAT THEY NEVER USED THE PRODUCT ON A PATIENT AND THE PRODUCT WAS NEVER REUSED PRIOR TO THIS OCCURRENCE. AS IT OCCURRED PRIOR TO PATIENT CONTACT, A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE AND THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18630864 ↗

FDA clearance and classification

The FDA regulates product code MND (ligator, esophageal) under 21 CFR 876.4400 as a Class II device. 21 CFR 876.4400 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Common questions about the Saeed G31917

Has the Saeed G31917 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002319172), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Saeed G31917?

The FDA's MAUDE database lists 118 adverse event reports for this device, received between Feb 19, 2016 and Jul 1, 2026. In those reports, the most common specific device problems are “Failure to Fire” (29), “Device Dislodged or Dislocated” (23) and “Material Too Soft/Flexible” (19); the most common patient problems are “Foreign Body In Patient” (11), “Hemorrhage/Blood Loss/Bleeding” (3) and “Gastrointestinal Hemorrhage” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Saeed G31917?

None of the reports are classified as deaths, 8 as injuries and 110 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Saeed G31917 FDA approved?

Devices of this type (product code MND) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the Saeed G31917, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Saeed, model G31917). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Ligator, esophageal devices

All Ligator, esophageal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page