SINGLE USE LIGATING DEVICE HX-400U-30 (REF N5382130)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of SINGLE USE LIGATING DEVICE (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Olympus single-use ligating device is used in the esophagus. Reports most often describe the device being difficult to remove or separate, handles or parts breaking, and loops that do not release. The database holds 244 reports, including no deaths and 124 injuries, with 120 malfunctions. There is 1 open recall, started in October 2025, because the loop may not release and could become fixed around patient tissue.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
239
Deaths reported
0
Injuries reported
120
Recalls naming it
1

Summary

The FDA's MAUDE database contains 239 adverse event reports that list this exact device (DI 04953170368615), received between Jul 16, 2018 and Sep 16, 2026 — classified as 120 injuries, 119 malfunctions.

The most frequently reported specific device problem is “Break” (75 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 3 reportsMar 2024: 9 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 5 reportsAug 2024: 9 reportsSep 2024: 6 reportsOct 2024: 4 reportsNov 2024: 5 reportsDec 2024: 6 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 7 reportsMay 2025: 5 reportsJun 2025: 4 reportsJul 2025: 9 reportsAug 2025: 2 reportsSep 2025: 11 reportsOct 2025: 1 reportsNov 2025: 6 reportsDec 2025: 11 reportsJan 2026: 6 reportsFeb 2026: 3 reportsMar 2026: 3 reportsApr 2026: 4 reportsMay 2026: 5 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 30, 2025Open, Classified

    Potential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.

    Z-0570-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break75
  • Difficult to Remove65
  • Difficult or Delayed Separation37
  • Separation Failure25
  • Mechanical Problem11
  • Detachment of Device or Device Component8
  • Entrapment of Device7
  • Physical Resistance/Sticking7

Patient problems

  • No Clinical Signs, Symptoms or Conditions185
  • No Consequences Or Impact To Patient19
  • Hemorrhage/Blood Loss/Bleeding13
  • Foreign Body In Patient8
  • No Known Impact Or Consequence To Patient5
  • Insufficient Information3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Laceration(s)2

Recent adverse event reports

The 25 most recent of 239 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Break; Separation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50797005 ↗
  • InjurySep 10, 2026Break; Detachment of Device or Device Component

    It was reported that during a therapeutic double balloon antegrade enterostomy, the Single use ligating device while being deployed inside the patient the wire snapped. As a result, the patient was taken for additional procedure into operation theater for Minilaparotomy, enterotomy, excision of polyp and release of entrapped detachable snare. The patient was operated and reported to be discharged.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50718473 ↗
  • MalfunctionAug 28, 2026Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501085 ↗
  • MalfunctionAug 17, 2026Separation Failure; Difficult to Open or Close; Mechanical Jam

    No additional information received from the customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50279389 ↗
  • MalfunctionAug 12, 2026Fracture; Difficult to Remove; Separation Failure

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50222543 ↗
  • MalfunctionJul 10, 2026Difficult or Delayed Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25783432 ↗
  • MalfunctionMay 15, 2026Difficult or Delayed Separation

    IT WAS REPORTED THE SINGLE USE LIGATING DEVICE LOOP'S RING DOESN'T COME OUT. THE ISSUE OCCURRED DURING A THERAPEUTIC EMR PROCEDURE. THE PROCEDURE WAS DELAYED, HOWEVER THE PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25187065 ↗
  • InjuryMay 12, 2026Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING A COLONOSCOPY, WHILE PERFORMING A POLYPECTOMY, THE LOOP OF THE LIGATING DEVICE WAS UNABLE TO DETACH, AND THE CANNULA/DEPLOYMENT DEVICE WAS UNABLE TO BE WITHDRAWN AFTER THE DEVICE WAS DEPLOYED. THE SLIDER LOCATED ON THE DEVICE HANDLE COULD NOT MOVE FORWARD OR BACKWARD. THE OPERATING WIRE HAD TO BE CUT TO REMOVE THE DEVICE FROM THE PATIENT/SCOPE. THE HOT SNARE WAS REQUIRED TO REMOVE THE STUCK POLYP LOOP. THE SCOPE WAS REMOVED AND REINSERTED. AN ENDOSCOPIC SUBMUCOSAL DISSECTION WAS PERFORMED TO COMPLETE THE PROCEDURE. IN ADDITION, IT WAS REPORTED THAT THE LESION WAS A 30 MM PEDUNCULATED POLYP ON A STALK LOCATED IN THE SIGMOID COLON. THE PROCEDURE WAS DELAYED FOR ABOUT 75 MINUTES, AND THE PATIENT WAS UNDER GENERAL ANESTHESIA WITH PROPOFOL. THE PATIENT REMAINED HEMODYNAMICALLY STABLE AND WAS ADMITTED TO THE HOSPITAL FOR OBSERVATION DUE TO THE LENGTH OF THE PROCEDURE TIME AND EXTENSIVE POLYPECTOMY TO ENSURE NO BLEEDING. THE PATIENT WAS DISCHARGED IN STABLE CONDITION. THIS REPORT IS 2 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25156699 ↗
  • InjuryMay 12, 2026Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING A COLONOSCOPY, WHILE PERFORMING A POLYPECTOMY, THE LOOP OF THE LIGATING DEVICE WAS UNABLE TO DETACH, AND THE CANNULA/DEPLOYMENT DEVICE WAS UNABLE TO BE WITHDRAWN AFTER THE DEVICE WAS DEPLOYED. THE SLIDER LOCATED ON THE DEVICE HANDLE COULD NOT MOVE FORWARD OR BACKWARD. THE OPERATING WIRE HAD TO BE CUT TO REMOVE THE DEVICE FROM THE PATIENT/SCOPE. THE HOT SNARE WAS REQUIRED TO REMOVE THE STUCK POLYP LOOP. THE SCOPE WAS REMOVED AND REINSERTED. AN ENDOSCOPIC SUBMUCOSAL DISSECTION WAS PERFORMED TO COMPLETE THE PROCEDURE. IN ADDITION, IT WAS REPORTED THAT THE LESION WAS A 30 MM PEDUNCULATED POLYP ON A STALK LOCATED IN THE SIGMOID COLON. THE PROCEDURE WAS DELAYED FOR ABOUT 75 MINUTES, AND THE PATIENT WAS UNDER GENERAL ANESTHESIA WITH PROPOFOL. THE PATIENT REMAINED HEMODYNAMICALLY STABLE AND WAS ADMITTED TO THE HOSPITAL FOR OBSERVATION DUE TO THE LENGTH OF THE PROCEDURE TIME AND EXTENSIVE POLYPECTOMY TO ENSURE NO BLEEDING. THE PATIENT WAS DISCHARGED IN STABLE CONDITION. THIS REPORT IS 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25156685 ↗
  • MalfunctionMay 6, 2026Break

    IT WAS REPORTED THE SINGLE USE LIGATING DEVICE'S CABLE BROKE, DISALLOWING RELEASE WHILST TIGHTENED AROUND A LESION. THIS OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095462 ↗
  • MalfunctionMay 1, 2026Material Deformation; Material Twisted/Bent; Activation Failure

    IT WAS REPORTED THE SINGLE USE LIGATING DEVICE DID NOT DEPLOY FROM DELIVERY SYSTEM ONCE PLACED ONTO POLYP. THE ISSUE OCCURRED DURING A THERAPEUTIC COLON POLYPECTOMY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056438 ↗
  • MalfunctionApr 20, 2026Difficult or Delayed Separation

    NO NEW ADDITIONAL INFORMATION WAS PROVIDED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24925110 ↗
  • InjuryApr 13, 2026Entrapment of Device; Difficult to Remove; Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING A DIAGNOSTIC COLONOSCOPY, THE SINGLE USE LIGATING DEVICE MAY NOT HAVE BEEN DEPLOYED CORRECTLY. UPON FURTHER FOLLOW UP, THE CUSTOMER INDICATED THAT THE DEVICE ANCHORED ITSELF AROUND THE PATIENT'S TISSUE. THE PROCEDURALIST COULD NOT REMOVE IT FOR A PROLONGED PERIOD BECAUSE THE SLIDER PART OF THE DEVICE DETACHED FROM THE OPERATION WIRE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH A DELAY GREATER THAN OR EQUAL TO 30 MINUTES. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24857501 ↗
  • InjuryApr 13, 2026Entrapment of Device; Difficult to Remove; Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING A DIAGNOSTIC COLONOSCOPY, THE SINGLE USE LIGATING DEVICE MAY NOT HAVE BEEN DEPLOYED CORRECTLY. UPON FURTHER FOLLOW UP, THE CUSTOMER INDICATED THAT THE DEVICE ANCHORED ITSELF AROUND THE PATIENT'S TISSUE. THE PROCEDURALIST COULD NOT REMOVE IT FOR A PROLONGED PERIOD BECAUSE THE SLIDER PART OF THE DEVICE DETACHED FROM THE OPERATION WIRE. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH A DELAY GREATER THAN OR EQUAL TO 30 MINUTES. THERE WERE NO REPORTS OF FURTHER PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED INCLUDING FURTHER PATIENT DETAILS, OUTCOME, AND METHOD OF DEVICE REMOVAL, BUT NO CUSTOMER RESPONSE HAS BEEN RECEIVED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24857457 ↗
  • InjuryApr 6, 2026Material Twisted/Bent; Difficult or Delayed Separation

    IT WAS REPORTED THAT DURING A THERAPEUTIC POLYPECTOMY, THE SINGLE USE LIGATING DEVICE FAILED TO DEPLOY. WHILE ATTEMPTING TO ATTACH AND CLOSE THE LOOP AROUND THE POLYP, THE DEVICE GOT STUCK RESULTING IN FAILURE TO RELEASE THE LOOP. THE CATHETER HAD TO BE CUT ON THE HANDLE END OF THE DEVICE AND THE SCOPE WAS REMOVED. THE SCOPE WAS THEN REINSERTED, AND THE LOOP WAS REMOVED USING A LOOP CUTTER. UPON INSPECTION AFTER THE PROCEDURE, IT WAS NOTED THAT THE WIRE ON THE CATHETER HANDLE WAS BENT IN A Z SHAPE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24796019 ↗
  • MalfunctionMar 20, 2026Physical Resistance/Sticking

    IT WAS REPORTED THE SINGLE USE LIGATING DEVICE SLIDER PRESENTED SOME RESISTANCE DURING USE, HOWEVER DEVICE DEPLOYED AND PRODUCT USE WAS SUCCESSFUL. THE ISSUE OCCURRED DURING AN UNKNOWN PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24651416 ↗
  • MalfunctionMar 11, 2026No Apparent Adverse Event

    IT WAS REPORTED THE SINGLE-USE LIGATING DEVICE'S DETATCHABLE LOOP WAS LOOSENING DURING A POLYPECTOMY PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24569776 ↗
  • InjuryMar 7, 2026Difficult to Remove; Separation Failure

    IT WAS REPORTED THAT DURING A THERAPEUTIC PROCEDURE UNDER SEDATION, THE SINGLE USE LIGATURE COULD NOT BE RELEASED. THE LIGATURE WAS CUT AND REMOVED. NO REPORTS OF ADVERSE PATIENT HEALTH CONSEQUENCES DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24539909 ↗
  • InjuryFeb 20, 2026Difficult or Delayed Separation

    IT WAS REPORTED DURING AN UNSPECIFIED PROCEDURE THE SINGLE USE LIGATING DEVICE FAILED. AFTER ATTEMPTING DEPLOYMENT OF THE SINGLE USE LIGATING DEVICE, THE MECHANISM WOULD NOT RELEASE RESULTING IN THE WIRE NEEDING TO BE CUT TO REMOVE FROM THE PATIENT. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405940 ↗
  • InjuryFeb 10, 2026Difficult or Delayed Separation

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24308645 ↗
  • InjuryFeb 2, 2026Difficult or Delayed Separation

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24236069 ↗
  • MalfunctionJan 23, 2026Difficult to Open or Close

    IT WAS REPORTED THE SINGLE USE LIGATING DEVICE POLYLOOP WAS DEFECTIVE WHEREIN THE POLYLOOP YELLOW SLIDER WAS NOT SLIDING PROPERLY TO CLOSE LOOP SEEMED TO GET STUCK ON SHEATH. THE ISSUE OCCURRED DURING AND UNKNOWN COLONOSCOPY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163144 ↗
  • MalfunctionJan 23, 2026Difficult to Open or Close

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163087 ↗
  • InjuryJan 15, 2026Firing Problem; Premature Separation

    THIS IS REPORT 1 OF 2: IT WAS REPORTED THAT DURING A PROCEDURE, THERE WAS AN ATTEMPT TO USE A LIGATING DEVICE. THE LIGATING DEVICE WAS PREMATURELY DEPLOYED AND FELL OUT OF THE LESION. THERE AS AN ATTEMPT TO RETRIEVE THE LOOP WITH FORCEPS THROUGH THE INSTRUMENT CHANNEL OF THE SCOPE BUT NOTHING WAS FOUND AND THE PROCEDURE WAS CLOSED. THE SCOPE WAS REPROCESSED AS USUAL. A FEW DAYS LATER, AFTER A CASE, WITH NO REPORTED ISSUES OF THE SCOPE, DURING REPROCESSING, THE LOOP DROPPED OUT OF THE INSTRUMENT CHANNEL AND DISPOSED OF. LEAKAGE TESTING WAS CONDUCTED AND NO ISSUE TO BE FOUND. THERE WAS NO HARM OR INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24082993 ↗
  • InjuryJan 9, 2026Difficult or Delayed Separation

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24021262 ↗

Other Ligator, Esophageal devices