Soft-Trans G26669Recalls & reported problems
Cook Incorporated · Soft-Trans Embryo Transfer Catheter · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002266698), received between Aug 10, 2017 and Aug 22, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material2
- Device Operates Differently Than Expected2
Patient problems
- No Patient Involvement2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2017Device Operates Differently Than Expected; Device Contamination with Chemical or Other Material; Device Operates Differently Than Expected; Device Contamination with Chemical or Other Material
THE COOK REPRESENTATIVE VISITED THE USER FACILITY TO PERFORM TESTING OF UNUSED SOFT-TRANS EMBRYO TRANSFER CATHETERS IN STOCK AT THE FACILITY IN AN EFFORT TO FIND A CATHETER THAT FLUSHES OUT YELLOW MEDIA FLUID. AFTER TESTING OVER 50 UNUSED CATHETERS, ONE CATHETER WAS FINALLY FOUND THAT THE MEDIA WAS FLUSHED THROUGH THE CATHETER AND CAME OUT YELLOW. THIS SAMPLE AND THE OTHER DEVICES USED DURING THE TEST WERE PACKAGED TOGETHER TO RETURN TO THE MANUFACTURER FOR FURTHER ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT DURING THIS TESTING. THE ISSUE OF MEDIA FLUID FLUSHING OUT "YELLOW" WAS PREVIOUSLY REPORTED BY THIS FACILITY. HOWEVER, SAMPLES WERE NOT PROVIDED FOR ANALYSIS. THE TWO PREVIOUS REPORTS CAN BE FOUND IN MFR. REPORT NUMBERS 1820334-2017-02298 AND 1820334-2017-02304.
Patient problems: No Patient Involvement
MAUDE report 6811082 ↗MalfunctionAug 10, 2017Device Operates Differently Than Expected; Device Contamination with Chemical or Other Material; Device Operates Differently Than Expected; Device Contamination with Chemical or Other Material
THE USER FACILITY REPORTED PRIOR TO ALL EMBRYO TRANSFERS THE CATHETERS ARE FLUSHED USING CLEAR VITROLIFE MEDIA. DURING THE LAST EMBRYO TRANSFER OF THE DAY, THE SOFT-TRANS EMBRYO TRANSFER CATHETER WAS FLUSHED WITH MEDIA BUT THE FLUID CAME OUT YELLOW. THE CATHETER WAS NOT USED. ALL REMAINING MEDIA FROM THE SAME LOT WAS CHECKED AND NO SIGN OF DISCOLORATION WAS FOUND. A NEW CATHETER WAS FLUSHED WITH A FRESH TUBE OF MEDIA AND THE LIQUID CAME OUT CLEAR. THE NEW CATHETER WAS USED FOR THE EMBRYO TRANSFER. ACCORDING TO THE REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. THE YELLOW EXPELLED LIQUID WAS VIEWED UNDER THE MICROSCOPE. NO SIGN OF GROWTH WAS OBSERVED.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 6785574 ↗
FDA clearance and classification
The FDA regulates product code MQF (catheter, assisted reproduction) under 21 CFR 884.6110 as a Class II device. 21 CFR 884.6110 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Soft-Trans G26669
Has the Soft-Trans G26669 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002266698), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Soft-Trans G26669?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 10, 2017 and Aug 22, 2017. In those reports, the most common specific device problems are “Device Contamination with Chemical or Other Material” (2) and “Device Operates Differently Than Expected” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Soft-Trans G26669?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Soft-Trans G26669, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Soft-Trans, model G26669). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ASSISTED REPRODUCTION devices
- Guardia G34783Cook Incorporated · CATHETER, ASSISTED REPRODUCTION61 reports
- Pivet G30457Cook Incorporated · CATHETER, ASSISTED REPRODUCTION29 reports
- Guardia G53007Cook Incorporated · CATHETER, ASSISTED REPRODUCTION9 reports
- Soft-Pass G18955Cook Incorporated · CATHETER, ASSISTED REPRODUCTION8 reports
- Guardia G24216Cook Incorporated · CATHETER, ASSISTED REPRODUCTION5 reports
- Sydney IVF G18740Cook Incorporated · CATHETER, ASSISTED REPRODUCTION3 reports
- Soft-Pass G30399Cook Incorporated · CATHETER, ASSISTED REPRODUCTION2 reports
- Soft-Pass G18658Cook Incorporated · CATHETER, ASSISTED REPRODUCTION2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page