Soft-Pass G30399Recalls & reported problems
Cook Incorporated · Soft-Pass Embryo Mock Transfer Catheter Set with Echogenic Tip · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002303997), received between Jun 25, 2014 and Apr 3, 2019 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Separation2
Patient problems
- Foreign body, removal of1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2019Material Separation; Material Separation
IT IS REPORTED WHILE PERFORMING A MEASUREMENT IN PREPARATION FOR EMBRYO TRANSFER USING AN EMBRYO TRANSFER CATHETER ASSIST DEVICE (RPNJ-SPPE-681710-ET-MC) THE ECHO TIP BAND BECAME CAUGHT IN THE OUTER CATHETER. PRIOR TO EMBRYO TRANSFER, THE PHYSICIAN REPORTS USING THE EMBRYO TRANSFER CATHETER ASSIST DEVICE (RPNJ-SPPE-681710-ET-MC) TO MEASURE WHERE THE EMBRYO WOULD BE PLACED. WHILE REMOVING THE CATHETER FROM THE OUTER SHEATH, THE PHYSICIAN REPORTED "THE ECHO TIP BAND BECAME CAUGHT IN THE OUTER CATHETER. A GREAT DEAL OF FORCE WAS REQUIRED TO SEPARATE THEM. METAL ECHO TIP DISLODGED FROM THE INNER CATHETER AND STUCK ONTO OUTER CATHETER " A SECOND SOFT-PASS ECHO TIP EMBRYO TRANSFER CATHETER SET (K-J-SPPE-681710-ET) WAS USED TO COMPLETE THE MEASUREMENT PRIOR TO EMBRYO TRANSFER WITHOUT DIFFICULTY. AFTER SUCCESSFUL MEASUREMENT, THE EMBRYO TRANSFER WAS COMPLETED USING A SOFT-PASS ECHO TIP EMBRYO TRANSFER CATHETER SET (K-J-SPPE-681710-ET). IT IS REPORTED THAT THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR REQUIRE ANY ADDITIONAL PROCEDURES AS A RESULT OF THIS ALLEGED PRODUCT MALFUNCTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8480402 ↗InjuryJun 25, 2014Material Separation
DURING A MOCK EMBRYO TRANSFER PROCEDURE, THE PHYSICIAN WAS ATTEMPTING TO REMOVE THE INNER CATHETER AND IT GOT STUCK ON THE OUTER CATHETER. WHEN THE PHYSICIAN REMOVED THE SOFT-PASS ECHOTIP EMBRYO TRANSFER CATHETER IT WAS NOTED THAT THE ECHO TIP BAND WAS NOT THERE. AN ULTRASOUND WAS DONE AND THE TIP WAS FOUND IN THE PATIENT. THE PATIENT HAD A SEPARATED BAND RETRIEVED VIA HYSTEROSCOPY ON (B)(6) 2014. IT WAS INDICATED THAT THEY HAD TO GO IN WITH A RIGID SCOPE TO REMOVE THE BAND AS IT WAS DIFFICULT TO RETRIEVE. THIS WAS THE SECOND IVF PROCEDURE FOR THE PATIENT. THERE WAS NO SHARP CERVICAL ANGLE. NO HARM WAS NOTED TO THE PATIENT.
Patient problems: Foreign body, removal of
MAUDE report 3918817 ↗
FDA clearance and classification
The FDA regulates product code MQF (catheter, assisted reproduction) under 21 CFR 884.6110 as a Class II device. 21 CFR 884.6110 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Soft-Pass G30399
Has the Soft-Pass G30399 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002303997), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Soft-Pass G30399?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 25, 2014 and Apr 3, 2019. In those reports, the most common specific device problem is “Material Separation” (2); the most common patient problem is “Foreign body, removal of” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Soft-Pass G30399?
None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Soft-Pass G30399, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Soft-Pass, model G30399). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, ASSISTED REPRODUCTION devices
- Guardia G34783Cook Incorporated · CATHETER, ASSISTED REPRODUCTION61 reports
- Pivet G30457Cook Incorporated · CATHETER, ASSISTED REPRODUCTION29 reports
- Guardia G53007Cook Incorporated · CATHETER, ASSISTED REPRODUCTION9 reports
- Soft-Pass G18955Cook Incorporated · CATHETER, ASSISTED REPRODUCTION8 reports
- Guardia G24216Cook Incorporated · CATHETER, ASSISTED REPRODUCTION5 reports
- Sydney IVF G18740Cook Incorporated · CATHETER, ASSISTED REPRODUCTION3 reports
- Soft-Pass G18658Cook Incorporated · CATHETER, ASSISTED REPRODUCTION2 reports
- Soft-Trans G26669Cook Incorporated · CATHETER, ASSISTED REPRODUCTION2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.