Skip to content

Tracer Metro G26862Recalls & reported problems

Cook Incorporated · Tracer Metro Direct Wire Guide · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00827002268623), received between May 27, 2016 and Jun 23, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated6
  • Detachment Of Device Component1
  • Device-Device Incompatibility1

Patient problems

  • No Consequences Or Impact To Patient3
  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE. IT WAS REPORTED THAT AFTER PLACEMENT OF THE WIRE GUIDE IN THE PANCREATIC DUCT, THE USER ATTEMPTED TO INSERT ANOTHER DEVICE FROM THE GUIDEWIRE END BUT FELT RESISTANCE. THE PHYSICIAN CONTINUED TO ATTEMPT TO INSERT, BUT THE GUIDEWIRE COATING NOTICED PEELED [WIRE GUIDE COATING DAMAGE], THEREFORE, THE PHYSICIAN DISCONTINUED TO USE THE WIRE GUIDE [LOSS OF WIRE GUIDE ACCESS]. THEY WERE REPLACED WITH ANOTHER MANUFACTURE'S DEVICES TO CONTINUE TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22309275 ↗
  • MalfunctionFeb 25, 2025Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE. IT WAS REPORTED THAT THE WIRE GUIDE WAS PLACED IN THE PANCREATIC DUCT, BUT THE WIRE GUIDE BECAME UNABLE TO PASS MIDWAY. USE OF DEVICE WAS DISCONTINUED [LOSS OF WIRE GUIDE ACCESS], AND THE PROCEDURE WAS COMPLETED USING OTHER PRODUCTS. PER EVALUATION OF PRODUCT RETURN ON 4 FEB 2025, DAMAGE TO THE PROXIMAL WIRE GUIDE COATING WAS NOTED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21462205 ↗
  • MalfunctionSep 14, 2018Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE. THE WIRE STARTED TO UNRAVEL AND COME OFF TOWARDS THE PROXIMAL END WHERE THE MARKINGS ARE [COATING DAMAGE]. THEY REMOVED THE WIRE AND USED ANOTHER DEVICE TO COMPLETE THE CASE SUCCESSFULLY. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7876552 ↗
  • InjuryNov 3, 2017Detachment Of Device Component; Detachment Of Device Component

    THE FOLLOWING INFORMATION WAS PROVIDED TO COOK VIA MEDWATCH REPORT MW5072361: "GUIDEWIRE PASSED THROUGH THE SPHINCTEROTOME, DIFFICULT CANNULATION. [THE] WIRE WAS REMOVED FROM THE SPHINCTEROTOME AND IT WAS NOTED THAT THE WIRE WAS SHEARED. IT WAS NOTED ON FLUOROSCOPY THAT PART OF THE WIRE WAS PRESENT IN THE PATIENT. [THE USER] ATTEMPTED TO REMOVE THE WIRE UNSUCCESSFULLY. ANOTHER GUIDEWIRE WAS USED TO PERFORM THE PROCEDURE WITH A NEW SPHINCTEROTOME." PER THE RECEIVED MEDWATCH REPORT, STATES THAT A SERIOUS INJURY OCCURRED AND ADVERSE EVENT WAS CHECKED "Y". THERE WERE NO OTHER DETAILS INCLUDED ON THE MEDWATCH REPORT. MULTIPLE ATTEMPTS TO DETERMINE THE CUSTOMER OR OTHER DETAILS RELATED TO THIS EVENT WERE COMPLETED WITH NO RESULTS.

    Patient problems: Foreign Body In Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7001253 ↗
  • MalfunctionJan 13, 2017Peeled/Delaminated; Device-Device Incompatibility; Peeled/Delaminated; Device-Device Incompatibility

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE (METII-21-480). AT THE BEGINNING OF THE CASE, THE DOCTOR CANNULATED WITH ANOTHER MANUFACTURERS CANNULA [RECOMMENDED WIRE GUIDE DIAMETER TO BE USED WITH THIS CANNULA IS .018 IN]. THE ASSISTANT ADVANCED THE METII-21-480 [.021 IN] THROUGH THE PROXIMAL END OF THE CATHETER. THE ASSISTANT COULD FEEL THE WIRE STICKING A LITTLE BIT. WHEN SHE REMOVED THE WIRE GUIDE SHE COULD SEE THE COATING OF THE WIRE WAS SEPARATING FROM THE WIRE AND THE TIP. THEY REMOVED IT COMPLETELY AND GOT A NEW METII-21-480 AND THE REST OF THE CASE WENT FINE AND WAS FINISHED SUCCESSFULLY WITH THE SECOND WIRE GUIDE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6249382 ↗
  • MalfunctionNov 17, 2016Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE. THE PHYSICIAN WAS USING THE WIRE GUIDE IN THE SMALL INTESTINE NEAR THE BILE DUCT. THE COATING OF THE WIRE GUIDE SHREDDED OFF WHEN HE WAS PULLING THE WIRE GUIDE BACK INTO THE ENDOSCOPE. SOME OF THE COATING STAYED IN THE PATIENT'S SMALL INTESTINE. THE PHYSICIAN SAID IT WAS NOT HARMFUL AND WILL COME OUT IN THE PATIENT'S STOOL.

    Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient

    MAUDE report 6109135 ↗
  • MalfunctionMay 27, 2016Peeled/Delaminated; Peeled/Delaminated

    DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRACER METRO DIRECT WIRE GUIDE. THE "WIRE COATING WAS PEELING BACK."

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5684877 ↗

FDA clearance and classification

The FDA regulates product code OCY (endoscopic guidewire, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Tracer Metro G26862

Has the Tracer Metro G26862 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00827002268623), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Tracer Metro G26862?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between May 27, 2016 and Jun 23, 2025. In those reports, the most common specific device problems are “Peeled/Delaminated” (6), “Detachment Of Device Component” (1) and “Device-Device Incompatibility” (1); the most common patient problem is “Foreign Body In Patient” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Tracer Metro G26862?

None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Tracer Metro G26862 FDA approved?

Devices of this type (product code OCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.

Who makes the Tracer Metro G26862, and is it still sold?

The FDA's device database lists Cook Incorporated as the company responsible for it (brand Tracer Metro, model G26862). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Endoscopic guidewire, gastroenterology-urology devices

All Endoscopic guidewire, gastroenterology-urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page