HiWire G30470Recalls & reported problems
Cook Incorporated · HiWire Nitinol Core Wire Guide Straight with Hydrophilic Coating Stiff Shaft and 3cm Flexible Tip · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002304703), received between Aug 23, 2018 and Oct 23, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated2
- Material Separation1
Patient problems
- Foreign Body In Patient2
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2024Material Separation; Material Separation
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 20516220 ↗InjuryJul 20, 2022Peeled/Delaminated; Peeled/Delaminated
AS REPORTED, THE PHYSICIAN WAS MET WITH RESISTANCE WHEN ATTEMPTING TO ADVANCE A HIWIRE NITINOL HYDROPHILIC WIRE GUIDE THROUGH A NEEDLE SEMI-RIGID URETEROSCOPE WHILE GAINING ACCESS TO THE URETER. THE PATIENT HAD RADIATION PRIOR TO THE PROCEDURE, AND THEY GAINED ACCESS BY WIRING IT UP THROUGH THE KIDNEY. THE PATIENT WAS NOTED TO HAVE SCARRED ANATOMY. A CATHETER WAS PLACED OVER THE WIRE, AND IT WAS FOUND THAT THE WIRE HAD SHEARED OFF. THE PHYSICIAN BELIEVES THEY REMOVED ALL OF THE WIRE (DEVICE USED UNKNOWN). A SECOND WIRE KINKED AND BEGAN TO SEPARATE, BUT IT DID NOT SHEAR. IT IS POSSIBLE THAT THE SCOPE OR ANGLE COULD HAVE CONTRIBUTED TO THE ISSUES. A SMALLER DIAMETER WIRE WORKED SUCCESSFULLY TO COMPLETE THE PROCEDURE. NO UNINTENDED SECTION OF THE DEVICE REMAINED INSIDE OF THE PATIENT'S BODY. NO ADDITIONAL PROCEDURES WERE REQUIRED DUE TO THE OCCURRENCE. NO ADVERSE EFFECTS WERE REPORTED DUE TO THE OCCURRENCE. DEVICES WERE RETURNED ON (B)(6) 2022: ATTACHED PHOTO'S SHOW COATING SEPARATION ON BOTH WIRE GUIDES. FOR THE FIRST WIRE GUIDE THE COATING SEPARATED AND CAME COMPLETELY OFF THE WIRE. FOR THE 2ND DEVICE THE COATING SEPARATED, BUT DID NOT COMPLETELY COME OFF THE WIRE. KINKED WIRE WAS N
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 15064204 ↗MalfunctionAug 23, 2018Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED, DURING AN UNKNOWN PROCEDURE, THE TEFLON ON THE HIWIRE NITINOL HYDROPHILIC WIRE GUIDE SPLIT/PEELED. ADDITIONAL INFORMATION HAS BEEN REQUESTED REGARDING THE PATIENT AND THE EVENT. AT THE TIME OF THIS REPORT, NO FURTHER INFORMATION HAS BEEN PROVIDED. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE ALLEGED MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7809360 ↗
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.