Hercules G31927Recalls & reported problems

Cook Incorporated · Hercules 3 Stage Balloon Esophageal · Class 2

Quick summary

This Cook Hercules balloon dilator is used to widen narrowed areas of the esophagus during endoscopy. Most of the 74 reports describe the balloon bursting or leaking while inflated, sometimes leaving a piece of the device behind in the patient, though reports generally say no additional procedure was needed. There are 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
74
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 74 adverse event reports that list this exact device (DI 00827002319271), received between Apr 26, 2016 and Sep 29, 2026 — classified as 4 injuries, 70 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (35 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel35
  • Use of Device Problem19
  • Fluid/Blood Leak17
  • Leak/Splash16
  • Material Puncture/Hole12
  • Hole In Material7
  • Improper or Incorrect Procedure or Method5
  • Material Rupture4

Patient problems

  • No Consequences Or Impact To Patient58
  • No Clinical Signs, Symptoms or Conditions11
  • Foreign Body In Patient6
  • No Known Impact Or Consequence To Patient5
  • Tissue Damage1

Recent adverse event reports

The 25 most recent of 74 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938347 ↗
  • MalfunctionJul 27, 2026Material Rupture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50039261 ↗
  • MalfunctionFeb 24, 2026Fluid/Blood Leak

    DURING AN ENDOSCOPIC PROCEDURE IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] WHEN THEY GOT TO 20 [ATM], THE BALLOON BURST AT THE DISTAL END. THEY FINISHED THE PROCEDURE BY REMOVING THE SCOPE AND BALLOON AND USED ANOTHER DEVICE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24433394 ↗
  • MalfunctionFeb 24, 2026Fluid/Blood Leak

    DURING AN ESOPHAGEAL DILATION, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431820 ↗
  • MalfunctionDec 12, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN ESOPHAGEAL DILATION, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON BROKE AT 19 MM, STAGE 2 [LEAKAGE]. NO BALLOON MATERIAL SEPARATED FROM THE DEVICE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE OF THE SAME TYPE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23788727 ↗
  • MalfunctionAug 26, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED AT ABOUT 4 ATM WHEN INFLATED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22897855 ↗
  • MalfunctionAug 26, 2025Material Rupture; Material Rupture

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED AT ABOUT 4 ATM WHEN INFLATED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22897775 ↗
  • MalfunctionAug 26, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN ESOPHAGEAL ENDOSCOPY, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED AT ABOUT 4 ATM WHEN INFLATED [SUBJECT OF REPORT]. DEVICE WAS EVALUATED ON 18AUG2025. APPROXIMATELY 5.5 CM OF THE BALLOON/DISTAL HAS BEEN CUT FROM THE DEVICE AND WAS NOT INCLUDED IN THE RETURN. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22897639 ↗
  • MalfunctionAug 18, 2025Fluid/Blood Leak; Fluid/Blood Leak

    DURING AN UPPER ENDOSCOPY IN THE ESOPHAGUS, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] WHILE INFLATING THE BALLOON, BEFORE IT REACHED 6 ATM, THE BALLOON STARTED TO LEAK. THE BALLOON WAS REMOVED AND ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO FURTHER COMPLICATIONS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22828034 ↗
  • MalfunctionMay 9, 2025Loss of or Failure to Bond; Loss of or Failure to Bond

    DURING AN ESOPHAGEAL DILATION, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21995795 ↗
  • MalfunctionMar 11, 2025Material Rupture; Material Rupture

    DURING AN UNKNOWN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. IT WAS REPORTED [THAT] THE BALLOON POPPED DURING DILATION WHILE IN THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21574183 ↗
  • MalfunctionSep 19, 2019Fluid/Blood Leak; Material Puncture/Hole; Use of Device Problem; Fluid/Blood Leak; Material Puncture/Hole; Use of Device Problem

    DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON WAS INTRODUCED DOWN THE ENDOSCOPE FINE. UPON THE SECOND STAGE OF INFLATION (4 ATM TO 6 ATM), THE BALLOON BURST INSIDE OF THE PATIENT. THE DEVICE WAS REMOVED FULLY INTACT, NOTHING DETACHED IN THE PATIENT. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9088257 ↗
  • MalfunctionSep 16, 2019Burst Container or Vessel; Use of Device Problem; Burst Container or Vessel; Use of Device Problem

    DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON BURST INSIDE THE PATIENT. THE CUSTOMER SAYS THAT THEY DID NOT OVER-INFLATE. THEY WERE USING ANOTHER MANUFACTURERS INFLATION DEVICE. NO PART OF THE BALLOON SEPARATED INSIDE THE PATIENT SO THEY DID NOT NEED TO RETRIEVE IT. THE FOLLOWING WAS RECEIVED ON (B)(6) 2019 PER MEDWATCH MW5089773: "PT IN MPU ROOM 7 FOR EGD PROCEDURE. WE HAVE SWITCHED OUR PRODUCT TO COOK FROM BOSTON SCIENTIFIC. WE USED THE COOK BALLOON TODAY, THE 18-20 MM THROUGH SCOPE DILATOR (HURRICANE IS THE SAME I BELIEVE). INFLATED 18MM IN STRICTURE AND HELD FOR 60 SEC THEN WENT UP TO 19MM AND THEN BALLOON RUPTURED WHEN WE WENT UP TO 20 MM MARK AFTER ABOUT 10 SEC. HAD TO REPEAT THE DILATION, CUT THE BALLOON AND REDO. THIS ADDED EXTRA TIME AS WELL AS EXTRA COST WHEN GOING THROUGH 2 BALLOONS INSTEAD OF 1. DURING PROCEDURE BALLOON WAS INFLATED AS DIRECTED BY PHYSICIAN. STAFF APPLIED NEGATIVE PRESSURE TO BALLOON AS INSTRUCTED, INFLATED TO DESIRED AND RECOMMENDED VOLUME. BALLOON WAS INITIATED TO 18 (4 PRESSURE), 19 (5 PRESSURE), 20 (6 PRESSURE). INFLATION WAS WITH THE BIG60 DEVICE AND WE INFLATED TO ATMOSPHERIC PRESSURE. N

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9071457 ↗
  • MalfunctionSep 13, 2019Burst Container or Vessel; Burst Container or Vessel

    AS INITIALLY REPORTED TO CUSTOMER RELATIONS: DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON WAS INFLATED HALF WAY AND THE BALLOON BURST. NOTHING DETACHED IN THE PATIENT AND THE DEVICE WAS REMOVED INTACT. ANOTHER BALLOON WAS USED TO CONTINUE THE PROCEDURE WITH NO FURTHER COMPLICATIONS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9046404 ↗
  • MalfunctionSep 6, 2019Burst Container or Vessel; Burst Container or Vessel

    DURING AN ESOPHAGEAL DILATION PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THERE WAS A SITUATION DURING DILATION PROCEDURE WHERE THE BALLOON BURST WHILE IN SITU [SIC] IN THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8972470 ↗
  • MalfunctionJul 26, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    DURING AN ESOPHAGEAL DILATION, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON WAS NOT INFLATING. THERE WAS NO REPORTABLE INFORMATION AT THIS TIME. THE DEVICE WAS EVALUATED ON 03-JUL-2019 AND A LEAK WAS SEEN FROM A HOLE IN THE MIDDLE OF THE BALLOON. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8831354 ↗
  • InjuryJul 10, 2019Burst Container or Vessel; Fluid/Blood Leak; Material Separation; Burst Container or Vessel; Fluid/Blood Leak; Material Separation

    DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON BURST DURING THE PROCEDURE. THE DEVICE WAS EVALUATED ON (B)(6) 2019. THE BALLOON MATERIAL HAD DETACHED FROM THE REST OF THE BALLOON. THE DETACHED PORTION INCLUDED THE BALLOON TIP. THE CANNULA USED TO ASSIST WITH DEFLATION HAD ALSO DETACHED AND WAS RETURNED KINKED IN ONE LOCATION. ADDITIONAL INFORMATION WAS RECEIVED ON (B)(6) 2019: THE USER INFLATED THE BALLOON TO 18 MM AND THEN ATTEMPTED TO INFLATE THE BALLOON TO 20 MM WHEN IT POPPED. AS THE USER WAS ATTEMPTING TO REMOVE THE BALLOON FROM THE ENDOSCOPE IN THE PATIENT, THE BALLOON DETACHED. AFTER THEY REMOVED THE DEVICE FROM THE ENDOSCOPE, THEY THEN REMOVED THE DETACHED PORTION FROM THE PATIENT WITH A SNARE. THERE WERE NO ADDITIONAL PROCEDURES REQUIRED AND THERE WERE NO ADVERSE EFFECTS TO THE PATIENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT REQUIRED REMOVAL OF THE DETACHED PORTION OF THE BALLOON WITH A SNARE DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: Foreign Body In Patient

    MAUDE report 8777885 ↗
  • MalfunctionJul 1, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE CUSTOMER HAD A BALLOON THAT WOULD NOT INFLATE. UPON INSPECTION, OUTSIDE THE PATIENT, THEY NOTED A HOLE IN THE SIDE OF THE BALLOON. THERE WAS NO HARM CAUSED TO THE PATIENT AND A DIFFERENT BALLOON WAS UTILIZED TO PROVIDE TREATMENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8748593 ↗
  • MalfunctionMay 10, 2019Burst Container or Vessel; Burst Container or Vessel

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE DEVICE WAS USED SUCCESSFULLY; HOWEVER, WHEN REMOVING THE BALLOON, THE BALLOON WOULD NOT COME OUT OF THE ENDOSCOPE. THE ENDOSCOPE AND THE BALLOON WERE REMOVED AS A UNIT. THE BALLOON WAS EVENTUALLY REMOVED, OUTSIDE OF THE PATIENT'S BODY, FROM THE ENDOSCOPE. THE BALLOON WAS INFLATED AS A TEST OUTSIDE OF THE PATIENT'S BODY AND THE BALLOON BURST [SUBJECT OF THIS REPORT]. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8601334 ↗
  • MalfunctionApr 30, 2019Material Puncture/Hole; Material Puncture/Hole

    DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE DEVICE DID NOT PERFORM PROPERLY. [THE DEVICE] HAD THREE (3) HOLES IN THE BALLOON. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8566528 ↗
  • MalfunctionApr 5, 2019Burst Container or Vessel; Use of Device Problem; Burst Container or Vessel; Use of Device Problem

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. AFTER REACHING THE FINAL DIAMETER (20 MM) AND LEVEL OF PRESSURE (6 ATM), DURING DILATION, THE BALLOON SUDDENLY BURST. THE DEVICE WAS TAKEN OUT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8487218 ↗
  • MalfunctionMar 7, 2019Burst Container or Vessel; Fluid/Blood Leak; Improper or Incorrect Procedure or Method; Burst Container or Vessel; Fluid/Blood Leak; Improper or Incorrect Procedure or Method

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH ESOPHAGEAL DILATION PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. ON THE SECOND INFLATION, THE BALLOON BURST. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8398688 ↗
  • MalfunctionMar 1, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON LEAKED TOWARD THE DISTAL TIP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8382631 ↗
  • MalfunctionMar 1, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON LEAKED TOWARD THE DISTAL TIP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8382629 ↗
  • MalfunctionMar 1, 2019Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole

    DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. THE BALLOON LEAKED TOWARD THE DISTAL TIP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8382628 ↗

Other Dilator, esophageal devices

All Dilator, esophageal devices →