Skip to content

Hercules G31928Recalls & reported problems

Cook Incorporated · Hercules 3 Stage Balloon Esophageal · Class 2

All models of Hercules (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002319288), received between Sep 15, 2016 and Sep 15, 2016 — classified as 1 malfunction.

The most frequently reported specific device problem is “Hole In Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Hole In Material1
  • Use of Device Problem1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2016Hole In Material; Use of Device Problem; Hole In Material; Use of Device Problem

    DURING AN ENDOSCOPY PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL. AS REPORTED TO CUSTOMER RELATIONS: "THE CUSTOMER ATTEMPTED TO ADVANCE A COOK HERCULES 3 STAGE BALLOON ESOPHAGEAL DOWN THE ENDOSCOPE AND THE DEVICE WOULD NOT GO [ADVANCE]. THEN, THE CUSTOMER USED A HERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC AND THE BALLOON BURST. SEE MDR REPORT NUMBER 1037905-2016-00351. THERE WAS NO HARM TO THE PATIENT AND NO ADDITIONAL PROCEDURES WERE REQUIRED DUE TO EITHER OCCURRENCE." THE DEVICE THAT WAS REPORTED TO BE UNABLE TO BE ADVANCED, WAS RECEIVED FOR EVALUATION ON 08/24/2016. AT THIS TIME, WATER WAS NOTICED TO BE LEAKING FROM A PINHOLE IN THE BALLOON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5953686 ↗

Other Dilator, esophageal devices

All Dilator, esophageal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page