Flexor Parallel G34872Recalls & reported problems

Cook Incorporated · Flexor, Parallel Rapid Release Ureteral Access Sheath with AQ, Hydrophilic Coating · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002348721), received between Jul 19, 2022 and Feb 21, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance1
  • Unsealed Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 21, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED, THE NURSE WENT TO OPEN THE PACKAGE OF A FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATORS AND NOTICED THE SEAL WAS OPEN. THE DEVICE WAS FOUND PRIOR TO PATIENT CONTACT. ANOTHER SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. A PHOTO OF WHAT APPEARS TO SHOW UNSEALED DEVICE PACKAGING WAS ATTACHED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21439143 ↗
  • MalfunctionJul 19, 2022Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT A FLEXOR PARALLEL URETERAL ACCESS SHEATH AND DILATOR PERFORATED A PATIENTS BLADDER. THE USER REPORTED THE TIP OF THE INNER STYLET SHEATH SPLIT EARLY AND THE WIRE CAME OFF. THE USER ALSO REPORTED THAT THE DEVICES WIRE RELEASE FUNCTION IS VERY ABRUPT AND RELEASES BEFORE THE STYLET IS PULLED FROM THE SHEATH FOR PROPER DEPLOYMENT. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ADDITIONAL INFORMATION REGARDING EVENT DETAILS, PATIENT ANATOMY AND OUTCOME HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THIS TIME.

    Patient problems: Perforation

    MAUDE report 15057441 ↗

Other endoscopic access overtube, gastroenterology-urology devices