Acrobat G47610Recalls & reported problems
Cook Incorporated · Acrobat 2 Calibrated Tip Wire Guide · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002476103), received between Jun 29, 2017 and Apr 1, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Use of Device Problem2
- Device-Device Incompatibility1
- Improper or Incorrect Procedure or Method1
- Migration or Expulsion of Device1
- Peeled/Delaminated1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2019Peeled/Delaminated; Use of Device Problem; Improper or Incorrect Procedure or Method; Peeled/Delaminated; Use of Device Problem; Improper or Incorrect Procedure or Method
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK ACROBAT® 2 CALIBRATED TIP WIRE GUIDE. [IT WAS] DIFFICULT TO PASS THROUGH [THE CATHETER] AND THE END PEELED OFF THE WIRE GUIDE. THE PHYSICIAN WAS USING A COOK CLASSIC ERCP CATHETER [ERCP-1-BT]. FINALLY THIS ONE [WIRE GUIDE] WAS PEELED [COATING DAMAGE]. THE PHYSICIAN USED ANOTHER WIRE GUIDE FROM ANOTHER MANUFACTURER AND IT WORKED WELL. ADDITIONAL INFORMATION WAS RECEIVED ON 19-MAR-2019 INDICATING THAT THE DAMAGE ON THE WIRE GUIDE WAS AT THE PATIENT END, 5 CENTIMETERS FROM THE TIP. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8471382 ↗MalfunctionJun 29, 2017Migration or Expulsion of Device; Use of Device Problem; Device-Device Incompatibility; Migration or Expulsion of Device; Use of Device Problem; Device-Device Incompatibility
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK ACROBAT 2 CALIBRATED TIP WIRE GUIDE. THE PHYSICIAN GAINED ACCESS, HAD THE WIRE IN PLACE AND WHEN HE WENT TO REMOVE THE COOK ULTRA TAPER SPHINCTEROTOME (UTS-20), THE WIRE WOULD NOT STAY IN PLACE. IT ALL CAME BACK TOGETHER AND HE LOST ACCESS. (THE WIRE WAS USED WITH A COOK NON-COMPATIBLE DEVICE. IT WAS USED WITH A (UTS-20). THE PHYSICIAN INSISTED ON TRYING IT AND REP DID NOT ENCOURAGE THE USE OF THE DEVICE.) THE UTS-20 AND WIRE STAYED OUT. THE PHYSICIAN THEN GAINED ACCESS WITH ANOTHER MANUFACTURER'S [.035 DEVICE] AND ADVANCED WIRE THROUGH THAT AND GAINED ACCESS WITHOUT PROBLEM. HE WENT TO REMOVE THE BALLOON AND KEEP WIRE ACCESS; HOWEVER, THE [OTHER] WIRE CAME OUT AGAIN. THEY COULD NOT GET THE WIRE TO STAY IN PLACE AND IT GRIPPED WITH THE BALLOON. THE WIRE WAS KEPT VERY WET AND THE RACETRACK WAS KEPT WET AS WELL.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6678498 ↗
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page