Acrobat G47615Recalls & reported problems
Cook Incorporated · Acrobat 2 Calibrated Tip Wire Guide · Class 2
All models of Acrobat (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002476158), received between Jun 16, 2017 and Mar 31, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated4
- Improper or Incorrect Procedure or Method1
- Mechanical Problem1
Patient problems
- No Consequences Or Impact To Patient4
- Perforation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 31, 2026Mechanical Problem
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE, THE PHYSICIAN USED A COOK ACROBAT 2 CALIBRATED TIP WIRE GUIDE. DURING A DISCUSSION THE PHYSICIAN REGARDING THE ACROBAT 2 DEVICE AND THE COOK SURETOME (CAPTURED IN MDR 1037905-2026-00106), THE PHYSICIAN DISCUSSED MANY ASPECTS OF ERCP TECHNIQUES AND CONSIDERATIONS. HE MENTIONED NEAR THE END OF THE VISIT AN INSTANCE IN WHICH HE WAS USING A WIRE GUIDE, WHICH PERFORATED THE LIVER DUCT. HE SAID IT WAS STIFFER THAN HE EXPECTED AND IT JUST "SAILED OUT". HE ALSO SAID IT DIDN'T "KNUCKLE" LIKE HE EXPECTED (WHICH WE INTERPRET AS ROTATING OR ORIENTING THE TIP). UPON FURTHER DISCUSSION, IT WAS DETERMINED THAT THIS EVENT WAS PREVIOUSLY REPORTED UNDER MDR 1037905-2026-00106 FOR THE COOK SURETOME. THIS MDR IS BEING SENT TO CAPTURE THE COOK ACROBAT 2, SINCE IT WAS NOT ORIGINALLY REPORTED BY THE INITIAL REPORTER AND IT IS UNABLE TO BE DETERMINED WHICH DEVICE CAUSED OR CONTRIBUTED. THE PATIENT WILL REQUIRE A SECOND ERCP TO RETRIEVE THE PLASTIC STENT THAT WAS PLACED. THERE WAS AN INTRAHEPATIC DUCTAL PERFORATION NOTED ON FLUORO WHERE THE WIRE TERMINATED.
Patient problems: Perforation
MAUDE report 24744617 ↗MalfunctionApr 13, 2018Peeled/Delaminated; Peeled/Delaminated
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED FIVE (5) TOTAL ACROBAT 2 CALIBRATED TIP WIRE GUIDES FROM LOT W3994086 [SUBJECT OF THIS REPORT] AND W3926765 [SEE RELATED MDR 1037905-2018-00144]. THE QUANTITY FROM EACH LOT IS UNKNOWN. THE WIRE WAS INSERTED AND THE PLASTIC COATING WAS UNRAVELING AND GETTING STUCK WITHIN THE ENDOSCOPE. THE USER TRIED NUMEROUS TIMES AND THE SAME THING WAS HAPPENING. THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED ON 03/29/18: THE COATING ON FIVE (5) OF THEIR WIRES SHREDDED FROM THE 5 CM MARK UPWARDS. THEY KEPT TWO (2) OUT OF THE FIVE (5) TO BE RETURNED. THE DOCTOR COULD NOT CONFIRM THAT THE TIP WAS WET BEFORE ENTERING THE SPHINCTEROTOME AND THE NURSES WERE NOT AVAILABLE TODAY TO SEE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7428735 ↗MalfunctionAug 10, 2017Peeled/Delaminated; Peeled/Delaminated
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK ACROBAT 2 CALIBRATED TIP WIRE GUIDE. THE WIRE STRIPPED WHEN PLACING A LONG COOK 7-15 ZIMMON PANCREATIC STENT IN THE BILE DUCT. THE PHYSICIAN USED EXTREME FORCE TO GET THE WIRE INSIDE STENT WHEN RELEASING THE STENT. ANOTHER MANUFACTURER'S EXTRACTION BALLOON WAS USED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6785096 ↗MalfunctionJun 30, 2017Peeled/Delaminated; Improper or Incorrect Procedure or Method; Peeled/Delaminated; Improper or Incorrect Procedure or Method
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK ACROBAT 2 CALIBRATED TIP WIRE GUIDE (AWG2-35-260-A). DURING AN ERCP WITH DIFFICULT ANATOMY, THE PHYSICIAN WAS UTILIZING THE AWG2-35-260-A THROUGH THE COOK FUSION OASIS ONE ACTION STENT INTRODUCTION SYSTEM FS-OA-8.5. THE PHYSICIAN REQUESTED THE NURSE FOR BACK TENSION. AS THE WIRE GUIDE WAS STILL ENGAGED, THE DISTRICT MANAGER CAUTIONED THE USER OF THE INSTRUCTIONS FOR USE FOR THE FS-OA-8.5. THE GUIDEWIRE BECAME STUCK BETWEEN THE FS-OA-8.5'S GUIDE CATHETER AND PUSHER. ONCE THE DEVICES WERE REMOVED, IT WAS NOTED THAT THE COATING ON THE WIRE HAD ACCORDIONED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6680699 ↗MalfunctionJun 16, 2017Peeled/Delaminated; Peeled/Delaminated
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK ACROBAT 2 CALIBRATED TIP WIRE GUIDE. THE WIRE WAS DIFFICULT TO REMOVE THROUGH A DEVICE AND THE NURSE TRIED TO PULL WITH THE HEMOSTAT. THE COATING ON THE WIRE STRIPPED. THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED ON 5/30/17 FROM THE COOK AREA REPRESENTATIVE: THERE WAS DAMAGE TO THE WIRE GUIDE AT BOTH ENDS OF THE WIRE. I DON'T KNOW EXACTLY HOW MANY CENTIMETERS AT PATIENT END AND I WOULD ESTIMATE APPROXIMATELY 5 CM. IT LOOKED ACCORDION LIKE TO MY EYE WHICH DISTORTS AN ESTIMATE. THE FOLLOWING ADDITIONAL INFORMATION WAS PROVIDED ON 6/7/17 FROM A PHONE CALL WITH THE COOK AREA REPRESENTATIVE: AFTER THE NURSE PULLED OUT THE WIRE GUIDE WITH THE HEMOSTAT, THE AREA REPRESENTATIVE NOTICED THE DAMAGE ON THE PROXIMAL END OF THE WIRE GUIDE. SHE THEN PUT THE WIRE GUIDE IN THE POUCH TO BE RETURNED TO COOK. A LITTLE LATER, WHILE THE PHYSICIAN WAS DOING BIOPSIES, THE PHYSICIAN HANDED HER [THE AREA REPRESENTATIVE] A PIECE OF WIRE GUIDE COATING THAT SEEMED TO HAVE BEEN RETRIEVED FROM THE BIOPSY PROCESS [THE WIRE GUIDE COATING WAS INADVERTENTLY RETRIEVED WHILE TAKING A BIOPSY]. THIS PIECE WAS ALSO PLACED IN
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6647268 ↗
FDA clearance and classification
The FDA regulates product code OCY (endoscopic guidewire, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Acrobat G47615
Has the Acrobat G47615 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00827002476158), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Acrobat G47615?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jun 16, 2017 and Mar 31, 2026. In those reports, the most common specific device problems are “Peeled/Delaminated” (4), “Improper or Incorrect Procedure or Method” (1) and “Mechanical Problem” (1); the most common patient problem is “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Acrobat G47615?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Acrobat G47615 FDA approved?
Devices of this type (product code OCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cook Incorporated.
Who makes the Acrobat G47615, and is it still sold?
The FDA's device database lists Cook Incorporated as the company responsible for it (brand Acrobat, model G47615). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology81 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page