Micropuncture G47947Recalls & reported problems
Cook Incorporated · Micropuncture Introducer Set Silhouette Transitionless With Nitinol Wire Guide · Class 2
All models of Micropuncture (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002479470), received between Apr 23, 2018 and Oct 19, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation1
- Unraveled Material1
Patient problems
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2020Unraveled Material; Unraveled Material
AS REPORTED, DURING A PRE-FONTAN CARDIAC CATHETERIZATION THE WIRE GUIDE OF A MICROPUNCTURE TRANSITIONLESS PEDIATRIC ACCESS SET UNRAVELED. ACCESS WAS GAINED FROM THE RIGHT INTERNAL JUGULAR VEIN WITH THE SUPPLIED NEEDLE WITH GOOD BACKFLOW FROM THE VESSEL AND THE 0.018 WIRE WAS SMOOTHLY INTRODUCED. AFTER REACHING THE "EXPECTED DEPTH" THE USER ATTEMPTED TO WITHDRAW THE NEEDLE TO USE THE WIRE TO SELDINGER THE DILATOR. THE USER NOTED UNUSUAL RESISTANCE WHEN TRYING TO REMOVE THE NEEDLE AND THUS TRIED TO GENTLY PULL ON WIRE AND FOUND THAT IT STARTED TO UNRAVEL. THEY THEN SLOWLY PULLED BOTH THE WIRE AND NEEDLE OUT, BUT FOUND THAT THE THIN UNRAVELED WIRE WAS STILL IN THE VESSEL, WHICH GOT LONGER AS IT WAS REMOVED. THE ENTIRETY OF THE WIRE WAS REMOVED, DESPITE THE UNRAVELING. A HEMATOMA HAD FORMED AFTER THIS AND MADE THE NEXT PUNCTURE MORE CHALLENGING. THIS RESULTED IN THE NEED TO ACCESS THROUGH THE NECK AT A CRITICAL VESSEL AND WAS ULTIMATELY SUCCESSFUL. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10698261 ↗InjuryApr 23, 2018Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN ANGIOPLASTY OF THE PULMONARY VALVE, THE MICROPUNCTURE TRANSITIONLESS PEDIATRIC ACCESS SET SHEATH WAS FOUND TO BE SPLIT INTO TWO PIECES. THIS EVENT OCCURRED AT SOME POINT DURING CANNULATION OF THE FEMORAL VEIN AND WAS DISCOVERED WHEN THE SHEATH WAS REMOVED. ONLY THE PROXIMAL PORTION WAS EXTRICATED OVER ANOTHER MANUFACTURER'S WIRE. THE DISTAL PORTION OF THE SHEATH WAS SEEN LODGED IN THE FEMORAL VEIN. THE GROIN AREA HAD SCARRING FROM A PREVIOUS CUT DOWN PROCEDURE BUT REPORTEDLY, THIS DID NOT MAKE THE INSERTION OF THE SHEATH DIFFICULT. THE COMPLAINANT POSTULATED THE SHEATH MUST HAVE SPLIT WHILE IN THE PATIENT; THEREFORE, IT COULD NOT HAVE BEEN CUT SINCE THE WIRE WAS WHOLLY INTACT UPON REMOVAL. THE COMPLAINANT REPORTED, THE SEPARATED PORTION OF THE SHEATH WAS SURGICALLY REMOVED AND THE PATIENT IS DOING WELL.
Patient problems: No Code Available
MAUDE report 7451365 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page