Micropuncture G47948Recalls & reported problems
Cook Incorporated · Micropuncture Introducer Set Silhouette Transitionless With Stainless Steel Wire Guide · Class 2
All models of Micropuncture (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00827002479487), received between Apr 28, 2021 and Apr 28, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 28, 2021Material Separation; Material Separation
ADDITIONAL INFORMATION WAS RECEIVED VIA MEDWATCH MW5101055. RESISTANCE WAS ENCOUNTERED UPON WITHDRAWAL OF THE WIRE GUIDE. FLUOROSCOPY WAS USED, AND MULTIPLE ATTEMPTS WERE MADE TO REMOVE THE WIRE, USING SHEATHS AND A CATHETER. THE WIRE SEPARATED AFTER SEVERAL RETRIEVAL ATTEMPTS, LEAVING THE TIP IN THE PATIENT. A VENOTOMY OF THE BRACHIAL VEIN WAS PERFORMED TO REMOVE THE TIP OF THE WIRE, AND THE VEIN WAS REPAIRED.
Patient problems: Foreign Body In Patient
MAUDE report 11738170 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page