Flexor G49043Recalls & reported problems
Cook Incorporated · Flexor, Check-Flo, Introducer Ansel 1 Modification - with High-Flex Dilator and Hydrophilic Coating · Class 2
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00827002490437), received between Jun 21, 2018 and Nov 27, 2024 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Material Separation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation4
- Detachment of Device or Device Component3
- Break1
- Fluid/Blood Leak1
- Material Split, Cut or Torn1
- Peeled/Delaminated1
- Stretched1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions4
- No Code Available1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20799527 ↗MalfunctionNov 10, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12791914 ↗MalfunctionSep 1, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED ON 08DEC2021: THE DILATOR WAS IN PLACE WHEN THE SEPARATION OCCURRED. IT IS PRESUMED THAT THERE WAS A WIRE IN THE LUMEN OF THE SHEATH WHEN THE SEPARATION OCCURRED. THE SHEATH HUB WAS NOT USED TO PULL THE DEVICE FROM THE PATIENT. THE DEVICE WAS INSIDE OF THE PATIENT WHEN THE SEPARATION OCCURRED. THERE WERE NO SHARP EDGES NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12408112 ↗MalfunctionJul 28, 2021Break; Material Separation; Break; Material Separation
ADDITIONAL INFORMATION WAS RECEIVED 24SEP2021. THE PATIENT LOST AN ESTIMATED 20ML BLOOD DURING THE PROCEDURE. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL INTERVENTION OR EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS BLOOD LOSS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12238602 ↗MalfunctionAug 19, 2020Material Separation; Stretched; Material Separation; Stretched
AS REPORTED, PRIOR TO USE, AS THE DOCTOR WAS ATTEMPTING TO SHAPE THE DEVICE, A FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH STRETCHED AND "RIPPED" AT THE SHAFT, EXPOSING THE COILS OF THE DEVICE. THE DILATOR WAS IN PLACE AT THE TIME OF THE EVENT. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE, AN ANGIO OF THE RIGHT LEG, SUCCESSFULLY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10427182 ↗MalfunctionJul 21, 2020Peeled/Delaminated; Peeled/Delaminated
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10301483 ↗MalfunctionMar 26, 2020Fluid/Blood Leak; Fluid/Blood Leak
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9884499 ↗MalfunctionMar 3, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING AN UNKNOWN PROCEDURE, A FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH SEPARATED AT THE HUB DURING REPOSITIONING OF THE DEVICE. AN UNKNOWN GUIDE WIRE WAS IN THE LUMEN OF THE DEVICE WHEN THE HUB SEPARATED. NORMAL RESISTANCE WAS REPORTED AS THE SHEATH WAS MOVED OVER THE CALCIFIED AORTIC BIFURCATION. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9781404 ↗MalfunctionJan 24, 2019Material Separation; Material Separation
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8274966 ↗InjuryJun 21, 2018Material Separation; Material Separation
IT WAS REPORTED, DURING RECANALIZATION OF THE PATIENT'S CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA), THE TIP OF THE FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH BROKE AND BECAME LODGED IN THE PATIENT'S ANATOMY. AN INCISION WAS MADE IN THE PATIENT'S LEG IN ORDER TO RETRIEVE THE SEPARATED PORTION OF THE COMPLAINT DEVICE. THE RETRIEVAL WAS SUCCESSFUL AND THE PATIENT REPORTEDLY WAS "OK" AFTER THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE AFTER THE RETRIEVAL OF THE BROKEN DEVICE COMPONENT.
Patient problems: No Code Available
MAUDE report 7625473 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.