Fusion G50976Recalls & reported problems
Cook Incorporated · Fusion Looptip Wire Guide · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00827002509764), received between Jun 29, 2016 and Aug 19, 2016 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Bent1
- Fracture1
- Material Deformation1
- Migration or Expulsion of Device1
Patient problems
- No Consequences Or Impact To Patient2
- Foreign Body In Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2016Bent; Migration or Expulsion of Device; Bent; Migration or Expulsion of Device
DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK FUSION LOOP TIP WIRE GUIDE. AFTER A VERY DIFFICULT CANNULATION USING THE FUSION LOOP TIP WIRE GUIDE (INSTEAD OF THE PRE-LOADED WIRE GUIDE), THE SPHINCTEROTOME WAS HARD TO PEEL OFF [EXCHANGE]. FINALLY, THEY LOST ACCESS FROM THE BILE DUCT. THE WIRE GUIDE WAS BENT, SO IT COULD NOT BE USED ANYMORE. THEY HAD TO USE NEW PRODUCTS AND RE-CANNULATE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5888880 ↗InjuryJun 29, 2016Fracture; Material Deformation; Fracture; Material Deformation
DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD), THE PHYSICIAN USED A COOK FUSION LOOP TIP WIRE GUIDE. THE DISTAL END OF THE WIRE UNRAVELED IN THE PSEUDOCYST. A PORTION OF THE DEVICE WAS REMOVED USING A PEDIATRIC BIOPSY FORCEP. A DIFFERENT WIRE GUIDE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: Foreign Body In Patient; No Consequences Or Impact To Patient
MAUDE report 5759803 ↗
Other Endoscopic guidewire, gastroenterology-urology devices
- HiWire G30474Cook Incorporated · Endoscopic guidewire, gastroenterology-urology82 reports
- Roadrunner G18154Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall18 reports
- Roadrunner G17542Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall9 reports
- Roadrunner G18153Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall6 reports
- Roadrunner G18155Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall3 reports
- Roadrunner G17539Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall2 reports
- Roadrunner G17540Cook Incorporated · Endoscopic guidewire, gastroenterology-urology1 recall0 reports
- Jagwire M00556580Boston Scientific Corporation · Endoscopic guidewire, gastroenterology-urology1 death49 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page