Flexor G56227Recalls & reported problems
Cook Incorporated · Flexor, Check-Flo, Introducer Ansel 1 Modification - with High-Flex Dilator and Hydrophilic Coating · Class 2
All models of Flexor (134)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00827002562271), received between Feb 11, 2019 and Sep 30, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Contamination /Decontamination Problem1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2025Break; Break
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23178473 ↗MalfunctionSep 25, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING A FENESTRATED ENDOVASCULAR DEVICE IMPLANTATION FOR A TYPE 1A ENDOLEAK INVOLVING ANOTHER MANUFACTURER¿S DEVICE THAT HAD BEEN INDWELLING FOR APPROXIMATELY EIGHT YEARS, THE HUB OF A FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH SEPARATED UPON INTRODUCTION OF THE DEVICE. THE FLEXOR SHEATH AND DILATOR WERE INSERTED INTO A RENAL ARTERY OVER A 260-CENTIMETER ROSEN WIRE, THROUGH A LARGE COOK SHEATH, FOR A FENESTRATED PROXIMAL EXTENSION AND RE-LINE OF ANOTHER MANUFACTURER¿S ¿OLD¿ INDWELLING GRAFT. PER THE REPORTER, RESISTANCE WAS ENCOUNTERED UPON INSERTION OF THE SHEATH, RESULTING IN THE DEVICE TWISTING DURING ADVANCEMENT. THE ILIAC ARTERIES WERE TORTUOUS, AND THE PRESENCE OF THE INDWELLING GRAFT REPORTEDLY MADE IT DIFFICULT TO TRACK THE SHEATH. PER THE REPORTER, EVERY DEVICE THAT WAS REMOVED THROUGHOUT THE CASE SHOWED SIGNIFICANT DISTORTION FROM THE INDWELLING GRAFT AND THE PATIENT¿S ANATOMY. AFTER THE HUB SEPARATED, THE SHEATH WAS EASILY REMOVED FROM THE PATIENT "IN THE NORMAL MANNER." THERE WERE NO ADDITIONAL COMPLICATIONS. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES OR EXPERIENCE ANY ADVERSE EFFECTS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20305416 ↗MalfunctionFeb 11, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Patient Involvement
MAUDE report 8325106 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page