Flexor G56228Recalls & reported problems
Cook Incorporated · Flexor, Check-Flo, Introducer Ansel 1 Modification - with High-Flex Dilator and Hydrophilic Coating · Class 2
All models of Flexor (134)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00827002562288), received between Jan 31, 2017 and Dec 10, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component8
- Material Separation2
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20891530 ↗MalfunctionMay 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING A PROCEDURE INVOLVING "TBRANCH/VISCERAL VESSEL PROBING" VIA TRANS-AXILLARY ACCESS, THE SIDE-ARM SEPARATED FROM A FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH'S HUB WHEN THE USER CONNECTED A SYRINGE FOR INJECTION OF CONTRAST. THE DILATOR WAS NOT IN THE SHEATH WHEN THE SIDE-ARM SEPARATED, NOR WAS ANY OTHER DEVICE. THE ANATOMY WAS REPORTEDLY NORMAL, AND RESISTANCE WAS NOT ENCOUNTERED UPON INSERTION OR REMOVAL OF THE SHEATH. THE PATIENT RECEIVED HEPARIN DURING THE PROCEDURE. A POWER INJECTOR WAS NOT USED. THE PROCEDURE WAS COMPLETED WITH A NEW SHEATH. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19393352 ↗MalfunctionJan 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED (VIA A PROSPECTIVE, POST-MARKET, MULTICENTER, OBSERVATIONAL STUDY), DURING A PROCEDURE INVOLVING TREATMENT FOR A THORACOABDOMINAL AORTIC ANEURYSM, THE HUB SEPARATED FROM A FLEXOR HIGH-FLEX ANSEL GUIDING SHEATH UPON REMOVAL OF THE DEVICE. PER THE REPORTER, THE HUB SEPARATED DUE TO EXTREME FRICTION WITH THE SHEATH AND WIRES. ACCESS WAS OBTAINED IN THE RIGHT AND LEFT FEMORAL ARTERIES AND THE LEFT AXILLARY ARTERY DURING THE PROCEDURE. THE ACCESS SITE WAS NOT SCARRED, AND THE ANATOMY WAS NOT STENOSED, TORTUOUS, OR CALCIFIED. THE COMPLAINT SHEATH WAS ADVANCED THROUGH AN UNSPECIFIED 12-FRENCH SHEATH VIA THE LEFT AXILLARY ARTERY. RESISTANCE WAS ENCOUNTERED UPON BOTH INSERTION AND REMOVAL OF THE COMPLAINT SHEATH, REPORTEDLY DUE TO AN UNSPECIFIED WIRE THAT WAS INSERTED PARALLEL TO THE DEVICE, INSIDE THE UNSPECIFIED 12-FRENCH SHEATH. THE SHEATH SUCCESSFULLY ACCESSED THE TARGET VESSEL AND PERFORMED AS INTENDED, AND AN UNSPECIFIED BRIDGING STENT WAS DELIVERED THROUGH THE SHEATH. THE HUB SEPARATED AS THE COMPLAINT DEVICE WAS REMOVED OVER AN UNSPECIFIED AMPLATZ STIFF WIRE, BY PULLING FROM THE SHEATH HUB. THE DILATOR WAS NOT REINSERTED PRIOR TO SHEATH REMOVAL. SIDE-BRANCH CATHETERIZATI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18598528 ↗MalfunctionFeb 11, 2021Material Separation; Unraveled Material; Detachment of Device or Device Component; Material Separation; Unraveled Material; Detachment of Device or Device Component
NO NEW PATIENT OR EVENT INFORMATION TO REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11314533 ↗MalfunctionJul 17, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10290171 ↗MalfunctionJun 10, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10138719 ↗MalfunctionApr 20, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING A T-BRANCH PROCEDURE, TWO FLEXOR HIGH-FLEX ANSEL GUIDING SHEATHS SEPARATED AT THE HUB. ACCESS WAS OBTAINED IN THE BRACHIAL ARTERY FOR CANNULATION OF THE RENAL ARTERY. AN UNKNOWN 5 FRENCH CATHETER AND UNKNOWN HYDROPHILIC WIRE GUIDE WERE USED DURING THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT IS NOT AVAILABLE AT THIS TIME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9977612 ↗MalfunctionJul 17, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED, DURING REMOVAL OF TWO FLEXOR HIGH-FLEX ANSEL GUIDING SHEATHS (SAME PRODUCT AND LOT NUMBER FOR BOTH DEVICES), THE "STOP RING" (VALVE) WAS TORN OFF. NO ADDITIONAL INFORMATION WAS PROVIDED. NO ADVERSE EFFECT HAS BEEN REPORTED. ADDITIONAL INFORMATION REGARDING EVENT DETAILS, PATIENT ANATOMY AND OUTCOME HAS BEEN REQUESTED, BUT IS NOT AVAILABLE AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7694037 ↗MalfunctionJan 31, 2017Material Separation; Material Separation
INTERNATIONAL CUSTOMER REPORTED THAT "THE VENT GOT TORN OFF." ADDITIONAL INFORMATION WAS REQUESTED BUT NOT PROVIDED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6291431 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page