Blue Rhino G57685Recalls & reported problems

Cook Incorporated · Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set · Class 2

All models of Blue Rhino (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00827002576858), received between Sep 14, 2021 and Sep 9, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K193133); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 5, 2026Class IIOpen, Classified

    Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

    Z-1779-2026 · Root cause: Incorrect or no expiration date · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device-Device Incompatibility1
  • Difficult to Advance1
  • Insufficient Device Problem Information1
  • Material Deformation1
  • Material Separation1
  • Material Split, Cut or Torn1
  • Material Too Soft/Flexible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1
  • Pneumothorax1
  • Upper Respiratory Tract Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE TIP OF THE 14 FRENCH PUNCH DILATOR INCLUDED IN THE BLUE RHINO G2-MULTI PERCUTANEOUS TRACHEOSTOMY INTRODUCER SET SEPARATED. AT THIS TIME, NO ADVERSE EFFECTS OR ADDITIONAL PROCEDURES FOR THE PATIENT WERE REPORTED DUE TO THIS OCCURRENCE. ADDITIONAL INFORMATION REGARDING THE EVENT AND PATIENT OUTCOME HAS BEEN REQUESTED BUT IS CURRENTLY UNAVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23005640 ↗
  • InjuryMar 24, 2025Material Separation; Insufficient Device Problem Information; Material Separation; Insufficient Device Problem Information

    IT WAS REPORTED THAT TRACHEAL INJURY OCCURRED DURING USE OF A BLUE RHINO G2-MULTI PERCUTANEOUS TRACHEOSTOMY INTRODUCER SET. A 61-YEAR-OLD MALE UNDERWENT A BEDSIDE BRONCHOSCOPY AND PERCUTANEOUS TRACHEOSTOMY PLACEMENT PROCEDURE. THE GUIDE WIRE WAS INSERTED FIRST, AND THEN THE BLUE RHINO DILATOR WAS USED TO FURTHER DILATE THE TRACHEOSTOMY SITE. A #8 CUFFED TRACHEOTOMY TUBE WAS INSERTED AND THEN WAS REMOVED TO CLARIFY THE PLACEMENT. THE BRONCHOSCOPY WAS ADVANCED THROUGH THE ENDOTRACHEAL TUBE (ETT), AND A TEAR IN THE POSTERIOR TRACHEAL WALL WAS VISUALIZED. VENTILATION WAS THEN RESUMED. DIAGNOSTIC STUDIES, INCLUDING A COMPUTED TOMOGRAPHY SCAN AND CHEST X-RAY, AND A THORACIC CONSULT WERE ORDERED. THE PATIENT WAS NOT A SURGICAL CANDIDATE DUE TO HIGH CO-MORBIDITIES AND REQUIRED A DOUBLE LUMEN ETT. PER PHYSICIAN NOTES, BILATERAL CHEST TUBES WERE INSERTED SECONDARY TO THE INCIDENT. THE INDICATION FOR THE RIGHT SIDED CHEST TUBE WAS A PNEUMOTHORAX RESULTING FROM THE INCIDENT. THE LEFT SIDED CHEST TUBE WAS INSERTED AS A PROPHYLACTIC MEASURE SECONDARY TO THE HIGH RISK OF PNEUMOTHORAX. IT IS UNKNOWN IF THE TRACHEAL WALL INJURY WAS CAUSED BY THE WIRE GUIDE OR THE BLUE RHINO DILATOR. THE CUSTOMER BE

    Patient problems: Pneumothorax; Upper Respiratory Tract Injury

    MAUDE report 21678694 ↗
  • MalfunctionAug 4, 2023Device-Device Incompatibility; Difficult to Advance; Device-Device Incompatibility; Difficult to Advance

    NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17465631 ↗
  • MalfunctionSep 14, 2021Material Deformation; Material Too Soft/Flexible; Material Deformation; Material Too Soft/Flexible

    IN ADDITIONAL INFORMATION RECEIVED ON 22SEP2021, IT WAS REPORTED THAT THIS WAS A COVID PATIENT THAT REQUIRED VENTILATION. IT WAS CONFIRMED THAT DIFFICULTY WAS EXPERIENCED DURING INITIAL PLACEMENT OF THE DEVICE INTO THE TRACHEA, "WHEN RAILROADING THE HORN OVER THE WIRE, THE TIP BUCKLED AND WAS NOT FIRM ENOUGH TO ENTER THE TRACHEA." THE TUBE WAS SUCCESSFULLY PLACED USING A SET FROM ANOTHER MANUFACTURER. THE PATIENT DID NOT HAVE ANATOMY THAT WOULD HAVE MADE PERCUTANEOUS TRACHEOSTOMY DIFFICULT TO PERFORM. IMAGING WAS NOT BEING USED AT THE TIME OF THE PROCEDURE. THE ACCESS SITE WAS PRE-DILATED WITH THE 14FR INTRODUCER DILATOR, AND THE HYDROPHILIC COATING OF THE BLUE RHINO WAS ACTIVATED. THE BLUE RHINO WAS ADVANCED WITH THE WHITE GUIDING CATHETER OVER THE WIRE. NO ADDITIONAL LUBRICANT WAS REQUIRED FOR ADVANCEMENT. IT WAS ALSO REPORTED THAT THE USER MAY HAVE NOT "DISSECTED DOWN ENOUGH", AS A "FIRMER" MODEL WOULD NOT REQUIRE THEM TO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12468196 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K193133. One submission often covers many models, so check the product and lot details in each record.

  • Z-1780-2026 · Mar 5, 2026 · Open, Classified
    COOK MEDICAL Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray: Reference Part Number C-PTISY-100-HC-G-NA, Order Number G57703; Reference Part Number C-PTISY-100-UNL-
  • Z-1428-2023 · Apr 7, 2023 · Open, Classified
    Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Tray, Reference Part Numbers C-PTISYJ-100-HC-G-EU-FLEX8.5 (G57725), C-PTISYJ-100-HC-G-NA-FLEX7.5 (G57720), C-PTISYJ-100-HC-
  • Z-1427-2023 · Apr 7, 2023 · Open, Classified
    Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FL

Other Tube tracheostomy and tube cuff devices