Skip to content

EchoTip AcuCore G59746Recalls & reported problems

Cook Ireland Ltd · EchoTip AcuCore EUS Biopsy Needle · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00827002597464), received between Mar 12, 2024 and Mar 18, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method3
  • Physical Resistance/Sticking2
  • Break1
  • Loss of or Failure to Bond1
  • Material Twisted/Bent1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2025Retraction Problem; Material Twisted/Bent; Retraction Problem; Material Twisted/Bent

    THE PROVIDE EXPLAINED THAT AFTER PASS 4 WITH THE NEEDLE THEY WERE UNABLE TO FULLY RETRACT THE NEEDLE INTO THE SHEATH AND WAS ALSO UNABLE TO ADVANCE THE STYLET ALL THE WAY BACK INTO THE NEEDLE. THE CASES WERE COMPLETED WITH A COMPETITOR DEVICE. THE FOLLOWING INFORMATION HAS BEEN RECEIVED VIA EMAIL ON (B)(6) 2025. 2. IS AN ACKNOWLEDGMENT LETTER (FORMAL NOTIFICATION OF THE COMPLAINT BEING RECEIVED) REQUIRED? NO 3. IS ANY ACCOUNT ACTION NEEDED (CREDIT, NO CHARGE REPLACEMENT, NO ACTION, INVESTIGATION RESULTS, ETC.)? NO CHARGE REPLACEMENT. 4. WILL THE FACILITY BE REPORTING THIS TO A GOVERNMENT AGENCY (FDA, COMPETENT AUTHORITY, ETC.) NO 5. CAN YOU PROVIDE ANY PATIENT INFORMATION (AGE, WEIGHT, GENDER, PREEXISTING CONDITIONS)? NO 6. WHAT TYPE OF PROCEDURE WAS BEING PERFORMED? ENDOSCOPIC ULTRASOUND WITH FINE NEEDLE BIOPSY. 7. PLEASE DESCRIBE THE LOCATION IN THE BODY FOR THE INTENDED TARGET SITE (PANCREAS, STOMACH, LUNGS, ETC.) PANCREAS. 1. (IF LUNGS, WHICH LYMPH NODE WAS BEING TARGETED? E.G., 2R, 2L, 4R, AO, AR, 11RI, 11S, ETC.). 8. PLEASE DESCRIBE THE SIZE OF THE INTENDED TARGET SITE. UNKNOWN. 9. WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? NO 10. WAS PUNCTURE OF THE TARGETED SITE DIFFI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21632645 ↗
  • MalfunctionOct 23, 2024Break; Break

    A SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO COMPLETION OF THE INVESTIGATION ON 14-FEB-2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20513615 ↗
  • MalfunctionMay 29, 2024Improper or Incorrect Procedure or Method; Physical Resistance/Sticking; Improper or Incorrect Procedure or Method; Physical Resistance/Sticking

    DESCRIPTION OF EVENT: AS INITIALLY REPORTED TO CUSTOMER RELATIONS: A PATIENT OF UNDISCLOSED GENDER AND AGE UNDERWENT AN EBUS WITH FINE NEEDLE BIOPSY PROCEDURE OF THE GASTRIC HEPATIC LYMPH NODE IN WHICH THE ECHOTIP ACUCORE ULTRASOUND BIOPSY NEEDLE, G59746, WAS USED. THE ECHOTIP NEEDLE WOULD NOT PENETRATE THE LESION. DEVICE WAS REMOVED FROM THE PATIENT AND PROCEDURE COMPLETED WITH AN OLYMPUS VISISHOT NEEDLE. PATIENT/EVENT INFO - NOTES: 4.1 FOR ALL COMPLAINTS, ASK: ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? NO. IF THE REPORT INVOLVES A KINK OR BEND IN THE NEEDLE, WHERE IS THIS LOCATED ON THE DEVICE (HANDLE END (PROXIMAL END) OR PATIENT END (DISTAL END))? NO BEND OR KINK WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DEVICE PRIOR TO RETURN (E.G. KINK)? NO. PLEASE SPECIFY IF YES. IF THE DEVICE WAS KINKED BELOW THE SHEATH EXTENDER, WAS THE KINK OBSERVED BEFORE INSERTING THE DEVICE INTO THE SCOPE? N/A. IF THE DEVICE IS A PROCORE NEEDLE, IS THE DEVICE DAMAGE LOCATED AT THE NOTCH / CORE TRAP? N/A. IF NO, PLEASE SPECIFY WHERE THE DAMAGE IS LOCATED: WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? NO. WAS THE DEVICE USED IN A TORTUOUS POSITION? NO. WAS PUNCTURE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19413307 ↗
  • MalfunctionMay 22, 2024Improper or Incorrect Procedure or Method; Physical Resistance/Sticking; Improper or Incorrect Procedure or Method; Physical Resistance/Sticking

    DESCRIPTION OF EVENT: PER REP - 25APR2024 - THE DOCTOR FOUND IT DIFFICULT TO PUNCTURE THE LESION. THEY DID GET IT WORK WITH DIFFICULTY FOR TWO PASSES BUT USED A COMPETITIVE NEEDLE FOR A THIRD PASS. PATIENT/EVENT INFO - NOTES: 4.0 RPN PREFIX ECHO 4.1 FOR ALL COMPLAINTS, ASK: ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? N/A, YES, NO -NO IF THE REPORT INVOLVES A KINK OR BEND IN THE NEEDLE, WHERE IS THIS LOCATED ON THE DEVICE (HANDLE END (PROXIMAL END) OR PATIENT END (DISTAL END))? N/A, HANDLE END, PATIENT END -N/A WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DEVICE PRIOR TO RETURN (E.G. KINK)? N/A, YES, NO -NO PLEASE SPECIFY IF YES. IF THE DEVICE WAS KINKED BELOW THE SHEATH EXTENDER, WAS THE KINK OBSERVED BEFORE INSERTING THE DEVICE INTO THE SCOPE? N/A, YES, NO -N/A IF THE DEVICE IS A PROCORE NEEDLE, IS THE DEVICE DAMAGE LOCATED AT THE NOTCH / CORE TRAP? N/A, YES, NO -N/A IF NO, PLEASE SPECIFY WHERE THE DAMAGE IS LOCATED: WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? N/A, YES, NO -NO WAS THE DEVICE USED IN A TORTUOUS POSITION? N/A, YES, NO -NO WAS PUNCTURE OF THE TARGET SITE DIFFICULT? N/A, YES, NO -YES PLEASE DESCRIBE THE ANATOMICAL LOCATION OF THE I

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19370086 ↗
  • MalfunctionMar 12, 2024Loss of or Failure to Bond; Improper or Incorrect Procedure or Method; Loss of or Failure to Bond; Improper or Incorrect Procedure or Method

    AS REPORTED TO CUSTOMER RELATIONS VIA COMPLAINT EMAIL "THE DISTAL END OF THE HANDLE BECAME DETACHED FROM THE NEEDLE. THE NEEDLE WAS REMOVED FROM THE SCOPE AND THE HANDLE WAS REATTACHED TO THE NEEDLE AND SPECIMEN WAS OBTAINED. THE PROCEDURE WAS COMPLETED USING A COMPETITIVE DEVICE." PATIENT/EVENT INFO - NOTES: 4.1 FOR ALL COMPLAINTS, ASK: ARE IMAGES OF THE DEVICE OR PROCEDURE AVAILABLE? N/A, YES, NO IF THE REPORT INVOLVES A KINK OR BEND IN THE NEEDLE, WHERE IS THIS LOCATED ON THE DEVICE (HANDLE END (PROXIMAL END) OR PATIENT END (DISTAL END))? N/A, HANDLE END, PATIENT END WERE ANY OTHER DEFECTS (OTHER THAN THE COMPLAINT ISSUE) OBSERVED ON THE DEVICE PRIOR TO RETURN (E.G. KINK)? N/A, YES, NO ¿ PLEASE SPECIFY IF YES. IF THE DEVICE WAS KINKED BELOW THE SHEATH EXTENDER, WAS THE KINK OBSERVED BEFORE INSERTING THE DEVICE INTO THE SCOPE? N/A, YES, NO IF THE DEVICE IS A PROCORE NEEDLE, IS THE DEVICE DAMAGE LOCATED AT THE NOTCH / CORE TRAP? N/A, YES, NO ¿ IF NO, PLEASE SPECIFY WHERE THE DAMAGE IS LOCATED: WAS GAINING ACCESS TO THE TARGET SITE DIFFICULT? N/A, YES, NO WAS THE DEVICE USED IN A TORTUOUS POSITION? N/A, YES, NO WAS PUNCTURE OF THE TARGET SITE DIFFICULT? N/A, YES, NO PLEASE DESCRIBE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18886258 ↗

Other Endoscopic ultrasound system, gastroenterology-urology devices

All Endoscopic ultrasound system, gastroenterology-urology devices →

Sources and updates