Skip to content

EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE OLYMPUS GF TYPE UCT180 (UDI 04953170287879)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

All models of EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Olympus EVIS EXERA II ultrasound endoscope is an endoscopic ultrasound system for gastroenterology and urology. Reports most often describe parts of the scope peeling, cracking, breaking or separating, and some mention contamination of the device itself. There are 260 reports, no deaths and 2 injuries. Reports rose to 78 in the last 12 months from 21 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
266
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 266 adverse event reports that list this exact device (DI 04953170287879), received between Dec 28, 2018 and Sep 28, 2026 — classified as 2 injuries, 264 malfunctions.

The most frequently reported specific device problem is “Microbial Contamination of Device” (52 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 4 reportsJan 2024: 7 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 4 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 7 reportsSep 2024: 5 reportsOct 2024: 7 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 4 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 1 reportsNov 2025: 9 reportsDec 2025: 5 reportsJan 2026: 8 reportsFeb 2026: 6 reportsMar 2026: 3 reportsApr 2026: 9 reportsMay 2026: 18 reportsJun 2026: 4 reportsJul 2026: 10 reportsAug 2026: 5 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Microbial Contamination of Device52
  • Material Separation42
  • Peeled/Delaminated35
  • Break29
  • Crack27
  • Degraded20
  • Material Split, Cut or Torn15
  • Detachment of Device or Device Component14

Patient problems

  • No Clinical Signs, Symptoms or Conditions227
  • No Consequences Or Impact To Patient16
  • No Known Impact Or Consequence To Patient11
  • No Patient Involvement10
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Bacterial Infection1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 266 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Break

    The customer reported the ultrasound gastrovideoscope for a separate issue. During the initial evaluation of the returned device, a chip was observed in the adhesive area between the acoustic lens and the ultrasound probe. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917967 ↗
  • MalfunctionSep 24, 2026Microbial Contamination of Device

    It was reported that, during reprocessing, the Ultrasound Gastrovideoscope tested positive for less than 1 Colony Forming Units (CFUs) of Penicillium, Micrococcus luteus, and Coagulase-negative Staphylococcus. All channels were sampled. The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50888206 ↗
  • MalfunctionSep 14, 2026Device Reprocessing Problem; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50761643 ↗
  • MalfunctionSep 11, 2026Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50743381 ↗
  • MalfunctionSep 11, 2026Crack

    The customer reported the Ultrasound Gastrovideoscope had a separate issue. During the initial evaluation of the device it was observed, Ultrasonic probe has a crack noted. No patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50725421 ↗
  • MalfunctionSep 10, 2026Detachment of Device or Device Component

    The customer returned the ultrasound gastrovideoscope, an Olympus asset, with no reported complaint. During the evaluation of the device, wear of the adhesive around the objective lens was observed. The service repair technician team assessed the condition and determined that the issue was most likely due to component failure. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717496 ↗
  • MalfunctionSep 1, 2026Break; Crack; Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50550669 ↗
  • MalfunctionAug 27, 2026Material Separation

    The customer reported that the Ultrasound Gastrovideoscope had a separate issue. During the initial evaluation of the device, a gap between acoustic lens and ultrasound probe was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478198 ↗
  • MalfunctionAug 11, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50212712 ↗
  • MalfunctionAug 7, 2026Corroded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177941 ↗
  • MalfunctionAug 4, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50130997 ↗
  • MalfunctionAug 3, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50112065 ↗
  • MalfunctionJul 30, 2026Degraded

    The customer returned the Ultrasound Gastrovideoscope to the service center for a separate issue. During the device analysis, the investigation indicates that the adhesive around the light guide lens had wear. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50082336 ↗
  • MalfunctionJul 28, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50060018 ↗
  • MalfunctionJul 28, 2026

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50058339 ↗
  • MalfunctionJul 27, 2026Degraded

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50040134 ↗
  • MalfunctionJul 27, 2026Device Reprocessing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50025121 ↗
  • MalfunctionJul 27, 2026Break; Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50007871 ↗
  • MalfunctionJul 22, 2026Crack; Degraded

    THE CUSTOMER RETURNED THE ULTRASOUND GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING THE REPAIR CENTER IDENTIFIED THE DEVICE HAD ADHESIVE AROUND OBJECTIVE LENS CHIPPED AND ADHESIVE AROUND LIGHT GUIDE LENS HAD WEAR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25904623 ↗
  • MalfunctionJul 21, 2026Break; Peeled/Delaminated

    THE CUSTOMER REPORTED SIGNIFICANT LOSS OF ADHESIVE AROUND THE IMAGE LENS DURING SERVICE/MAINTENANCE INVOLVING AN OLYMPUS ULTRASOUND GASTROVIDEOSCOPE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25887008 ↗
  • MalfunctionJul 1, 2026Crack

    THE CUSTOMER RETURNED THE ULTRASOUND GASTROVIDEOSCOPE WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING THE DEVICE ANALYSIS, THE INVESTIGATION INDICATES THAT CHIP IN ADHESIVE AROUND OBJECTIVE LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681999 ↗
  • MalfunctionJul 1, 2026Degraded

    THE CUSTOMER RETURNED THE ULTRASOUND GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE INVESTIGATION INDICATES THAT THERE WAS WEAR ON THE ADHESIVE AROUND THE OBJECTIVE LENS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25680592 ↗
  • MalfunctionJun 29, 2026Crack; Failure to Clean Adequately

    THE CUSTOMER RETURNED THE ULTRASOUND GASTROVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING DEVICE ANALYSIS, THE SERVICE TEAM INDICATED THAT THE ADHESIVE AROUND THE OBJECTIVE LENS HAD A CRACK AND FOREIGN OBJECTS CAME OUT OF THE DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25651011 ↗
  • MalfunctionJun 2, 2026Biofilm coating in Device; Detachment of Device or Device Component; Gas/Air Leak; Naturally Worn

    IT WAS REPORTED THE ULTRASOUND GASTROVIDEOSCOPE HAD THE DISTAL COVER WITH PRESENCE OF BIOFILM. THE ISSUE WAS FOUND SET UP. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25367707 ↗
  • MalfunctionJun 2, 2026Obstruction of Flow

    THE CUSTOMER REPORTED A PROTHESIS OBSTRUCTING THE INSTRUMENT CHANNEL DISPLAYED DURING INSPECTION BEFORE USE INVOLVING AN OLYMPUS ULTRASOUND GASTROVIDEOSCOPE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25364425 ↗

Other Endoscopic ultrasound system, gastroenterology-urology devices

All Endoscopic ultrasound system, gastroenterology-urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page